M501/M303 series scale operation manual
REV: Y1, July 2, 2013
M501/M303 series scale operation manual
4. Calibration
Turn on the scale.
Press
key during self checking, display will show
Clear the platform, press
key, display will show the
calibration value, you can change the value by pressing
and
key, then press
key to confirm.
display will show
Place the test weight on the platform. When scale get stable,
Press
key, display will show
Then will start self-test and will come to normal display
5. Precision*10
If you want to see more accurate weight value, press and hold
for 2 seconds, display will show one more decimal place, the last digit will
twink for 5 seconds, then it will go back to normal weighing value
automatically.
-4-
1. INTRODUCTION
The M501/M303 scale is accurate, fast and versatile series of chair
scales with tare, BMI functions, all models with ECTAC(Class III) and
conform 93/42/EC directive, materials used according to RoHS directive.
The display is 25 mm LCD with white LED back light display.
The capacity is up to 250kg.
2. KEY DESCRIPTIONS
Turns the scale power On / Off.
Set the zero point for all subsequent weighing / Move digit left.
Tares the scale / Escape from the menu.
Set to hold mode
Set to BMI mode / Enter into the menu
1. Enter parameter settings
2. Choose menu block or options one by one
3. Increase the digital number
Change unit: kg / lb
-1-
unload
pass
load