SeQual Regalia Instruction Manual

Summary of Regalia

  • Page 1

    Regalia oxygen concentrator regalia instruction manual p/n 5167 rev a march 2006.

  • Page 2

    Sequal technologies inc. Part no: 5167 2 table of contents warnings and cautions .............................................................................................................................. 3 indications for use .........................................................................

  • Page 3

    Sequal technologies inc. Part no: 5167 3 warnings and cautions warning and caution notices used in this manual apply to hazards or unsafe practices that could result in personal injury or property damage. Not for medical purposes. "read instruction manual" do notoperate this oxygen concentrator with...

  • Page 4

    Sequal technologies inc. Part no: 5167 4 indications for use the sequal regalia oxygen concentrator is intended to provide only non-medical oxygen. It is not for medical purposes. This oxygen concentrator is not intended for life supporting or life sustaining applications. Introduction sequal's rega...

  • Page 5

    Sequal technologies inc. Part no: 5167 5 "electrical shock” indicates the presence of an electric shock hazard. Warning:to reduce the risk of electric shock, do not remove cover. There are no user serviceable parts inside. "audible alarm" an audible indicator is provided to announce, in a tone disti...

  • Page 6

    Sequal technologies inc. Part no: 5167 6 recommended operating environment guidelines temperature 50 º f to 104 º f (10 º c to 40 º c) electrical use no extension cords or electrical outlets controlled by a switch. Placement no closer than 3 inches (7.6 cm) from the wall, draperies and furniture. En...

  • Page 7

    Sequal technologies inc. Part no: 5167 7 alarm system - the oxygen concentrator has an alarm system with a red visual alarm indicator light and buzzer. The alarm system will activate in the event of an internal fault or power outage. The red and/or yellow light may stay on for up to ten (10) minutes...

  • Page 8

    Sequal technologies inc. Part no: 5167 8 installation instructions 1. Select a location caution: locate oxygen concentrator in a well-ventilated space that provides adequate airflow and check if air is being prevented from entering the oxygen concentrator by furniture, draperies or clothing. 2. Insp...

  • Page 9

    Sequal technologies inc. Part no: 5167 9 power management feature in the event of a power interruption: warning:because the duration of a power interruption is unknown, seek an alternate supply of oxygen during the power interruption. Resume use of the regalia when power is restored. If a power outa...

  • Page 10

    Sequal technologies inc. Part no: 5167 10 2. Display of operating conditions flow rate shown? Green light yellow light red light audible alarm unit state operating condition no off off off off not energized unit not running / flow indicator display not lit no off off off off energized unit not runni...

  • Page 11

    Sequal technologies inc. Part no: 5167 11 maintenance routine maintenance: routine maintenance consists of cleaning the air inlet filter at regular intervals to keep the unit functioning for years. The user can easily achieve this maintenance. Warning: unplug the oxygen concentrator from the power o...

  • Page 12

    Sequal technologies inc. Part no: 5167 12 trouble shooting guide table 1: concentrator is not running symptom probable cause remedy cannot start unit flow indicator blank back light off power cord not plugged into wall outlet. Plug in power cord. Internal fuse blown service required. Audible alarm b...

  • Page 13

    Sequal technologies inc. Part no: 5167 13 technical data flow rate 0.5 to 10.0 lpm (for the 10 lpm version) 0.5 to 7 lpm (for the 7 lpm version) oxygen concentration 1.0 to 10.0 lpm 91 3% at sea level (for the 10 lpm version) 1.0 to 7.0 lpm 91 3% at sea level (for the 7 lpm version) oxygen concentra...

  • Page 14

    Sequal technologies inc. Part no: 5167 14 electromagnetic compatibility this equipment has been tested and found to comply with the limits for medical devices to the iec 60601-1-2: 1994 electromagnetic compatibility standard. These limits are designed to provide reasonable protection against harmful...