User's guide vaisala carbocap® carbon dioxide calibrator for the gmp220 series probes gm k220 u421en-1.1.
Page 2
Published by vaisala oyj phone (int.): +358 9 8949 1 p.O. Box 26 fax: +358 9 8949 2227 fin-00421 helsinki finland visit our internet pages at http://www.Vaisala.Com/ © vaisala 2006 no part of this manual may be reproduced in any form or by any means, electronic or mechanical (including photocopying)...
Page 3: Table of Contents
______________________________________________________________________________________ vaisala _____________________________________________________________________________1 table of contents chapter 1 general information...................................................................... 2 safety...
Page 4: General Information
User's guide __________________________________________________________________________ 2 __________________________________________________________________________ u421en-1.1 chapter 1 general information safety throughout the manual, important safety considerations are highlighted as follows: warn...
Page 5: Product Description
Chapter 2______________________________________________________________product description vaisala _____________________________________________________________________________3 chapter 2 product description introduction although vaisala's co 2 instruments are inherently stable, periodical checkings...
Page 6: Probes
User's guide __________________________________________________________________________ 4 __________________________________________________________________________ u421en-1.1 gmk220 calibrator for the gmp220 series probes figure 1 gmk220 calibrator. Caution when handling the gmk220 calibrator, plea...
Page 7: Additional Equipment
Chapter 2______________________________________________________________product description vaisala _____________________________________________________________________________5 additional equipment the following additional equipment is needed for checking and calibration operations: • traceable ref...
Page 8: Operation
User's guide __________________________________________________________________________ 6 __________________________________________________________________________ u421en-1.1 chapter 3 operation spot checking spot checking is carried out when there is a need to check if the probe (gmp221 or gmp222)...
Page 9
Chapter 3_____________________________________________________________________ operation vaisala _____________________________________________________________________________7 6. Let the calibrator settle for 10 minutes. 7. Connect the reference gas and adjust the flow rate to 0.6 l/min with a screw...
Page 10: Note
User's guide __________________________________________________________________________ 8 __________________________________________________________________________ u421en-1.1 11. Stop the output with the esc-button (or by typing mf _ mode 0). 12. The difference between the average reading and the a...
Page 11: Two-Point Calibration
Chapter 3_____________________________________________________________________ operation vaisala _____________________________________________________________________________9 two-point calibration the reference gases used for the two-point calibration shall represent the low and high end of the pro...
Page 12: Note
User's guide __________________________________________________________________________ 10 _________________________________________________________________________ u421en-1.1 9. Give the input data for the calibration. Calicali l 2 concentration > ambient > ambient > where co 2 concentration is giv...
Page 13: Note
Chapter 3_____________________________________________________________________ operation vaisala ____________________________________________________________________________11 12. Give the input data for the calibration. Calicali h 2 concentration > ambient > ambient > where co 2 concentration is gi...
Page 14
User's guide __________________________________________________________________________ 12 _________________________________________________________________________ u421en-1.1 probe_save cr > >probe_save new calibration parameters are now saved in the probe. 14. Disconnect the gas flow and remove th...
Page 15: Technical Data
Chapter 4_________________________________________________________________ technical data vaisala ____________________________________________________________________________13 chapter 4 technical data operating temperature range +15...30ºc dimensions 20 x 18 x 16 (height) cm materials metal frame p...