Rae RAEGuard 2 PID User Manual

Manual is about: Fixed VOC monitor

Summary of RAEGuard 2 PID

  • Page 1

    User’s guide raeguard 2 pid p/n d03-4001-000 rev d april 2012.

  • Page 2: Contents

    Raeguard 2 pid user’s guide contents section 1: raeguard 2 pid user’s guide 1 general information ............................................................ 2 2 general specifications ......................................................... 4 2.1 raeguard 2 pid specifications .......................

  • Page 3

    Raeguard 2 pid user’s guide 2 set span value ................................................................. 23 meas. Gas* ....................................................................... 23 ext. Alarm delay* ............................................................ 23 pump duty (%)* ......

  • Page 4

    Raeguard 2 pid user’s guide 3 section 2: digipid user’s guide safety instructions .................................................................... 57 17 read before operating ............................................... 58 17.1 digipid marking ....................................................

  • Page 5

    Raeguard 2 pid user’s guide 4 © copyright 2012 rae systems by honeywell.

  • Page 6

    Raeguard 2 pid user’s guide 1 section 1: raeguard 2 pid user’s guide.

  • Page 7: General Information

    Raeguard 2 pid user’s guide 2 1 general information the raeguard 2 pid is a fixed photoionization detector (pid) that measures a broad range of volatile organic compounds (vocs). It operates on 10 to 28 vdc and provides an analog (4-20ma) three- wire signal output, and rs-485 modbus digital signal o...

  • Page 8: Applications:

    Raeguard 2 pid user’s guide 3 applications: • refineries, petrochemical and natural gas plants • metallurgical • chemical, medication • environmental protection • electricity, communications • fire protection • utilities • pulp and paper, printing • storage • sewage disposal • food, brewing • scient...

  • Page 9: General Specifications

    Raeguard 2 pid user’s guide 4 2 general specifications 2.1 raeguard 2 pid specifications basic parameters principle pid (photoionization detector) sensor digital smart sensor sampling internal diaphragm pump working current dc 10 to 28v, 210ma at 24v power output • 4-20ma • three-level programmable ...

  • Page 10

    Raeguard 2 pid user’s guide 5 raeguard 2 pid specifications (continued) physical parameters dimensions, l x w x h 257 x 201 x 107 mm (10.1" x 7.9" x 4.2") material stainless steel weight 3.5 kg (7.7 lbs) certification atex ii 2g ex d ia ii ct4 gb ul/csa class i, div. 1, groups a b c d t4 - pending i...

  • Page 11: Operation

    Raeguard 2 pid user’s guide 6 3 operation the calibration of all newly purchased rae systems instruments should be tested by exposing the sensor to a known concentration calibration gas before the instrument is used or put into service. Prior to factory shipment, the raeguard 2 pid is calibrated and...

  • Page 12: Physical Description

    Raeguard 2 pid user’s guide 7 4 physical description the raeguard 2 pid can be easily installed and integrated with various control systems. It is designed with flexible pipe- holding/wall-mounting options and standard connection terminals. 5 physical dimensions the physical dimensions are as shown:...

  • Page 13: 6.1

    Raeguard 2 pid user’s guide 8 6 installation and access instructions 6.1 mounting first, decide where the transmitter will be mounted. (refer to installation drawing, below.) drill two holes in mounting surface, with the center of the holes 5.25" (133mm) apart. Besides directly mounting the raeguard...

  • Page 14: Note:

    Raeguard 2 pid user’s guide 9 note: when installing the raeguard 2 pid, make sure the sensor is vertically oriented (pointing straight down). Wall pole.

  • Page 15: 6.2

    Raeguard 2 pid user’s guide 10 6.2 instrument disassembly prior to service: make sure power is off. Observe all hazardous location safety procedures. 1. Loosen the fastening bolt before unscrewing the housing lid. Unscrew the housing lid from the housing body by rotating it counter- clockwise. (as s...

  • Page 16: 6.3

    Raeguard 2 pid user’s guide 11 to reassemble the instrument: 1 reconnect the 24-pin connector of the ribbon cable and lock it in place. 2 fasten the cable to the board. 3 mate the board with the clip on the left side and tilt it over. 4 click the board into place. 5 screw on the sensor, making sure ...

  • Page 17: 6.4

    Raeguard 2 pid user’s guide 12 6.4 wiring procedure 1. Inside the housing bottom, unplug the two green terminal block plugs from the terminal block on the pc boards. Note : the terminal block plugs accept 12 awg to 24 awg wire. 2. Lace the wires through the raeguard 2 pid’s wire hole(s) and connect ...

  • Page 18: 6.5

    Raeguard 2 pid user’s guide 13 6.5 earth grounding instructions 6.5.1 external earth grounding fasten the crimped ground wire with hardware as illustrated below. The wire should have a minimum cross-section area of 4mm 2 for its conductor. 6.5.2 internal earth grounding use the same hardware as show...

  • Page 19: 6.6

    Raeguard 2 pid user’s guide 14 6.5.3 finished grounding wires internal and external grounding are shown here, as well as an alternate external grounding point. Always follow local electrical guidelines. 6.6 alarm contact setup the alarm contacts can drive external alarms such as a light or buzzer. B...

  • Page 20: Important!

    Raeguard 2 pid user’s guide 15 three jumper blocks are located below the three green terminal blocks. They are labeled (from left to right): j4, j3, and j1. Here are their functions: jumper function j1 high j3 low j4 error lift off the jumper for each one that you want to change and slide it back on...

  • Page 21: Display And User Interface

    Raeguard 2 pid user’s guide 16 7 display and user interface 7.1 user interface the raeguard 2 pid’s user interface consists of a status led, an lcd display, and three keys, [+], [mode], and [-]. The three keys are operated by using the magnet key. Raeguard 2 pid user interface. 7.2 magnet key the ra...

  • Page 22: 7.3

    Raeguard 2 pid user’s guide 17 7.3 using the magnet key using the magnet end of the magnet key, briefly touch the glass above the mode circle or the triangles labeled [+] and [-]. Then remove the key straight out and away from the raeguard 2 pid. Raeguard magnet key touching glass above the [+] tria...

  • Page 23: 7.6

    Raeguard 2 pid user’s guide 18 condition. If there are no errors or alarm conditions, the green “ok” led is lit and the gas concentration is displayed. If there is an error, the “fault” led blinks and an error message blinks. Each alarm condition has a corresponding led that blinks an amber color wh...

  • Page 24: Navigating Settings

    Raeguard 2 pid user’s guide 19 8 navigating settings you can view the basic settings, clear the peak and minimum readings, and perform a test of the analog system without entering programming mode. At the main reading screen, hold the magnetic key over [-] for two seconds. The raeguard 2 screen chan...

  • Page 25

    Raeguard 2 pid user’s guide 20 • if you do not want to clear the minimum value, press [-]. • to clear the minimum value (reset it), press [+]. • if you do not want to clear the minimum value, press [-]. The screen advances to min. Clear. • to confirm that you want to clear the minimum value (reset i...

  • Page 26

    Raeguard 2 pid user’s guide 21 press [-] to advance to the screen showing the measurement range (m.Range) and high alarm (h.Alarm) and low alarm (l.Alarm) values: press [-] to advance to the screen showing the firmware version and build time and date, and the serial number of the instrument: press [...

  • Page 27: Note:

    Raeguard 2 pid user’s guide 22 for example, if you have a sensor range of 0 to 1000 ppm, the detector lcd displays in sequence 0, 125, 250, 375…all the way up to 1000 ppm. These are the ppm equivalent values of 4,6, 8, 10, etc., all the way up to 20ma. Note: you can stop the test and exit at any tim...

  • Page 28: Programming Menus

    Raeguard 2 pid user’s guide 23 9 programming menus programming allows anyone with the password to change the instrument’s settings, calibrate the instrument, etc. Depending on whether you use the basic or advanced mode, both of which have different passwords, you can access different menus and subme...

  • Page 29: 9.1

    Raeguard 2 pid user’s guide 24 9.1 entering programming mode enter programming mode by using the magnetic key to press these buttons in sequence: 1. Press [+]. 2. Press [-]. 3. Press [mode]. The password screen appears, with the cursor on the first digit: input the 4-digit password: • increase the n...

  • Page 30: 9.2

    Raeguard 2 pid user’s guide 25 as you repeatedly press [-], the selection moves from left to right, and you see these screens in sequence: 9.2 calibration two types of calibration are available: zero (fresh air) and span. Note: complete calibration procedures are detailed on page 39. Press [+] to se...

  • Page 31: Zero/fresh Calibration

    Raeguard 2 pid user’s guide 26 zero/fresh calibration your options are indicated along the bottom of the screen: • press [+] to start. • press [mode] to quit and return to the menu. If you choose to start zero/fresh air calibration, you see this message: please apply zero gas… connect the zero gas, ...

  • Page 32: Gas Library

    Raeguard 2 pid user’s guide 27 • gas library is a library that consists of all the gases found in rae systems’ technical note tn-106 (available online at www.Raesystems.Com). • press [-] to step between the two options. • press [mode] to go back to the menu level above it. • press [+] to select your...

  • Page 33

    Raeguard 2 pid user’s guide 28 1. Press [-] to move the cursor from “ppm” to “mg/m3” or cycle through them. 2. Press [+] to register your choice. 3. Press [mode] to save the selection. 4. Press [+] to save the measurement unit, [mode] to return to editing it, and [-] to cancel the change. The new me...

  • Page 34: Meas. Gas (Advanced Mode)

    Raeguard 2 pid user’s guide 29 if you have set a value that is outside the sensor’s range, you see this, along with a message that says, “invalid input!”: you have two options: • press [-] to exit without making a change. • press [+] to go back to change the value. Meas. Gas (advanced mode) pids can...

  • Page 35: Last Ten

    Raeguard 2 pid user’s guide 30 then follow the steps for the type of selection group you have chosen: last ten 1. Step through the list’s gases by pressing [-]. 2. Register your selection by pressing [+]. 3. Save your choice by pressing [mode]. Gas library 1. Step through the list’s gases by pressin...

  • Page 36: 9.4

    Raeguard 2 pid user’s guide 31 9.4 alarm setting during each measurement period, the gas concentration is compared with the programmed high and low gas concentration alarm limits. If the concentration exceeds any of the preset limits, the alarms (and relays) are activated immediately to warn of the ...

  • Page 37: Note:

    Raeguard 2 pid user’s guide 32 ext. Alarm delay (advanced mode) in order to prevent false alarms, you can set a delay time between when an alarm event occurs and when the raeguard 2 sends out the alarm signal, up to 60 seconds. If the alarm duration is less than the alarm delay setting, then no rela...

  • Page 38: 9.5

    Raeguard 2 pid user’s guide 33 9.5 monitor setup this menu has sub-menus for setting the date, time, and other parameters. Temp. Unit set the temperature units to either fahrenheit or celsius. 1. Press [-] to place the cursor on the temperature unit you want. 2. Press [+] to make the selection. 3. P...

  • Page 39: Note:

    Raeguard 2 pid user’s guide 34 measurement is greater than the 2ppm threshold and is re-enabled when the reading falls below 90% of the threshold (1.8 ppm). 1. Press [+] repeatedly to select a desired number. Numbers increase from 0 to 9. Once 9 is reached, pressing [+] again “wraps” around back to ...

  • Page 40: Note:

    Raeguard 2 pid user’s guide 35 pump cycle(s) (advanced mode) this sets the length of time for one cycle of the pump running. 1. Press [+] repeatedly to select a desired number. Numbers increase from 0 to 9. Once 9 is reached, pressing [+] again “wraps” around back to 0. 2. Press [-] to advance to th...

  • Page 41: Bus Baudrate

    Raeguard 2 pid user’s guide 36 bus baudrate choose from three different baud rates for modbus data transmission: 4800, 9600, or 19200. 1. Press [-] to place the cursor on the baud rate you want. 2. Press [+] to make the selection. 3. Press [mode] to save your selection. 4. Press [+] to store the sel...

  • Page 42: Unit Id

    Raeguard 2 pid user’s guide 37 unit id when using modbus communication, you can give each raeguard 2 pid on a network a unique unit id that helps identify it. 1. Press [+] repeatedly to select a desired number. Numbers increase from 0 to 9. Once 9 is reached, pressing [+] again “wraps” around back t...

  • Page 43: Note:

    Raeguard 2 pid user’s guide 38 change password (advanced mode) you can change the four-digit password for accessing basic mode. The default basic password is: 1111 the advanced password is: 1250 note: you cannot change the password for accessing advanced mode. 1. Press [+] to step through all 10 num...

  • Page 44: Calibration

    Raeguard 2 pid user’s guide 39 10 calibration the raeguard 2 pid is calibrated using a two-point calibration process. First, use a “zero gas.” then use a “span gas” containing a known concentration of a standard reference gas, to set the second point of reference. Note: zero calibration must be perf...

  • Page 45: 10.1  Zero Calibration

    Raeguard 2 pid user’s guide 40 10.1 zero calibration 1. Connect the zero gas cylinder to the gas inlet port on the raeguard2 pid using the provided calibration adapter. 2. In measurement mode, use the magnet key to press [+], [-], and then [mode] in sequence. 3. Input your password, followed by [-]....

  • Page 46: Note:

    Raeguard 2 pid user’s guide 41 7. Press [+]. A countdown screen is shown as the calibration is performed: note: you may abort the calibration during the countdown by pressing [-]. If you stop the calibration, the process stops and this screen is displayed before the menu screen is shown: 8. Once the...

  • Page 47: 10.2  Span Calibration

    Raeguard 2 pid user’s guide 42 10.2 span calibration 1. Connect the calibration gas cylinder to the gas inlet port on the raeguard2 pid using the provided calibration adapter. Note: make sure the labeled concentration of calibration gas matches the value that is set in the raeguard 2 pid. 2. If the ...

  • Page 48: Note:

    Raeguard 2 pid user’s guide 43 6. Press [+], and this screen appears: 7. Start the gas flow. 8. Press [+]. A countdown screen is shown as the calibration is performed: note: you may abort the calibration during the countdown by pressing [-]. If you stop the calibration, the process stops and this sc...

  • Page 49: Alarm Signal Summary

    Raeguard 2 pid user’s guide 44 11 alarm signal summary the following are reading-related alarms. Status lcd led indication relay status analog output normal green failure: disable low alarm: disable high alarm: disable according to reading value low alarm red (2 flashes per second) failure: disable ...

  • Page 50: Maintenance

    Raeguard 2 pid user’s guide 45 12 maintenance as a guide, it is recommended to regularly “bump test” a raeguard 2 pid unit with a known percentage gas. Also, periodically check the sensor inlet and outlet to make sure they are clean and unobstructed. 13 troubleshooting note: before attempting to dia...

  • Page 51: Related Configuration

    Raeguard 2 pid user’s guide 46 14 appendix a: range, sensor and related configuration the following table shows the measurement range and related configurations for four different digipid sensor modules: sensor (ppm) resolution (ppm) span gas setup default value default alarm setup (ppm) low alarm h...

  • Page 52: Answering Message:

    Raeguard 2 pid user’s guide 47 2. Data structure/communication steps raeguard 2 pid only supports function code 0x03 (read-only registers). The detector only supports value reading. 0x03: read-only register requesting message: device address func- tion code register address high byte register addres...

  • Page 53: Raeguard 2 And Digipid

    Raeguard 2 pid user’s guide 48 16 appendix b: controlled section raeguard 2 and digipid 16.1 scope the scope of this document is to describe the system aspects for use of the raeguard 2 pid by connection of the raeguard2 and digipid according to en60079-25; ul913 and can/csa-c22.2 no. 157-92 in haza...

  • Page 54

    Raeguard 2 pid user’s guide 49 cid1 cid1 same -40º c +55º c - um: 28vdc - - uo: 5.91v ui: 6.13v uo ok io: 0.8a ii: 1.5a io ok po: 1.16w pi: 1.225w po ok co: 30µf ci: 20.2µf ci ok lo: 55µh li: 1µh li ok as demonstrated in the table above, raeguard 2 and digipid have matching entity parameters, which ...

  • Page 55: Raeguard 2

    Raeguard 2 pid user’s guide 50 raeguard 2 16.3 scope the scope of this document is to identify the section of the raeguard2 controlled part of manual. 16.4 responsibility the included sections cannot be changed without prior approval from the notified body. 16.5 contents below are the sections contr...

  • Page 56: - Read Before Operating -

    Raeguard 2 pid user’s guide 51 1. Warnings and directive information - read before operating - this manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, mai...

  • Page 57: 2. Marking Of Raeguard 2

    Raeguard 2 pid user’s guide 52 2. Marking of raeguard 2 the raeguard 2 is certified according to atex and the iecex schemes as protected by a flameproof enclosure provided one or two associated equipment output port protected as intrinsically safe [ia], one of which can include a set of flame arrest...

  • Page 58

    Raeguard 2 pid user’s guide 53 5. Instructions for safe use it must be ensured that the supply voltage for raeguard 2 comply with the um input entity parameter. The 8-pin output interface connector facilitates connection to intrinsically safe equipment, e.G. Sensors or modems. The connected devices ...

  • Page 59: 7.  Maintenance

    Raeguard 2 pid user’s guide 54 input/output entity parameters the rated raeguard2 input/output parameters are as follows: input: um: 28vdc output: uo: 5.91v co: 30µf lo: 55uh io: 0.8a po: 1.16w cable entry type and size raeguard2 is provided with three female 3/4" - 14 npt cable entry holes tapped i...

  • Page 60: External Earth Grounding

    Raeguard 2 pid user’s guide 55 8. Earth grounding instructions external earth grounding fasten the crimped ground wire with hardware as illustrated below. The wire should have a minimum cross-section area of 4mm 2 for its conductor. Internal earth grounding use the same hardware as shown in the illu...

  • Page 61: 9.  Physical Dimensions

    Raeguard 2 pid user’s guide 56 9. Physical dimensions raeguard 2 can be easily installed and integrated with various control systems with its flexible pipe-holding/wall-mounting options and standard connection terminals. The physical dimensions are as follows: 130mm 187mm 112mm 118mm.

  • Page 62

    Raeguard 2 pid user’s guide 57 section 2: digipid user’s guide.

  • Page 63: Safety Instructions

    Raeguard 2 pid user’s guide 58 safety instructions 17 read before operating this manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and servic...

  • Page 64: 17.1  Digipid Marking

    Raeguard 2 pid user’s guide 59 17.1 digipid marking the digipid is certified according to the iecex scheme, atex and csa for us and canada as protected by intrinsic safety. The product is marked with the following information: rae systems 3775 n. 1 st . St., san jose ca 95134, usa digi pid type ds10...

  • Page 65

    Raeguard 2 pid user’s guide 60 17.2 operation area and conditions 17.2.1 hazardous areas classified by zones digi pid is intended to be used in hazardous areas or mines susceptible to fredamp classified zone 0, zone 1 or zone 2, within the temperature range of -40º c to +55º c, where gases of explos...

  • Page 66

    Raeguard 2 pid user’s guide 61 17.4 use in hazardous areas equipment which is intended for use in explosive atmospheres and which has been assessed and certified according to international regulations may be used only under specified conditions. The components may not be modified in any way. The app...

  • Page 67: 17.6  Specifications

    Raeguard 2 pid user’s guide 62 17.6 specifications power supply 5v ±0.25v dc current 110 ma max power consumption measuring range 0 to 100 ppm, 1 to 1000 ppm resolution 10 ppb, 100 ppb, 1 ppm (depends on model) response time diffusion (t 90 ): 30 s pumped (t 90 ): 5 s calibration two-point off-line ...

  • Page 68: General Information

    Raeguard 2 pid user’s guide 63 18 general information the digipid sensor is a self-contained intelligent sensor module with built-in photoionization detector (pid) sensor processor, lamp driver, and analog and digital interface circuits. It is designed to detect volatile organic compounds (voc) and ...

  • Page 69: Control Signal:

    Raeguard 2 pid user’s guide 64 control signal: this optional pole only delivers a high (3.3v) or low (0v) signal level. Caution: 6.2v is the maximum input voltage for all of the sensor module’s input poles. 19 grounding (earth connection) digipid includes a grounding connection via pin 6 of the 8-pi...

  • Page 70: Physical Description

    Raeguard 2 pid user’s guide 65 20 physical description the digipid is a pumped sensor. The pump is in this version intended to pump the sample gas in though one of the digipid gas tubes and return it through the other..

  • Page 71

    Raeguard 2 pid user’s guide 66 21 sensor parts and dimensions digipid sensor shown with optional mounting adaptor 8-pin interface connector sample outlet pump connections sensor module pid lamp sensor detector teflon o-ring sensor cover with filter installed optional mounting adapter fastening nut o...

  • Page 72: Note:

    Raeguard 2 pid user’s guide 67 22 operating the sensor module 22.1 preparing the sensor for use make sure the 8-pole input of the external intrinsically safe equipment matches the interface connector pinout configuration of the digipid. Make sure the input power supply range is between 4.75v and 5.2...

  • Page 73: Using The Sensor Module

    Raeguard 2 pid user’s guide 68 23 using the sensor module 23.1 gas flow routing connection the digipid has an “in” and “out” port for sampling gas. Make sure that the in and out sample tubes are properly oriented. After making all gas connections and ensuring connection to the 8-pin connector, turn ...

  • Page 74: Important!

    Raeguard 2 pid user’s guide 69 23.2 sensor module calibration the sensor module is calibrated before it leaves the factory. However, you can also calibrate the sensor module. Allow it to warm up for two hours before calibrating it. Calibration should be performed using external equipment, using a tw...

  • Page 75: Note:

    Raeguard 2 pid user’s guide 70 23.3 maintenance and calibration the sensor module should be calibrated if it does not pass a bump test, but no less than every 6 months, depending on use and exposure to gas and contamination. 23.4 replacing the lamp and filter the sensor module is shipped with the la...

  • Page 76

    Raeguard 2 pid user’s guide 71 housing teflon o-ring pid lamp sensor module attachment ring important! Only handle lamp by holding sides of end. Do not touch flat surface. Filter important! Align hole in filter with hole in top of pid sensor to ensure unobstructed sample flow inlet/outlet.

  • Page 77: Note:

    Raeguard 2 pid user’s guide 72 24 replacing the sensor’s teflon uv shield on the underside of the sensor module is a teflon uv shield that should be replaced every 90 days, in order to ensure sensor accuracy. 1. Dismantle the digipid as shown on page 70, and remove the sensor module. Note: be carefu...

  • Page 78

    Raeguard 2 pid user’s guide 73 4. Unscrew and remove all four of the gold-plated contact “legs” from the sensor module. 5. Use a small flat-bladed screwdriver or similar tool to get between the sensor module and the teflon uv shield at the point nearest the shield’s large “notch,” and gently pry it ...

  • Page 79

    Raeguard 2 pid user’s guide 74 8. Slide a new teflon uv shield over the contact “legs,” and gently press it down until it is snug against the surface of the sensor module. 9. Use your fingers to screw on the four contact “legs.” 10. Use a soft, clean cloth to wipe any finger oils from the surface of...

  • Page 80

    Raeguard 2 pid user’s guide 75 25 sensor & lamp cleaning/replacement clean the pid sensor module, the lamp and housing only if: 1. The reading is inaccurate even after calibration. 2. The reading is very sensitive to air moisture. 3. A liquid has been sucked into the unit and damaged the unit. Use o...

  • Page 81: Changing The Lamp

    Raeguard 2 pid user’s guide 76 25.2 cleaning the lamp housing or changing the lamp if the lamp does not turn on, the instrument will display an error message to indicate replacement of the lamp may be required. 1. If the lamp is operational, clean the lamp window surface and the lamp housing by wipi...

  • Page 82

    Raeguard 2 pid user’s guide 77 25.3 cleaning the instrument occasional cleaning with a soft cloth is recommended. Do not use detergents or chemicals. 25.4 ordering replacement parts if you need replacement parts, contact your local rae systems distributor. A list is available online: http://www.Raes...

  • Page 83: Electronic Waste Disposal

    Raeguard 2 pid user’s guide 78 26 electronic waste disposal at the end of its life, this product must undergo separate collection and recycling from general or household waste..

  • Page 84: Technical Support

    Raeguard 2 pid user’s guide 79 27 technical support to contact rae systems technical support: monday through friday, 7:00am to 5:00pm pacific (us) time phone (toll-free) : +1 888-723-4800 phone : +1 408-952-8461 email : tech@raesystems.Com 28 rae systems contacts rae systems by honeywell world headq...

  • Page 85: Rae Systems Europe Aps

    Raeguard 2 pid user’s guide 80 rae systems europe aps kirstinehøj 23 a dk-2770 kastrup denmark phone: +45 86 52 51 55 fax: +45 86 52 51 77 orders@raeeurope.Com sales@raeeurope.Com service@raeeurope.Com web: www.Raesystems.Eu rae systems uk ltd d5 culham innovation centre culham science centre abingd...

  • Page 86: Rae Benelux Bv

    Raeguard 2 pid user’s guide 81 rae benelux bv rijndal 20 2904 dc capelle a/d ijssel phone: +31 10 4426149 fax: +31 10 4426148 email: info@rae.Nl web: www.Rae.Nl rae systems spain, s.L. Av. Remolar, 31 08820 el prat de llobregat spain phone: +34 933 788 352 fax: +34 933 788 353 mobile: +34 687 491 10...

  • Page 87: Rae Systems Japan

    Raeguard 2 pid user’s guide 82 rae systems japan 403 plaza ochanomizu bldg. 2-1 surugadai kanda chiyoda-ku tokyo, japan phone: 81-3-5283-3268 fax: 81-3-5283-3275 email: jpsales@raesystems.Com rae systems korea #1010, daemyungansvill first, sang-dong 412-2, wonmi-gu, bucheon, kyungki-do, korea phone:...

  • Page 88: Rae Systems By Honeywell

    Rae systems by honeywell world headquarters 3775 n. First st. San jose, ca 95134-1708 usa phone: 408.952.8200 toll-free: 888.723.4800 fax: 408.952.8480 e-mail (sales support): raesales@raesystems.Com web site: www.Raesystems.Com technical support phone: 408.952.8461 (mon-fri, 7am-5pm, pst) email: te...