R&S 1331.6539.85 Manual - 4.3

Manual is about: External Preamplifier

Summary of 1331.6539.85

  • Page 1

    R&s ® ha-z24e external preamplifier manual manual 1178.6140.02 ─ 02 (;ÜËx2).

  • Page 2

    This manual describes the following r&s ® ha-z24e model: ● 1331.6539.85 © 2017 rohde & schwarz gmbh & co. Kg mühldorfstr. 15, 81671 münchen, germany phone: +49 89 41 29 - 0 fax: +49 89 41 29 12 164 email: info@rohde-schwarz.Com internet: www.Rohde-schwarz.Com subject to change – data without toleran...

  • Page 3

    Contents r&s ® ha-z24e 3 manual 1178.6140.02 ─ 02 contents 1 about the r&s ha-z24e....................................................... 3 2 for your safety...................................................................... 4 3 product description....................................................

  • Page 4

    For your safety r&s ® ha-z24e 4 manual 1178.6140.02 ─ 02 2 for your safety the r&s ha-z24e is designated for use in industrial, administrative, and labora- tory environments. Use the r&s ha-z24e only for its designated purpose. Observe the safety and usage instructions in this manual and the supplie...

  • Page 5

    Product description r&s ® ha-z24e 5 manual 1178.6140.02 ─ 02 the preamplifier unit consists of the following main parts: figure 3-2: main parts of the preamplifier unit 1 = rf input connector 2 = rf output connector 3 = lemosa 5-pole usb connector and status led for power supply 4 = adjustable feet ...

  • Page 6

    Product description r&s ® ha-z24e 6 manual 1178.6140.02 ─ 02 possible damage to measurement device due to excess power to avoid damage to your measurement device, ensure that it supports the maximum output power of the r&s ha-z24e of +18 dbm. Direct connection cable connection lemosa 5-pole usb conn...

  • Page 7: 4.1

    Retrieving the s-parameters r&s ® ha-z24e 7 manual 1178.6140.02 ─ 02 4 retrieving the s-parameters the r&s ha-z24e stores the s-parameters as a touchstone (.S2p) file on the device. You can read out those values from a connected pc. 4.1 prerequisites to read out values from the r&s ha-z24e, the pc m...

  • Page 8: 4.3

    Retrieving the s-parameters r&s ® ha-z24e 8 manual 1178.6140.02 ─ 02 3. On the pc, start the ha_z24e-1.06.Msi file to install the driver. In case the windows user account does not have sufficient rights: a) search for the command-line tool cmd. B) right-click the tool and select the "run as administ...

  • Page 9

    Measurement setup with r&s ha-z24e r&s ® ha-z24e 9 manual 1178.6140.02 ─ 02 the software uses the installed driver and reads out the touchdown (.S2p) file from the eeprom of the r&s ha-z24e. The file is stored on the pc, in the same directory the software was started from. The name of the file is ha...

  • Page 10

    Measurement setup with r&s ha-z24e r&s ® ha-z24e 10 manual 1178.6140.02 ─ 02 precautions to avoid damage to the rf connectors when setting up a measurement with the r&s ha-z24e, consider the fol- lowing to avoid damage to the rf connectors: ● use the cables provided with the r&s ha-z24e to connect t...

  • Page 11

    Performance test r&s ® ha-z24e 11 manual 1178.6140.02 ─ 02 a) connect one end of the 1-mm cable to the rf out connector of the r&s ha-z24e. B) connect the other end of the 1-mm cable to the rf input connector of the measurement device. 4. Connect the usb power supply connector of the r&s ha-z24e to ...

  • Page 12

    Performance test r&s ® ha-z24e 12 manual 1178.6140.02 ─ 02 evaluation the measured s-parameter s21 should comply with the datasheet specification for small signal gain..