Troubleshooting:
Error Message
Problem
Solution
Device stabilization is in process.
Wait until stops flashing.
Battery is low and no more measurements
are possible.
Replace the battery.
Measurement before device stabilization.
Wait until stops flashing.
The device shows a rapid ambient
temperature change.
Allow the thermometer to rest in a room for at
least 30 minutes at room temperature:
10
o
C and 40
o
C (50
o
F ~104
o
F).
The ambient temperature is not within the
normal temperature range between 10
o
C and
40
o
C (50
o
F ~104
o
F).
Allow the thermometer to rest in a room at least
30 minutes at room temperature: 10
o
C and 40
o
C (50
o
F ~104
o
F).
Error 5~9, the system is not functioning
properly.
Unload the battery, wait for 1 minute and
repower it. If the message reappears, contact
the retailer for service.
Temperature taken is higher than 42.2
o
C
(108.0
o
F).
Check the integrity of the probe cover and take a
new temperature measurement.
Temperature taken is lower than 34
o
C
(93.2
o
F).
Make sure the probe cover is clean and take a
new temperature measurement.
Device can not be powered on to the ready
stage.
Change with a new battery.
*Note: Suggest that the device is performed the re-test for accuracy after 3 years. Please send the complete
device to the dealers or nearest service address. However, if this device is used according to the
operation instruction, periodic re-calibration is not required.
Please read the instructions for use
BF type applied part
Medical Technology Promedt Consulting GmbH
Add: Altenhofstrasse 80, D-66386 St. Ingbert, Germany
Radiant Innovation Inc.
Tel: +886 (3) 5644185 Fax: +886 (3) 5644170
E-mail:service@radiantek.com.tw Http://www.radiantek.com.tw
Add: 1F, No.3, Industrial East 9
th
Road, Science-Based Industrial Park, HsinChu, Taiwan 300.