    Printed in china 120 80 no.1 en,sp © 2002 matsushita electric works, ltd. All rights reserved. For questions or assistance with your blood pressure monitor, call us at 1-800-338-0552. Panasonic consumer electronics company division of panasonic corporation of north america one panasonic way 3d-1 sec...

  • Page 2

    2 1 table of contents introduction ...................................................................... 3 basics of blood pressure................................................. 3 important instructions before use ................................... 4 precautions to ensure safe, reliable operati...

  • Page 3

    4 3 important instructions before use 1. Do not confuse self-monitoring with self-diagnosis. Blood pressure measurements should only be interpreted by a health professional who is familiar with your medical history. 2. If you are taking medication, consult with your physician to determine the most a...

  • Page 4

    6 5 precautions to ensure safe, reliable operation 1. Do not drop the unit. Protect it from sudden jars or shocks. 2. Do not insert foreign objects into any openings. 3. Do not attempt to disassemble the unit. 4. Do not crush the pressure cuff. 5. If the unit has been stored at temperatures below 0°...

  • Page 5

    8 7 name and function of each product part 12 0 80 1 2 blood pressure indicator led display memory/recall button set button adjust button main body storage case view from side pressure cuff battery cover on/off and start button 2cm gauge fig. 1 important: • do not be alarmed by temporarily high or l...

  • Page 6

    10 9 setting the time and date having inserted the batteries, set the time and date so that any future stored measurements are saved in the memory with the correct time and date. 1. Press the set button “1” on the side of the unit for at least 3 seconds. The displayed date will flash on and off (fig...

  • Page 7

    12 1. Remove your watch and any jewelry from your left hand. 2. Wrapping the cuff around the wrist and securing with hook-and-loop fastener. Leave main unit at the distance of 2 cm (3/4") from the first wrinkle line of wrist-palm joint, mount the cuff on the wrist in such way as the center of the ma...

  • Page 8

    14 13 if the blood pressure monitor is positioned lower than the heart, the measured blood pressure value may be high. If the blood pressure monitor is positioned higher than the heart, the measured blood pressure value may be low. If the posture applies pressure to the stomach (leaning forward), ac...

  • Page 9

    16 15 measuring your blood pressure relax during measurements. 1. Press the on/off and start button (fig. 7-a). • all displays on the unit will light up for approx. 2 seconds (fig. 7-b). • the cuff will begin to inflate automatically. Changing inflation speeds and operating sounds occur during infla...

  • Page 10

    18 17 m / r caution: • for persons with poor circulation resulting from diabetes, liver disease, arteriosclerosis, high blood pressure or other conditions, there may be significant differences in blood pressure values measured at the wrist versus at the upper arm. • do not confuse self-monitoring wi...

  • Page 11

    20 setting (checking) the measurement notification time 19 m / r calling up readings from memory readings can be recalled after being saved without having to switch the unit on. 1. Press the memory/recall button. • the average readings of all recorded measurements are displayed. • the mark will be d...

  • Page 12

    22 21 5. (1) oprima el botón puesta. El valor de los minutos destellará (fig. 12-f). (2) oprima el botón ajustar repetidamente para establecer los minutos deseados. • para ajustar sólo la hora de notificación 1, oprima el botón arrancar/parar para comletar el ajuste. • para ajustar las horas de noti...

  • Page 13

    24 23 troubleshooting display the “e” blinks on the display status before error pressurization to above 280 mmhg was performed. (re-pressurization was performed several times.) pressure decreased suddenly and appeared in the display. Mark only blinks a few times, or not at all. Cuff does not inflate...

  • Page 14

    26 25 specifications method of measurement: display: measuring range: measurement accuracy: power source: operating environment: storage environment: wrist circumference measurement range: dimensions (h#w#d): weight: oscillometric system digital lcd blood pressure: 0 - 280 mmhg pulse rate: 30 - 160 ...