Rae ToxiRAE II PGM-1100 Series User Manual

Manual is about: Personal Toxic Gas Monitor

Summary of ToxiRAE II PGM-1100 Series

  • Page 1

    Toxirae ii pgm-1100 series personal toxic gas monitor user manual 045-4003-001, rev a december 2012 ex-ox-tox gasdetectie westerdreef 5v 2152 cs nieuw-vennep telefoon: 0252 620885 e-mail: info@exoxtox.Nl website: www.Exoxtox.Nl.

  • Page 2: - Read Before Operating -

    - read before operating - this manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the manufacturer’s instructi...

  • Page 3: Table of Contents

    Table of contents standard contents . . . . . . . . . . . . . . . . . . . . . . 1 calibration kit . . . . . . . . . . . . . . . . . . . . . . . . . . 1 general information . . . . . . . . . . . . . . . . . . . . . 2 physical description . . . . . . . . . . . . . . . . . . . . . 3 display features . ...

  • Page 4

    Proper product disposal at end of life the waste electrical and electronic equipment (weee) directive (2002/96/ec) is intended to promote recycling of electrical and electronic equipment and their components at end of life . This symbol (crossed-out wheeled bin) indicates separate collection of wast...

  • Page 5: Standard Contents

    1 standard contents toxirae ii monitor with sensor as specified high-capacity lithium battery installed gas adapter alligator clip stainless steel belt clip / hardhat adapter operation instructions calibration kit test gas (for sensor as specified) gas regulator with flow controller hard transport c...

  • Page 6: General Information

    2 general information the toxirae ii single-gas monitor continuously displays toxic gas concentrations . The toxirae ii is a full-featured gas monitor providing continuous, digital display of the selected toxic gas concentration, stel, twa and peak values, as well as high, low, twa and stel alarms ..

  • Page 7: Physical Description

    3 physical description 1 bright red led alarm 2 display 3 audio alarm 4 gas inlet 5 button 6 alligator clip 2 3 1 6 4 5.

  • Page 8: Display Features

    4 display features 1 zero calibration 2 low battery indicator 3 short term exposure limit (stel) 4 time weighted average (twa) 5 low alarm 6,7 high alarm 8 parts per million (ppm) 9,10 span calibration 11 gas concentration 2 3 1 6 4 5 7 8 9 10 11.

  • Page 9: Operating The Toxirae II

    5 operating the toxirae ii turning the monitor on make sure the lithium battery is installed . Hold the button down and release when the display says “on .” the monitor is now on . The monitor performs a self-test preceding the warm-up and zero calibration . You will see the firmware version display...

  • Page 10: Time Weighted Average (Twa)

    6 short term exposure limit (stel) the stel is the average reading of the gas concentration for the last 15 minutes . Note: the stel reading does not appear until 15 minutes have elapsed . Time weighted average (twa) the twa is the accumulated reading of the gas concentration since the monitor was t...

  • Page 11: Program Mode

    7 program mode use the program mode to perform any of the following actions (listed in order of appearance): calibration changing preset limits or span gas values entering program mode (firmware version 1.4, sensors for hcn, cl 2, clo 2, no, no 2 , ph 3 , so 2, or nh 3 ):the instrument must first be...

  • Page 12: Calibration

    8 calibration all newly purchased toxirae ii instruments should be bump tested by exposing the sensor to known concentration calibration gas before the instrument is used or put into service . A bump test is defined as a brief exposure of the monitor to the calibration gas and the sensors to show re...

  • Page 13: Only)

    9 you will see the “cal” (calibration) and “go” messages switch back and forth on the screen . To advance or skip to the next submenu, press the button once . Zero calibration (firmware version 1.7 only) note: zero calibration is performed auto– matically at startup with firmware version 1.4. To per...

  • Page 14: Span Gas Calibration

    10 span gas calibration to perform span gas calibration, press and hold the button . Release when “go” displays continuously without flashing on and off. The “gas” and span gas preset value switches back and forth on the screen . During these 10 seconds, connect the span gas to the monitor . (see pi...

  • Page 15: Span Gas Cylinder

    11 span gas cylinder regulator tygon tubing calibration adapter to interrupt and exit span calibration during the countdown, press the button and “no” appears on the screen . Press the button again to confirm and advance to the next submenu ..

  • Page 16: Stel

    12 stel high alarm span twa low alarm changing preset limits or span gas values cycle through the submenus by pressing the button once after each; stop at appropriate screen . (with firmware version 1 .4, the monitor returns to the monitor mode after the span gas value .) you will see “set” and “go”...

  • Page 17: Caution!

    13 caution! Change the battery only in an area known to be non- hazardous. Warning! Use only the following battery model: eve er14335 ! Maintenance follow these instructions carefully and calibrate and test your monitor according to this user manual’s instructions before placing it in service again ...

  • Page 18: Performance of This Product.

    14 step 1. To replace the battery, remove the four screws from the back of the monitor . Step 2. Remove the back cover . Step 3. Gently pull the ribbon to remove the battery from the monitor . Step 4. Install a lithium battery (voltage: 3 .6v, capacity: 1 .65ah, size: 2/3aa, p/n: 500-0076-100) . Imp...

  • Page 19

    15 replacing the filter (all versions) the toxirae ii has a removable sensor cover that allows you to change the sensor filter. Replace the filter (p/n: 045-2054-000) when it appears dirty or when the sensor is replaced . To remove the sensor cap, use the calibration adapter to turn the sensor cap o...

  • Page 20: Replacing The Sensor

    16 replacing the sensor toxirae ii models are designes so that you can easily change the sensor . New sensors need to warm up before first calibration and use. (see table a for required warm-up times.) to replace the sensor, use the calibration adapter to turn the sensor cap one quarter-turn counter...

  • Page 21: Specifications

    17 specifications size 3 .6" (9 .3 cm) x 1 .9" (4 .9 cm) x 0 .9" (2 .2 cm) weight 3 .6 oz (102 gm) with clip battery user-replaceable 2/3 aa high-capacity lithium battery operating period 2 years typical battery life for co/h 2 s/o 2 sensors or 730 minutes of alarm . All other low- or high-bias sens...

  • Page 22: Sensor Configuration

    18 table a: sensor configuration sensors are not interchangeable. Use only rae systems sensors, and use only the sensor type specified for your toxirae ii monitor. Use only rae systems batteries. Use of non-rae systems components will void the warranty and can compromise the safe performance of this...

  • Page 23: Alarm Signals (H

    19 table b: alarm signals (h 2 s) alarm type display cause buzzer & led vibration alarm over range reading > maximum range 3 beeps/sec once every second high reading > high alarm limit 3 beeps/sec once every second low reading > low alarm limit 2 beeps/sec once every second twa twa reading > twa lim...

  • Page 24