Radiantek TH859S Manual - page 2
Ref No.:062015
Enclosure Rating: IP22
Dimensions: 144.8 x 48.5 x 39.8 mm
Weight: 68.8 grams including battery
Battery life: around 3,000 continuous readings.
Expected Service Life: 4 years
The device should not submerge into any liquids and expose it to direct moisture.
There is no gender and age limitation for using infrared thermometer.
This is not an AP or APG product.
Troubleshooting:
Error Message
Problem
Solution
Device stabilization is in process.
Wait until stops flashing.
Battery is low and no more measurements
are possible.
Replace the battery.
Measurement before device stabilization.
Wait until stops flashing.
The ambient temperature is not within the
normal temperature range between 10
o
C and
40
o
C (50
o
F ~104
o
F).
Allow the thermometer to rest in a room at least
15 minutes at room temperature: 10
o
C and 40
o
C (50
o
F ~104
o
F).
Error 5~9, the system is not functioning
properly.
Unload the battery, wait for 1 minute and
repower it. If the message reappears, contact
the retailer for service.
Temperature taken is higher than 42.2
o
C
(108.0
o
F).
Check the integrity of the probe cover and take a
new temperature measurement.
Temperature taken is lower than 34
o
C
(93.2
o
F).
Make sure the probe cover is clean and take a
new temperature measurement.
Device can not be powered on to the ready
stage.
Change with a new battery.
Warranty :12 months
Manufacture Date:as the serial number( please open the battery cover, it is shown on the inside of the device.)
Ex.SN:E
5
12A000001, the first “E” is External, the second number “
5
”
is the last number of manufacture year,
the
third and the fourth number “12” is the manufacture month, the others is the serial number.
Note: The thermometer is calibrated at the time of manufacture. If at any time you question the accuracy of
temperature measurements, please contact the dealers or nearest service address.
Warning: No modification of this equipment is allowed.
Please read the instructions for use
BF type applied part
Medical Technology Promedt Consulting GmbH
Add: Altenhofstrasse 80, D-66386 St. Ingbert, Germany
Radiant Innovation Inc.
Tel: +886 (3) 5644185 Fax: +886 (3) 5644170
E-mail:
MKT@radiantek.com.tw Http://www.radiantek.com.tw
Add: 1F, No.3, Industrial East 9
th
Road, Science-Based Industrial Park, HsinChu, Taiwan 300.