    Scout tm series balances - skx instruction manual.

  • Page 3: 1. Introduction

    Scout tm series balances - skx en-1 1. Introduction this manual contains installation, operation and maintenance instructions for the scout skx series balances. Please read the manual completely before using the balance . 1.1 definition of signal warnings and symbols safety notes are marked with sig...

  • Page 4: 2.2.3 Security Slot

    En-2 scout tm series balances - skx 2.1.1 releasing the shipping lock release the red transportation lock on the sub-platform of the balance by turning the red pointer 90 counter- clockwise. Figure 2-1. Transportation lock 2.1.2 installing the weighing pan balances with a rectangular pan are placed ...

  • Page 5: 2.4  Connecting Power

    Scout tm series balances - skx en-3 2.4 connecting power ac adapter installation ac power is used to power the scale when battery power is not needed. First, connect the ac adapter (supplied) to the ac adapter input receptacle at the rear of the balance then connect the ac plug to an electrical outl...

  • Page 6: 3. Operation

    En-4 scout tm series balances - skx 3. Operation 3.1 controls figure 3-1. Scout control panel table 3-1. Button functions button primary function (short press) zero/on turns the balance on if balance is on, sets zero print sends the current value to the selected com ports if autoprint is set to off....

  • Page 7: 3.3 Calibration Operation

    Scout tm series balances - skx en-5 figure 3-2. Scout display table 3-2. Display symbols item description item description 1 battery charge symbol 7 pointer symbols 2 stable weight symbol 8 kilogram, grain symbols 3 negative symbol 9 pound, ounce, pound:ounce symbols 4 center of zero symbol 10 g/mol...

  • Page 8: 3.5 Density

    En-6 scout tm series balances - skx 3.5 density mode this mode allows the user to calculate the specific gravity of a sample when using the density kit. Density mode has been initiated when “g” and the current unit is displayed. The sample is weighed in air suspended from the hook and then weighed i...

  • Page 9: 4. Menu Settings

    Scout tm series balances - skx en-7 4. Menu settings the user menu allows the customizing of balance settings. Note: additional sub-menus may be available if interface options are installed. See interface user manual for the additional setting information. 4.1 menu navigation user menu: sub-menus c....

  • Page 10: 4.3 Setup Menu

    En-8 scout tm series balances - skx 4.3 setup menu enter this menu to set balance parameters. Note: bold always represents factory default value. Reset [reset] reset the setup menu to factory defaults. No = not reset yes = reset filter [filter] set the amount of signal filtering. Low = less stabilit...

  • Page 11: 4.4 Mode Menu

    Scout tm series balances - skx en-9 auto off [a.Off] set the automatic shut off functionality. Off = disabled 1 = powers off after 1 minute of no activity 5 = powers off after 5 minutes of no activity 10 = powers off after 10 minutes of no activity end setup [end] advance to the next menu or return ...

  • Page 12: 4.6 Additional Features

    En-10 scout tm series balances - skx 4.6 additional features weigh below hook the scout balance is equipped with a weigh below hook for weighing below the balance. The weigh below hook is located at the reverse side of the battery cover as shown below. To use this feature, remove the red protective ...

  • Page 13: 5. Maintenance

    Scout tm series balances - skx en-11 5. Maintenance 5.1 cleaning warning: electric shock hazard. Disconnect the equipment from the power supply before cleaning. Electric shock hazard. The housing may be cleaned with a cloth dampened with a mild detergent if necessary. Attention: do not use solvents,...

  • Page 14: 6. Technical Data

    En-12 scout tm series balances - skx 6. Technical data the technical data is valid under the following ambient conditions: indoor use only operating temperature range: +5 °c to +40 °c relative humidity: 10% to 80% at 31°c, decreasing linearly to 50% at 40°c, non-condensing altitude: up to 2000 m pow...

  • Page 15

    Scout tm series balances - skx en-13 table 6-2. Specifications (cont.) capacity x readability: model skx123 skx222 skx422 skx622 skx1202 skx2202 gram (g) 120 x 0.001 220 x 0.01 420 x 0.01 620 x 0.01 1200 x 0.01 2200 x 0.01 kilogram (kg) / / / / 1.2 x 0.00001 2.2 x 0.00001 newton (n) 1.17679 x 0.0000...

  • Page 16: 6.2

    En-14 scout tm series balances - skx 6.2 drawings figure 6.1 dimensions model a b c with draftshield 202 mm / 8.0 in. 222 mm / 8.7 in. 103 mm / 4.1 in. W/o draftshield 202 mm / 8.0 in. 224 mm / 8.8 in. 54 mm / 2.1 in. 6.3 compliance compliance to the following standards is indicated by the correspon...

  • Page 17

    Scout tm series balances - skx en-15 fcc note this equipment has been tested and found to comply with the limits for a class b digital device, pursuant to part 15 of the fcc rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. ...

  • Page 18

    En-16 scout tm series balances - skx limited warranty ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period ohaus will repair, or, at its option, replace any component(s) that proves...

  • Page 20

    En-18 scout tm series balances - skx ohaus corporation 7 campus drive suite 310 parsippany, nj 07054 usa tel: +1 973 377 9000 fax: +1 973 944 7177 with offices worldwide / con oficinas alrededor del mundo / avec des bureaux dans le monde entier / weltweite geshäftsstellen / con uffici in tutto il mo...