RKI Instruments GX-2012 Quick Reference Manual
1. Turning the GX-2012 ON
Note: This start up sequence assumes the following maintenance mode items
are turned on: LNCH BRK, ID DISP, CL RMNDR, and BP RMNDR
Set to CONFIRM
a. Attach the rubber nozzle or sample hose to the GX-2012’s
quick connect inlet fitting.
b. If a hose is used, attach probe to hose.
c. To turn on the GX-2012, press and briefly hold the POWER
ENTER button until you hear a beep. The GX-2012 will begin
its warm up sequence. The following screens may appear
during the warm up sequence:
CAL TIME REMAINING (factory set to appear)
• If the instrument if overdue for calibration, the GX-2012
displays CAL. Press the RESET SILENCE button to
continue.
• Calibration should be performed as soon as possible.
BUMP TIME REMAINING (does not appear with
factory setting)
• If the instrument is overdue for bump testing, the
GX-2012 displays “tESt B--Limit”. Press the RESET
SILENCE button to continue.
• A bump test should be performed as soon as possible.
SENSOR FAILURE
• If the GX-2012 experiences a sensor failure during start
up, the LCD will display FAIL and indicate which sensor(s)
failed. To continue operation press and release RESET
SILENCE to acknowledge the failure. Gas readings for
that sensor will be replaced by “- - -”.
• It is necessary to replace the sensor and recalibrate
before the GX-2012 is used.
2. Measuring Mode Screen
a. After warm up the GX-2012 will display the following: CH4
0%LEL, OXY 20.9%, CO 0ppm, H2S 0.0ppm (some readings
may not appear if sensors are not installed).
b. If the readings are not displaying fresh air values as above,
you must perform a Fresh Air Adjustment.
Quick Reference Guide For Model GX-2012
Note: Turn on and adjust the GX-2012 gas monitor in a known fresh air area.