Tanita Body Composition Analyzer BC-418MA Instruction Manual

Manual is about: BODY COMPOSITION ANALYZER

Summary of Body Composition Analyzer BC-418MA

  • Page 1

    Gb bc4187011(1) gb please read this instruction manual carefully and keep it handy for future reference. Body composition analyzer bc-418ma instruction manual 14-2,1-chome,maeno-cho,itabashi-ku tokyo,japan tel:(03)3968-2123 / (03)3968-7048 fax:(03)3967-3766 dresdener strasse 25 d-71065 sindelfingen,...

  • Page 2

    Gb gb warning ● individuals with a pacemaker or other internal medical devices this equipment sends a weak electrical current through the body during measurement. Individuals who have internally implanted medical devices, such as pacemakers, should not use this equipment due to the risk of malfuncti...

  • Page 3

    Gb gb 6 1.Product assembly and components printer paper paper width : 58mm paper roll diameter : maximum 55mm 쎱 product structure please verify that the following components are in the package. Syringe instruction manual (technical notes) bottom cover accessories 7 쎱 rear view of control box column ...

  • Page 4

    R r l l %body fat range : 1%~75% %body fat increments : 0.1% max 200kg e=d= 0.1kg min 2kg t= -200kg operation temperature range 0 / +35˚c bc-418ma p t type gb gb 8 feed key advances the print paper. On/off key turns the power on or off. Clothes weight setting key sets the clothes weight. Even if the...

  • Page 5

    Gb gb 11 쎱 connection and installation 3.Preparation for use for accurate measurement, place the weighing platform on as flat a surface as possible and adjust the level with the adjustable feet so the bubble in the level gauge is in the center of the frame. Caution make sure you place the weighing p...

  • Page 6

    Gb gb 12 13 쎱 setting the date and time 4.Setting date and time athletic step on female male kg cm / ft.In stabilized age % fat p t female male athletic step on time set kg cm / ft.In stabilized age % fat p t athletic step on ~ female male kg cm / ft.In stabilized age % fat p t athletic step on time...

  • Page 7

    Gb gb 14 15 the equipment will start up with these settings next time you use it. This process is utilized to select whether or not you use the target body fat ratio function. (see the printout sample on page 16.) athletic step on female male kg cm / ft.In stabilized age % fat p t p t athletic step ...

  • Page 8

    Gb gb 16 17 b o d y t y p e g e n d e r a g e h e i g h t w e i g h t b m i b m r f a t % f a t m a s s f f m t b w d e s i r a b l e r a n g e f a t % f a t m a s s t a r g e t b f % i s : p r e d i c t e d w e i g h t : p r e d i c t e d f a t m a s s : f a t t o g a i n : s t a n d a r d m a l e ...

  • Page 9

    Gb gb female male athletic step on kg cm / ft.In stabilized age % fat female male p t athletic step on kg cm / ft.In stabilized age % fat p t 19 posterior electrodes anterior electrodes athletic step on ~ female male kg cm / ft.In stabilized age % fat p t 1. Turn on the power. Press the [ ] key. “0....

  • Page 10

    Gb gb 쎱 setting of clothes weight 1. After turning on the unit, press the [ ] key. After a momentary display check, “0.0” will appear on the lcd. If measuring units need to be changed, do so at this time by pressing the [ ]key. An arrow on the lcd will follow the selection of weighing units. 2. Weig...

  • Page 11

    20-39 40-59 60-79 20-39 40-59 60-79 0% 10% 20% 30% 40% gb gb 23 22 쎱 explanation of the print out b o d y t y p e g e n d e r a g e h e i g h t w e i g h t b m i b m r f a t % f a t m a s s f f m t b w d e s i r a b l e r a n g e f a t % f a t m a s s t a r g e t b f % i s : p r e d i c t e d w e i ...

  • Page 12

    Gb gb caution please pay careful attention to avoid injury from the sharp edge. Please turn off the machine before clearing paper jams. 1. Remove the paper dispenser cover. If you lift up the paper dispenser cover from the back of the control box, it can be easily removed. 2. Remove the printer cove...

  • Page 13

    Gb gb 26 if problems occur, please refer to the following instructions before you ask for repairs. 쎱 problem list 8.Troubleshooting problem solution nothing appears on the display after pushing [ ] key • please confirm that the control box is properly connected to the ac adapter and the ac adapter i...

  • Page 14

    Gb ● transmission data will be output after measurements regardless of the status of the personal computer or other device at the receiving end. For that reason, the personal computer or other device to which data will be transmitted must be put in a reception-ready status before measurement. Cautio...

  • Page 15

    Gb gb output (fat monitor mode) format 쎱 transmission data 쎱 control commands remote mode all types of data may be entered by connecting the bf-350 to a pc and switching to remote mode. When in remote mode, the body fat analyzer may be operated remotely by sending control commands from a host. 쎱 err...

  • Page 16

    Gb gb switch on body composition analyzer. 1 choose the units of measurement. 2 first, switch on the body composition analyzer. Check that the unit is in remote mode. Also check that the unit is connected properly to the computer. Set tare weight 4 the tare weight need not be set if not required. It...

  • Page 17

    Gb gb host . / body composition analyzer lf cr 1 g host , . Body composition analyzer lf cr 1 g host , . Body composition analyzer lf cr 0 0 e if commands accepted if settings incorrect start measurement 11 readings will be outputted when measurement is finished. Host , if concluded normally host , ...

  • Page 18: Technical Notes

    Gb 36 reliability of body composition measurements with an 8-electrode bia introduction this device calculates body fat percentage, fat mass, fat free mass, and predicted muscle mass on the basis of data obtained by dual energy x-ray absorptiometry (dxa) using bioelectrical impedance analysis (bia)....

  • Page 19

    Gb gb 39 intra-day changes are thought to occur in unique cycles arising from differences in the rhythm of individual daily routines, occupations, and day-to-day activities. Consequently, to obtain consistent measurement figures it is recommended that measurements be taken under the following condit...

  • Page 20

    Gb gb 2) inter-day changes the diagrams below offer examples of actual measurements made of inter-day changes. A study was done to determine the degree of change in the impedance between the feet during dehydration; the first two days represent a normal daily routine, while in the latter two days a ...

  • Page 21

    Gb gb 43 product specifications output items measurement system measurement frequency impedance measurement weight measurement input items 8 polar bioelectrical impedance analysis 50khz measurement current 90µa electrode material pressure contact stainless steel foot pads, the hand grips are metal-p...