VAC V100 Troubleshooting Manual

Summary of V100

  • Page 1

    Trouble shooting guide for v100 positioner 1) 1.V100 not responding to input signal a) the tubing to the actuator may be incorrect i) plant air supply should be ported to the sport on the side of the positioner. Supply must be greater than 30 psi. Ii) the pneumatic control signal (3 to 15 psi) shoul...

  • Page 2

    D) cam i) if cam nut and locking screw are not correctly inserted the positioner will not work correctly, generally resulting in the cam slipping and a small input change will equal a full stroking of the actuator. Make sure that the locking screw is all the way out before you tighten the cam nut; t...

  • Page 3

    (1) the positioner will not work correctly if the positioner is not engaged properly in the actuator. Check the positioner to actuator coupling to see if the positioner and actuator are mated correctly. Ii) linear (1) slop – can result when the actuator to positioner components are loose or improper...

  • Page 4

    Reversing a positioners action a common question that valve accessories & controls often gets from the field is that a customer has a valve package with our positioner however they wish to reverse the input signal and action of the valve. For example a customer has a valve that is closed at 4ma and o...

  • Page 5

    1. The tubing must be reversed. C+ tube must now go to c- and c- must go to c+. So for double acting actuators you do exactly that, just reverse the tubing, for single acting, c+ would now be plugged and c- would now be piped to the actuator. 2. Remove the front cover and the indicator in order to e...