- DL manuals
- Vacuubrand
- Water Pump
- RP 12C
- Instructions For Use Manual
Vacuubrand RP 12C Instructions For Use Manual - page 11
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999019 / 22/12/2009
page 11 of 29
condensate drain plug
mains switch
motor rating plate
roots pump
inlet: small flange NW 25
oil filler plug
oil drain plug
rating plate
pumping unit
sight glass for
oil level with
marks
min./max.
RP 12C / 15C VARIO
inlet: small flange NW 25
motor rating plate
roots pump
condensate drain plug
frequency converter
oil drain plug
gas ballast valve
rating plate
pumping unit
mains switch
oil filler plug
sight glass for
oil level with
marks
min./max.
RP 22C VARIO with MBS 030
RP 35C VARIO with MBS 050
and frequency converter (both)
(fig. RP 15C VARIO)
frequency converter
RP 12C with MBS 030
RP 15C VARIO with MBS 030
and frequency converter
RP 22C/35C VARIO
(fig. RP 22C VARIO)
Summary of RP 12C
Page 1
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 1 of 29 roots pumping units rp 12c rp 15c vario rp 22c vario rp 35c vari...
Page 2
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 2 of 29 attention! Important notes! Not permitted! Misuse may cause dama...
Page 3: Contents
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 3 of 29 contents safety information! ......................................
Page 4: Safety Information!
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 4 of 29 safety information! Remove all packing material, remove the prod...
Page 5
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 5 of 29 comply with all relevant safety requirements (regulations and gu...
Page 6
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 6 of 29 the pumps are not suitable for pumping substances which may form...
Page 7
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 7 of 29 ☞ if the diaphragm pump is shut down, the roots pump is not shut...
Page 8
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 8 of 29 ☞ check every capacitor regularly by measuring its capacity and ...
Page 9: Technical Data
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 9 of 29 technical data * the pump achieves its ultimate pumping speed an...
Page 10
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 10 of 29 we reserve the right for modifications without prior notice! S ...
Page 11
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 11 of 29 condensate drain plug mains switch motor rating plate roots pum...
Page 12: Use and Operation
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 12 of 29 use and operation installation in a vacuum system ☞ avoid throt...
Page 13
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 13 of 29 ☞ the frequency is reduced at high inlet pressures (especially ...
Page 14: Troubleshooting
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 14 of 29 troubleshooting t l u a f e s u a c e l b i s s o p y d e m e r...
Page 15
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 15 of 29 a service manual with exploded view drawings, spare part lists ...
Page 16: Oil Change
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 16 of 29 before starting maintenance, isolate the pump from the vacuum s...
Page 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 17 of 29 replacing diaphragms and valves all bearings are encapsulated a...
Page 18
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 18 of 29 cleaning and inspecting the pump heads: rp 12c and rp 15c vario...
Page 19
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 19 of 29 ☞ check diaphragm for damage and replace if necessary. ➨ to do ...
Page 20
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 20 of 29 assembling pump heads ➨ use face wrench with torque indicator (...
Page 21
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 21 of 29 ☞ reassemble the pumping unit. Individual performance check of ...
Page 22
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 22 of 29 rp 22c/35c vario (with md 12c) cleaning and inspecting the pump...
Page 23
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 23 of 29 view of the disassembled pump head parts cover plate distributo...
Page 24
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 24 of 29 individual performance check of a pump head: ➃ ➃ ➃ ➃ ➃ ➀ ➀ ➀ ➀ ...
Page 25
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 25 of 29 montage of the connecting hose: if the pump does not achieve th...
Page 26
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 26 of 29 notes on return to the factory repair - return - dkd calibratio...
Page 27
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 27 of 29 health and safety clearance form declaration concerning safety,...
Page 28
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 28 of 29 konformitätserklärung declaration of conformity déclaration de ...
Page 29
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product. Manual-no.: 999019 / 22/12/2009 page 29 of 29 disclaimer: our technical literature is only intended to inform...