Vaisala HUMICAP HMP155 User Manual

Summary of HUMICAP HMP155

  • Page 1

    User's guide vaisala humicap® humidity and temperature probe hmp155 m210912en-c.

  • Page 2

    Published by vaisala oyj phone (int.): +358 9 8949 1 p.O. Box 26 fax: +358 9 8949 2227 fi-00421 helsinki finland visit our internet pages at www.Vaisala.Com . © vaisala 2012 no part of this manual may be reproduced, published or publicly displayed in any form or by any means, electronic or mechanica...

  • Page 3: Table of Contents

    ________________________________________________________________________________ vaisala ________________________________________________________________________ 1 table of contents chapter 1 general information ............................................................................ 5 about thi...

  • Page 4

    User's guide____________________________________________________________________ 2 ________________________________________________________________ m210912en-c send [addr] ................................................................30 sdelay .........................................................

  • Page 5

    ________________________________________________________________________________ vaisala ________________________________________________________________________ 3 serial line calibration ........................................................... 63 two-point humidity adjustment ......................

  • Page 6: List of Figures

    User's guide____________________________________________________________________ 4 ________________________________________________________________ m210912en-c list of figures figure 1 hmp155 probe .........................................................................13 figure 2 hmp155 with addit...

  • Page 7: General Information

    Chapter 1 ________________________________________________________ general information vaisala ________________________________________________________________________ 5 chapter 1 general information this chapter provides general notes for the manual and the product. About this manual this manual pr...

  • Page 8: Version Information

    User's guide____________________________________________________________________ 6 ________________________________________________________________ m210912en-c version information table 1 manual revisions manual code description m210912en-c november 2012. This manual. Added the membrane filter optio...

  • Page 9: Safety

    Chapter 1 ________________________________________________________ general information vaisala ________________________________________________________________________ 7 safety vaisala humicap ® humidity and temperature probe hmp155 delivered to you has been tested for safety and approved as shipped...

  • Page 10: Recycling

    User's guide____________________________________________________________________ 8 ________________________________________________________________ m210912en-c recycling recycle all applicable material. Dispose of batteries and the unit according to statutory regulations. Do not dispose of with regu...

  • Page 11: Warranty

    Chapter 1 ________________________________________________________ general information vaisala ________________________________________________________________________ 9 warranty visit our internet pages for standard warranty terms and conditions: www.Vaisala.Com/warranty . Please observe that any s...

  • Page 12

    User's guide____________________________________________________________________ 10 _______________________________________________________________ m210912en-c this page intentionally left blank..

  • Page 13: Product Overview

    Chapter 2 __________________________________________________________ product overview vaisala _______________________________________________________________________ 11 chapter 2 product overview this chapter introduces the features, advantages, and the product nomenclature of the vaisala humicap ® ...

  • Page 14: Basic Features and Options

    User's guide____________________________________________________________________ 12 _______________________________________________________________ m210912en-c table 4 quantities calculated by hmp155 quantity abbreviation metric unit non-metric unit dew point / frost point temperature (t d/f ) tdf º...

  • Page 15: Structure of The Hmp155

    Chapter 2 __________________________________________________________ product overview vaisala _______________________________________________________________________ 13 structure of the hmp155 0801-070 figure 1 hmp155 probe the numbers refer to figure 1 above. 1 = filter 2 = protective cover 3 = 8-p...

  • Page 16

    User's guide____________________________________________________________________ 14 _______________________________________________________________ m210912en-c additional temperature probe option 1206-038 figure 2 hmp155 with additional t-probe the active output version of hmp155 can be ordered with...

  • Page 17: Warmed Probe Option

    Chapter 2 __________________________________________________________ product overview vaisala _______________________________________________________________________ 15 warmed probe option in changing temperature conditions with high humidity, the probe temperature can lag behind the temperature of ...

  • Page 18

    User's guide____________________________________________________________________ 16 _______________________________________________________________ m210912en-c this page intentionally left blank..

  • Page 19: Installation

    Chapter 3 _______________________________________________________________ installation vaisala _______________________________________________________________________ 17 chapter 3 installation this chapter provides you with information on how to install hmp155 in different situations. Installation i...

  • Page 20: Installation In Dtr503

    User's guide____________________________________________________________________ 18 _______________________________________________________________ m210912en-c installation in dtr503 0801-072 figure 5 hmp155 installed in dtr503.

  • Page 21

    Chapter 3 _______________________________________________________________ installation vaisala _______________________________________________________________________ 19 installation in stevenson screen 0805-008 figure 6 hmp155 with t-probe installed in stevenson screen.

  • Page 22: 8-Pin Connector

    User's guide____________________________________________________________________ 20 _______________________________________________________________ m210912en-c 8-pin connector both the passive and active versions of hmp155 come with an 8-pin m12 male connector on the bottom of the probe. 0507-044 fi...

  • Page 23

    Chapter 3 _______________________________________________________________ installation vaisala _______________________________________________________________________ 21 the left side of figure 8 below shows the hmp155d block diagram. The right side shows an example of connections. For the pin layou...

  • Page 24

    User's guide____________________________________________________________________ 22 _______________________________________________________________ m210912en-c temporary rs-485 connection option it is possible to have a temporary rs-485 connection also in the passive output version by powering up th...

  • Page 25: Operation

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 23 chapter 4 operation this chapter contains information that is needed to operate this product. Getting started after the vaisala humi...

  • Page 26: Serial Line Communication

    User's guide____________________________________________________________________ 24 _______________________________________________________________ m210912en-c serial line communication connect the probe to a pc, for example, by using a usb cable (optional accessory item: 221040). Connect the usb ca...

  • Page 27

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 25 after power-up the probe (in stop-mode) outputs the software version and the command prompt. Hmp155 1.00 > in run mode a measurement...

  • Page 28

    User's guide____________________________________________________________________ 26 _______________________________________________________________ m210912en-c terminal application settings the steps below describe how to connect to the hmp155 using the putty terminal application for windows (availa...

  • Page 29

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 27 0903-025 figure 10 putty terminal application.

  • Page 30: List of Serial Commands

    User's guide____________________________________________________________________ 28 _______________________________________________________________ m210912en-c list of serial commands most of the commands listed below are relevant only in the digital output version of hmp155. These commands are mark...

  • Page 31

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 29 table 9 calibration and adjustment commands command description # force connection settings 19200 baud, n, 8, 1 acal [0/1] analog ou...

  • Page 32: Measurement Commands

    User's guide____________________________________________________________________ 30 _______________________________________________________________ m210912en-c measurement commands r enter the r command to start the continuous output of measurements. This command starts a temporary run mode. To chan...

  • Page 33

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 31 where 0 = no extra delay added 1 = 4-8 ms extra delay 255 = 1024 ms extra delay example: >sdelay 15 serial delay : 15 > seri [baud][...

  • Page 34: Smode

    User's guide____________________________________________________________________ 32 _______________________________________________________________ m210912en-c the settings are temporary, and last only for a single session at a time. On the next connection hmp155 reverts to the settings last set wit...

  • Page 35: Formatting Commands

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 33 formatting commands form use the serial line command form to change the format or select certain quantities for the output commands ...

  • Page 36

    User's guide____________________________________________________________________ 34 _______________________________________________________________ m210912en-c ok >send temperature= 24.23 > >form "twet=" 6.3 tw u3 #t "t=" t u3 #r#n ok >send twet= 11.290'c t= 24.231'c > >form 5.1 rh #t t #t tdf #r#n ...

  • Page 37: Unit [M/n]

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 35 unit [m/n] use the unit command to select metric or non-metric output units. Unit [m/n] where m= metric units n = non-metric units e...

  • Page 38: Pres and Xpres

    User's guide____________________________________________________________________ 36 _______________________________________________________________ m210912en-c pressure compensation commands pres and xpres use the pres serial line command to set ambient pressure value for fixed pressure compensation...

  • Page 39: System Commands

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 37 system commands the commands introduced in this section are used for setting the filtering level, checking the probe configuration, ...

  • Page 40

    User's guide____________________________________________________________________ 38 _______________________________________________________________ m210912en-c ? Use the ? Serial line command to check the current probe configuration. ?? Command is similar but can also be used if the probe is in poll...

  • Page 41: Help

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 39 help use the help command to list the available commands. Example: >help ? Acal ch0/ch1 addr 0...99 aerr err1 err2 amode ch1 ch2 ase...

  • Page 42: Errs

    User's guide____________________________________________________________________ 40 _______________________________________________________________ m210912en-c errs use the errs command to display probe error messages, see table 15 on page 54. Example: >errs no errors > vers use the vers command to ...

  • Page 43: Note

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 41 analog output configuration commands the analog output(s) are set in the factory according to the order form. In case you want to ch...

  • Page 44: Asel

    User's guide____________________________________________________________________ 42 _______________________________________________________________ m210912en-c asel use the asel command to select the quantities and scaling for analog outputs of the hmp155. Note that the optional quantities can be se...

  • Page 45: Atest

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 43 atest use the serial line to test the operation of all the analog outputs. Use the atest command to force the analog outputs to ente...

  • Page 46: Note

    User's guide____________________________________________________________________ 44 _______________________________________________________________ m210912en-c note the error output value is outputted only when there are minor electrical faults such as humidity sensor damage. When there is a severe ...

  • Page 47: Sensor Functions

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 45 sensor functions the active output version of hmp155 includes optional functions such as chemical purge and sensor heating. These fu...

  • Page 48: Chemical Purge In Power Up

    User's guide____________________________________________________________________ 46 _______________________________________________________________ m210912en-c 0508-035 figure 11 decrease of sensor gain automatic chemical purge (interval purge) when hmp155 leaves the factory the automatic chemical p...

  • Page 49: Pur

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 47 starting and configuring chemical purge in this section, the function and use of the chemical purge option is explained. Pur with pu...

  • Page 50: Sensor Heating

    User's guide____________________________________________________________________ 48 _______________________________________________________________ m210912en-c sensor heating this function is optionally available only in probes with humicap ® 180rc or humicap ® 180c composite sensors. It should be u...

  • Page 51: Xheat

    Chapter 4 ________________________________________________________________ operation vaisala _______________________________________________________________________ 49 xheat enables or disables the sensor heating. Example: >xheat on extra heat : on >xheat off extra heat : off > to configure the sens...

  • Page 52

    User's guide____________________________________________________________________ 50 _______________________________________________________________ m210912en-c this page intentionally left blank..

  • Page 53: Maintenance

    Chapter 5 ______________________________________________________________ maintenance vaisala _______________________________________________________________________ 51 chapter 5 maintenance this chapter contains information that is needed in basic maintenance of the product. Periodic maintenance cle...

  • Page 54: Changing The Sensor

    User's guide____________________________________________________________________ 52 _______________________________________________________________ m210912en-c changing the sensor you can change the humicap ® sensor yourself and the probe does not have to be sent in for service for a sensor change. ...

  • Page 55

    Chapter 5 ______________________________________________________________ maintenance vaisala _______________________________________________________________________ 53 0802-159 figure 12 changing the humicap ® 180/180r sensors the following numbers refer to figure 12 above: 1 = filter 2 = o-ring 3 =...

  • Page 56: Error States

    User's guide____________________________________________________________________ 54 _______________________________________________________________ m210912en-c error states in an error state, the quantity is not measured and the output is shown as follows: - analog channel outputs 0 v (you can use t...

  • Page 57: Technical Support

    Chapter 5 ______________________________________________________________ maintenance vaisala _______________________________________________________________________ 55 technical support for technical questions, contact the vaisala technical support by e-mail at helpdesk@vaisala.Com . Provide at leas...

  • Page 58

    User's guide____________________________________________________________________ 56 _______________________________________________________________ m210912en-c this page intentionally left blank..

  • Page 59: Calibration and Adjustment

    Chapter 6 ___________________________________________________calibration and adjustment vaisala _______________________________________________________________________ 57 chapter 6 calibration and adjustment the vaisala humicap ® humidity and temperature probe hmp155 is fully calibrated and adjusted...

  • Page 60: Note

    User's guide____________________________________________________________________ 58 _______________________________________________________________ m210912en-c 0801-076 figure 13 hmp155 adjustment buttons the following numbers refer to figure 13 above: 1 = protective cover (unscrewed) 2 = down butto...

  • Page 61: Push-Button Calibration

    Chapter 6 ___________________________________________________calibration and adjustment vaisala _______________________________________________________________________ 59 push-button calibration calibration and adjustment of the hmp155 can be done by using the adjustment buttons found on the probe. ...

  • Page 62: Note

    User's guide____________________________________________________________________ 60 _______________________________________________________________ m210912en-c 7. Insert the probe into the high end reference chamber (for example, nacl: 75 % rh chamber in the humidity calibrator hmk15) and do the hig...

  • Page 63: Adjustment

    Chapter 6 ___________________________________________________calibration and adjustment vaisala _______________________________________________________________________ 61 one-point humidity and temperature adjustment to make a one-point humidity or temperature adjustment: 1. Carry out the chemical p...

  • Page 64: Calibration

    User's guide____________________________________________________________________ 62 _______________________________________________________________ m210912en-c passive output version push-button calibration if you are operating a passive output version hmp155, the humidity calibration is carried out...

  • Page 65: Serial Line Calibration

    Chapter 6 ___________________________________________________calibration and adjustment vaisala _______________________________________________________________________ 63 serial line calibration two-point humidity adjustment note that the difference between the two humidity references must be at lea...

  • Page 66: Note

    User's guide____________________________________________________________________ 64 _______________________________________________________________ m210912en-c 13. Enter the adjustment information (date and text) to the memory of the probe; see the commands ctext and cdate. 14. Reset the probe with ...

  • Page 67: Sensor Change

    Chapter 6 ___________________________________________________calibration and adjustment vaisala _______________________________________________________________________ 65 relative humidity adjustment after sensor change after sensor change, carry out the procedure as described in previous sections. ...

  • Page 68: User Adjustment Commands

    User's guide____________________________________________________________________ 66 _______________________________________________________________ m210912en-c user adjustment commands you can use the user adjustment commands to display and set parameters. You can only use li command in adjust mode....

  • Page 69: Ctext

    Chapter 6 ___________________________________________________calibration and adjustment vaisala _______________________________________________________________________ 67 commands in adjust mode. To put hmp155 in adjustment mode, just hold down the adj button until the green led lights up. Ctext use...

  • Page 70: Acal [0/1]

    User's guide____________________________________________________________________ 68 _______________________________________________________________ m210912en-c acal [0/1] example : ch 0: >acal 0 u1 ( v ) ? 1.001 u2 ( v ) ? 9.011 ch 1: >acal 1 u1 ( v ) ? 2.0988 u2 ( v ) ? 8.8997 > mi70 check and adju...

  • Page 71

    Chapter 6 ___________________________________________________calibration and adjustment vaisala _______________________________________________________________________ 69 4. Press ok to move on to the adjustment and select the quantity you want to adjust. The list of quantities displayed in the figu...

  • Page 72

    User's guide____________________________________________________________________ 70 _______________________________________________________________ m210912en-c this page intentionally left blank..

  • Page 73: Technical Data

    Chapter 7 ____________________________________________________________ technical data vaisala _______________________________________________________________________ 71 chapter 7 technical data this chapter provides the technical data of the product. Performance relative humidity table 16 relative h...

  • Page 74: Temperature

    User's guide____________________________________________________________________ 72 _______________________________________________________________ m210912en-c temperature table 17 temperature performance description value measurement range -80 ... +60 °c (-112 ... +140 ºf) accuracy with voltage out...

  • Page 75: Operating Environment

    Chapter 7 ____________________________________________________________ technical data vaisala _______________________________________________________________________ 73 operating environment table 18 operating environment description value operating temperature range for humidity measurement -80 ......

  • Page 76: Calculated Variables

    User's guide____________________________________________________________________ 74 _______________________________________________________________ m210912en-c calculated variables table 19 calculated variables quantity abbreviation metric unit non metric unit dew point / frost point temperature (t ...

  • Page 77: Inputs and Outputs

    Chapter 7 ____________________________________________________________ technical data vaisala _______________________________________________________________________ 75 accuracy of wet bulb temperature °c relative humidity temp. 10 20 30 40 50 60 70 80 90 100 -40 0.20 0.20 0.20 0.20 0.20 0.20 0.20 0...

  • Page 78: Mechanics

    User's guide____________________________________________________________________ 76 _______________________________________________________________ m210912en-c table 21 average current consumption (+15 vdc, load 100 kohm) description value 0…1 v output 0…10 v output + 0.5 ma rs-485 during chemical p...

  • Page 79: Options and Accessories

    Chapter 7 ____________________________________________________________ technical data vaisala _______________________________________________________________________ 77 options and accessories table 24 options and accessories description item code sensors humicap180r humicap180r humicap180 humicap18...

  • Page 80: Dimensions In Mm (Inches)

    User's guide____________________________________________________________________ 78 _______________________________________________________________ m210912en-c dimensions in mm (inches) hmp155 probe Ø12 [0.47] 20 [0.79] 24 [0.94] 86 [3.39] 279 266 [10.47] 40 [1.57] [10.98] 0801-077 figure 16 probe d...

  • Page 81: Calculation Formulas

    Appendix a ______________________________________________________ calculation formulas vaisala _______________________________________________________________________ 79 appendix a calculation formulas this appendix contains the formulas used for the calculated output quantities. The hmp155 series p...

  • Page 82

    User's guide____________________________________________________________________ 80 _______________________________________________________________ m210912en-c enthalpy: x x t h 5 . 2 ) 00189 . 0 01 . 1 ( (4) the water vapor saturation pressure at temperature t pws is calculated over water and over ...

  • Page 83

    Appendix a ______________________________________________________ calculation formulas vaisala _______________________________________________________________________ 81 where: b0 = 0.13914993 * 101 b1 = -0.48640239 * 10-1 b2 = 0.41764768 * 10-4 b3 = -0.14452093 * 10-7 b4 = 0.65459673 * 101 a i = co...

  • Page 84