Setra Systems 267 Installation Manual

Summary of 267

  • Page 1

    Installation guide 1-800-257-3872 toll free 1-978-264-0292 fax www.Setra.Com web site model 267 and 267mr differential pressure transducers 159 swanson road, boxborough, ma 01719-1304 tel: 800-257-3872/978-263-1400 ss2039 r ev .G 11/2016.

  • Page 2

    Model 264 for static duct and flow measurement model 265 smaller size 10 psi overpressure model 230 wet/wet differential pressure measurement model 209 rugged low cost gauge pressure measurement from 0-5 to 0-1000 psig setra offers a complete line of hvac products 10.0 model 267 desiccant cover the ...

  • Page 3

    Table of contents 1.0 general information ................................................................................................1 2.0 mechanical installation ........................................................................................2 2.1 media compatibility .....................

  • Page 4

    7.0 model 267 & 267mr performance specifications accuracy rss * (at constant temperature.) ±1.0% fs non-linearity, bfsl ±0.98% fs hysteresis 0.2% fs non-repeatability 0.1% fs *rss of non-linearity, non-repeatability and hysteresis. Thermal effects compensated range °f(°c) +40 to +150 (+5 to +65) zer...

  • Page 5

    Setra systems model 267 and model 267mr 1.0 general information every model 267 and model 267mr (multi-range) has been tested and cali- brated before shipment. Specific performance specifications are listed on page 9 of this guide. The model 267 is single range only. The model 267mr has field select...

  • Page 6

    2.0 mechanical installation 2.1 media compatibility model 267 and 267mr transducers are designed to be used with air or noncon- ducting gases. Use with liquids or corrosive gases will damage the unit. 2.2 environment the operating temperature limits of the 267 and 267mr are as follows: operating tem...

  • Page 7

    The model 267 and 267mr voltage output is a 3-wire circuit, with three ter- minals available for wiring (see diagram 1). The -excitation and -output are commoned on the circuit (see diagram 2). The excitation/output specifications are: excitation output 9 to 30 vac / 11.5 to 42 vdc 0 to 5 vdc 12 to ...

  • Page 8

    *pins 8 and 9 are internally commoned together, user can connect to either pin. 3.3 current output units - 1/2’’ conduit opening, pg9 or pg-13.5 electrical termination the model 267 and 267mr is a two-wire loop-powered 4 to 20ma current output unit (see diagram 5). The current flows into +exc. Termi...