Sensor Electronics SEC 3000 Instruction And Operation Manual - page 4
3
I. SPECIFICATIONS
Model:
SEC 3000 Gas Detector
Available gases:
Ammonia
Carbon Monoxide
Hydrogen
Nitric Oxide
Oxygen
Phosgene
Bromine
Chlorine
Chlorine Dioxide
Fluorine Hydrogen
Peroxide
Ozone
Hydrogen Chloride
Hydrogen Cyanide
Hydrogen Fluoride
Hydrogen Sulfide
Nitrogen Dioxide
Sulfur Dioxide
Arsine Diborane
Germane
Hydrogen Selenide
Phosphine
Silane
Formaldehyde
Please note that this list is not all-inclusive. The SEC 3000 sensors can be calibrated for other toxic gases
provided a calibration gas is available. For more please contact Sensor Electronics Corporation.
Model Number: SEC3000
Detection Method: Electrochemical or Galvanic
Aspiration:
Diffusion
Optional sample draw (requires 1 liter per minute sample flow rate)
Output (Analog): 4-20 mA (Source type), max. 1000 Ohm load at 24 VDC supply voltage
Output (Digital) Interactive Interface Available On The Calibration (White) Wire
Response Time: Varies for type of sensing element
Construction: 316 Stainless Steel Explosion Proof
Accuracy: +/- 5%
Operating Temperature Rating:
-40
° to +70°C at 0 to 99% RH (non-condensing)
Operating Voltage:
24 VDC Operating range: 18 to 32 VDC measured at the detector head
Power Consumption: 1 Watt Max.
Max. Current Draw: 40 mA (at 24 VDC)
Approvals: CSA: Class 1, Division 1, Groups B,C,D
Installation Category: Cat. I, Pollution Degree 2