SeQual Eclipse 3 User Manual - Troubleshooting Table
Personal Ambulatory Oxygen System
Users Manual
No Battery installed
Install
Battery
or plug into external power
Battery is discharged
or no external power is present
Plug into external power
AC or DC Power Supply
is not plugged in correctly
Check plug at outlet, Power Supply box and
at the Eclipse to ensure properly plugged in.
Insufficient power from the outlet
Ensure there is adequate power from the outlet.
(Outlet may be controlled by a wall switch
or the circuit breaker may have engaged.)
Other
Contact your dealer/provider
Filter Blocked
Clean Air Inlet Filter
Humidifier
Check humidifier attachment
and tubing
Eclipse not ON
Power Eclipse ON
Tubing or cannula is not properly
connected or kinked
Check tubing, cannula
and connections
Other
Contact your dealer/provider
Restriction in tubing or humidifier
Repair or replace tubing or humidifier.
Place your Eclipse so there is adequate air flow.
Filter restricted
Clean or replace air inlet filter. Place your Eclipse
so there is adequate air flow.
Exceeding maximum breath rate in
Pulse Mode. *See Pulse Dose Mode
Breath Rate Chart (p. 18)
Switch to Continuous Flow Mode until respiratory
rate resumes to an acceptable rate.
Other
Contact your dealer/provider
Restriction in humidifier or tubing
Repair or replace humidifier or tubing
Filter Blocked
Clean or replace air inlet filter
Other
Contact your dealer/provider
Tubing/cannula longer
than 7 feet (2.1m)
Attach 7 foot (2.1m) tubing/cannula
Humidifier attached
Remove humidifier
No inspiration detected
Contact your dealer/provider
Unit in Continuous Mode
Switch to Pulse Mode
Nasal cannula not in the
nose or mouth breathing
Place nasal cannula in nostrils
and breath through your nose
Power Cartridge (Battery) is aging
Contact your dealer/provider
to replace Power Cartridge (Battery)
SympTOm
pRObAblE CAUSE
REmEDy
Eclipse does not
power on when
ON/OFF button
is pressed
No Oxygen
Delivered in
Pulse Flow Mode
Power Cartridge Status
Gauge never indicates
fully charged
Low Oxygen
Concentration
Low Oxygen Flow
No Oxygen
TROUblEShOOTING TAblE
DO NOT IGNORE AlARmS.
A BACk-UP SOURCE OF SUPPLEMENTAL OXYGEN IS RECOMMENDED IN CASE OF A POWER OUTAGE
OR A MECHANICAL FAILURE OF THE DEVICE. CONSULT YOUR PROVIDER FOR A BACk-UP SOURCE
OF OXYGEN.
WARNING:
20