OHAUS Valor 1000 Service Manual - page 36
APPENDIX A STANDARD CALIBRATION
A.2 SETUP
AND
CALIBRATION
TABLE A-1. INITIAL CONFIGURATION SETTINGS FOR EACH MODEL
MODEL
Division
(A----x)
Decimal point
(b----x)
Capacity
(c-----x)
Zero tracking
(d----x)
Calibration
Weight
V11P3 0.5g:
x=5
Kg: x=4
or g: x=1
3.0000kg
or 3000.0g
x=3
3.000
V11P6 1g:
x=1
Kg:x=3
or g: x=0
6.000kg
or 6000g
x=3
3.000
V11P15 2g:
x=2
Kg: x=3
or g: x=0
15.000kg
or 15000g
x=3
3.000
V11P30 5g:
x=5
Kg: x=3
or g: x=0
30.000kg
or 30000g
x=3
3.000
Follow these steps for initial Setup and Calibration:
1. Turn on the scale. When display flashes all segments (
888888
) press
and hold ZERO and SELECT at the same time.
SET--- appears at the end of power on sequence.
SET---
2. Press
SELECT until SET-F1 appears.
SET-F1
3. Press
ZERO to display A____x (x=divisions in grams: 1,2,5,10,20,50).
Press ZERO to accept the setting shown, or
press M+ or TARE to increment or decrement setting to match value
for specific model – see Table A-1.)
Note: Settings for each model must match Table A-1.
A___2
4. Press
ZERO to display b____x (x=decimal point: 0,1,2,3).
Press ZERO to accept the setting shown, or press M+ or TARE to
move up or down to match value for specific model – see Table A-1.
B___3
5. Press
ZERO to display Cxxxxx (xxxxx= capacity of scale).
Press ZERO to accept the setting shown, or press M+ or TARE to
move up or down to match value for specific model – see Table A-1.
(Press SELECT to move to the next digit.)
C 15.000
6. Press
ZERO to display dx (x=Zero tracking setting: 0-9 from 0.125d to
1.25d). Press ZERO to accept the setting shown, or M+ or TARE to
move up or down to match value for specific model – see Table A-1.
Setup is now complete. Calibration follows immediately:
D 3
7. Press
ZERO and display will show Load 0. Make sure pan is empty,
then press ZERO to accept. CAL 0 appears, then Load F.
Load 0
8. Press
ZERO to display Fxxxxx (xxxxx=Span weight ).
Note: Span weight is 3000 g (3 kg) for all models: DO NOT CHANGE.
Put displayed weight on the scale, and press ZERO to accept.
After completion, scale will be in weighing mode.
F 3.000
Ohaus Corporation www.ohaus.com
A-2
Valor™ 1000 Series Service Manual