User's manual carefully read and fully understand these instructions before using this unit. 2003.2 16004094080
I carefully read and fully understand these instructions before using this unit. T&d corporation accepts no responsibility for any malfunction of and / or trouble with this product or with your computer that is caused by the improper handling of this product and will deem such trouble or malfunction...
Ii safety precautions and instructions please carefully observe the following safety measures when using our product. To prevent any loss or damage to our customers, other people and / or property, and to ensure the proper use of our products we ask that before using our product you carefully read, ...
Iii if water or a foreign body enters the case, immediately cease using it . Continued use may cause fire, electrocution or damage. Store all batteries, sensors and thermo recorder units out of the reach of children. It is dangerous to swallow batteries. Connecting a communication cable, that is alr...
Iv make sure to set up this unit as a base unit, remote unit, or relay unit using the software [wireless for windows] (sold separately) it will not operate without doing this. Neither the base unit nor a relay unit can measure, record, or display data. Only remote units can measure, record, and disp...
V do not use or store the unit in places such as listed below: it may cause electrocution, fire or damage to the unit or to your computer. Lareas exposed to strong magnetic fields. It may cause damage lareas exposed to static electricity. It may cause damage. Lareas exposed to excess vibration. It m...
Vi this sensor cannot be used with rtr-72. ◆ about the tr-3220, the standard temp.&humidity sensor for rtr-72 use the sensor only within the measurable temperature and humidity range. (temperature: 0 to 50℃ , humidity: 10 to 95%rh) please make sure that the external sensor jack is securely inserted ...
Vii handling instruction for tr-3220 temp. And humidity sensor ● the replacement period for the standard sensor is 1 year. Replace it when 1 year has passed after opening the bag. When using the temperature and humidity sensor, the surface can become dirty and may decrease its sensitivity and accura...
Viii table of contents safety precautions and instructions------------------------ ii outline of wireless thermo recorder rtr-71/72 ------ 1 contents of package ------------------------------------------- 4 outline of part names and functions---------------------- 5 installing the battery-----------...
1 outline of wireless thermo recorder rtr-71/72 the system wireless downloading function prossible to gather and manage data from a variety of places without direct wire connnection a revolutionary system that allows you to measure and record temperature in each remote unit and by making use of wire...
2 relay unit: a relay unit measures and records temperature and / or humidity, and following orders from the base unit communicates that data to the base unit. Remote units can be registered into groups for easy management by the base unit. If an electrical outlet is nearby the ac adapter may be use...
3 software [wireless for windows ® ] the software is made up of the following three software programs wireless for windows (set up program) this is the main program and controls registration and settings for all units as well as manages the downloading of data via cable. Wireless for windows (contro...
4 contents of package the following items are included in the product package. Rtr-71 wireless thermo recorder rtr-71 temperature sensors tr-5320 × 2 lithium battery cr123a × 1 user's manual and warranty × 1 rtr-72 wireless thermo recorder rtr-72 temperature and humidity sensor tr-3220 × 1 lithium b...
5 outline of part names and functions outline of part names ① antenna ② sensor connection ③ battery compartment ④ lcd display ⑤ serial communication connection ⑥ ac adapter connection [temperature sensor tr-5320] ① thermistor ② stainless pipe (sus316) ③ cable (teflon resin coated) ④ connector [tempe...
6 ■ main unit lcd display ① recording indicator displayed when recording is in progress. ② antenna indicator displayed when wireless communication is possible. ③ displays strength of radio wave. ④ displayed when communication in progress. ⑤ displayed when either temperature or humidity exceeds upper...
7 lcd display-other displays ◆ units registered as a base unit ◆ units registered as a remote unit ◆ units registered as a relay unit ◆ unregistered units the display will alternate from go – pc – ch.1- ch.2 .
8 installing the battery [install the lithium battery cr123a] ch.1 ch.2 ● always put in a new battery. ● be sure that + and - are in the correct directions. ● if the unit is being used as a base unit or a relay unit, please connect the ac adapter as well as install the battery cr123a to help prevent...
9 ④ remove the battery change plug. Make sure that the ( ) indicator goes off and that the antenna indicator ( ) comes on in the lcd display. < figure 1 > < figure 2 > improper disposal may cause damage to our environment. Please be careful to properly dispose of all used batteri...
10 how to connect sensors ◆ connecting the included sensor (remote unit only) [rtr-71] ● insert the sensor plugs securely into the unit. Ch.1 ch.2 [rtr-72] ● insert the sensor plug securely into the unit. Ch.1 ch.2.
11 product specifications [rtr-71] measurement channels 2 channels (temperature) measurement -60 to 155℃ measurement accuracy ± 0.3℃ (-20 to 80℃ ) ± 0.5℃ (-40 to -20/80 to 110℃ ) ± 1.0℃ (-60 to -40/110 to 155℃ ) standard sensors tr-5320 x 2 (sensor with stainless protection) thermal-constant time in...
12 data back-up low battery power wireless method special short wave transmission range approximately 400m (may vary with conditions) dimensions h92mm x w66mm x d35mm (excludes antenna) weight approximately 120g (including one lithium battery) unit temperature resistance temperature: 0 to 50℃ humidi...
13 optional sensor guide ■ sensor for rtr-71 tr-5106 teflon-shielded sensor possible measurement range: -60 to 155℃ sensor temperature durability: -70 to 180℃ cable length: 0. 6m thermal-constant time in air: approx. 15 sec. In agitated water: approx. 2 sec. Water resistant ability: splash proof (se...
14 tr-5420 stainless protection sensor possible measurement range: -60 to 155℃ sensor temperature durability: -70 to 180℃ cable length: 2m thermal-constant time in air: approx. 12 sec in agitated water: approx. 2 sec. Water resistant ability: splash proof (sensor and cable) materials ① thermistor ② ...
15 ■ sensor for rtr-72 tr-3220 temperature and humidity sensor humidity measurement range: 10 to 95%rh humidity measurement accuracy: 5%rh (at 25℃ and 50%rh) temperature measurement range:0 to 50℃ sensor durability range:-10 to 55℃ cable length: 2m service life: 1year under normal conditions ① tempe...
16 warranty provisions for free repair 1 if the unit does not work properly despite the fact that the customer used it properly and in line with the user's manual, the unit shall be repaired free of charge through the agent which sold the unit. 2 if the customer requests free repair because of troub...