Vaisala PTB210 SERIES User Manual
Summary of PTB210 SERIES
Page 1
User's guide digital barometers with serial output ptb210 series m210669en-d.
Page 2
Published by vaisala oyj phone (int.): +358 9 8949 1 p.O. Box 26 fax: +358 9 8949 2227 fi-00421 helsinki finland visit our internet pages at http://www.Vaisala.Com/ © vaisala 2009 no part of this manual may be reproduced in any form or by any means, electronic or mechanical (including photocopying),...
Page 3: Table of Contents
________________________________________________________________________________ vaisala ________________________________________________________________________ 1 table of contents chapter 1 general information............................................................................ 3 safety.......
Page 4: List of Figures
User's guide ______________________________________________________________________ 2 ___________________________________________________________________ m210669en-d chapter 5 calibration and adjustment...........................................................21 wizard software........................
Page 5: General Information
Chapter 1 ________________________________________________________ general information vaisala ________________________________________________________________________ 3 chapter 1 general information safety throughout the manual, important safety considerations are highlighted as follows: warning wa...
Page 6: Recycling
User's guide ______________________________________________________________________ 4 ___________________________________________________________________ m210669en-d recycling recycle all applicable material. Dispose of batteries and the unit according to statutory regulations. Do not dispose of wit...
Page 7: Warranty
Chapter 1 ________________________________________________________ general information vaisala ________________________________________________________________________ 5 warranty vaisala hereby represents and warrants all products manufactured by vaisala and sold hereunder to be free from defects in...
Page 8
User's guide ______________________________________________________________________ 6 ___________________________________________________________________ m210669en-d this page intentionally left blank..
Page 9: Product Description
Chapter 2 _________________________________________________________product description vaisala ________________________________________________________________________ 7 chapter 2 product description the ptb210 series digital barometers are designed for various pressure ranges. The barometers with s...
Page 10
User's guide ______________________________________________________________________ 8 ___________________________________________________________________ m210669en-d the ptb210 barometers incorporate the barocap ® silicon capacitive absolute pressure sensor developed by vaisala. The sensor features ...
Page 11: Installation
Chapter 3 _______________________________________________________________ installation vaisala ________________________________________________________________________ 9 chapter 3 installation mounting due to the compact design of the ptb210, all you need to do is to mount the instrument, and to con...
Page 12: Note
User's guide ______________________________________________________________________ 10 __________________________________________________________________ m210669en-d note check that the cable is not taken upwards from the barometer. This is to prevent water flowing along the cable, and gathering clo...
Page 13: Connections
Chapter 3 _______________________________________________________________ installation vaisala _______________________________________________________________________ 11 connections there are three different serial output interfaces available with the ptb210 digital barometers: rs232c, rs485/232c, a...
Page 14: Rs232C/ttl
User's guide ______________________________________________________________________ 12 __________________________________________________________________ m210669en-d rs232c/ttl table 3 connections of the ptb210 barometer with rs232c/ttl interface wire color signal grey rx (with rs232c) green tx (wit...
Page 15: Operating
Chapter 4 ________________________________________________________________ operating vaisala _______________________________________________________________________ 13 chapter 4 operating general the ptb210 series barometers can be operated through a serial line by using suitable terminal software l...
Page 16: Settings
User's guide ______________________________________________________________________ 14 __________________________________________________________________ m210669en-d settings .Baud setting the baud rate .Baud.Xxxx xxxx = baud rate (1200 ... 19200) example: .Baud.9600 (commands invisible) .Reset note...
Page 17: Note
Chapter 4 ________________________________________________________________ operating vaisala _______________________________________________________________________ 15 .Id setting a identification code .Id.Xxx xxx = identification code (max. 15 characters) if there are several barometers on same lin...
Page 18: .Mpm Measurements Per Minute
User's guide ______________________________________________________________________ 16 __________________________________________________________________ m210669en-d .Mpm measurements per minute .Mpm.Xxx xxx = number of measurements per minute (6 ... 4200) the command sets the number of measurements...
Page 19: Operating Commands
Chapter 4 ________________________________________________________________ operating vaisala _______________________________________________________________________ 17 operating commands .P outputting one single message .P the command outputs the latest multi-point corrected pressure reading with tw...
Page 20: Note
User's guide ______________________________________________________________________ 18 __________________________________________________________________ m210669en-d output with or without the unit .Form.0 the command cancels the outputting of the unit after pressure reading. This is also the factor...
Page 21: Resistor On/off
Chapter 4 ________________________________________________________________ operating vaisala _______________________________________________________________________ 19 .Cald setting the calibration date with this command, the last calibration date can be stored in the memory of the barometer. .Cald....
Page 22: Self-Testing
User's guide ______________________________________________________________________ 20 __________________________________________________________________ m210669en-d .Ron the . Ron command sets the terminating resistor on. .Roff the . Roff command sets the terminating resistor off. Self-testing .? B...
Page 23: Calibration and Adjustment
Chapter 5 ___________________________________________________calibration and adjustment vaisala _______________________________________________________________________ 21 chapter 5 calibration and adjustment the ptb210 series digital barometers can be fine adjusted and calibrated against pressure st...
Page 24: Wizard Software
User's guide ______________________________________________________________________ 22 __________________________________________________________________ m210669en-d wizard software the user-friendly wizard software is designed to work in win3.X, win95, and nt environment. The multi-point pressure c...
Page 25
Chapter 5 ___________________________________________________calibration and adjustment vaisala _______________________________________________________________________ 23 1. Choose the com port in use from the file menu. 2. To open the connection click the read ptb210-button. 3. Give reference press...
Page 26: Note
User's guide ______________________________________________________________________ 24 __________________________________________________________________ m210669en-d note the values are given in hpa: the reference value as whole number, and the correction with two decimals. 4. To save the new correc...
Page 27: Technical Data
Chapter 6 ____________________________________________________________ technical data vaisala _______________________________________________________________________ 25 chapter 6 technical data technical specification operating range pressure range 500 ... 1100 hpa 50 ... 1100 hpa operating temperat...
Page 28: General
User's guide ______________________________________________________________________ 26 __________________________________________________________________ m210669en-d * defined as the ±2 standard deviation limits of end-point non-linearity, hysteresis error or repeatability error. ** defined as ±2 st...
Page 29
Chapter 6 ____________________________________________________________ technical data vaisala _______________________________________________________________________ 27 electromagnetic compatibility electromagnetic compatability en61326-1:1997+ am1:1998 + am2:2001 generic environment [cispr16/22 cla...
Page 30: Dimensions In Mm (Inches)
User's guide ______________________________________________________________________ 28 __________________________________________________________________ m210669en-d dimensions in mm (inches) 12 2. 5 ( 4. 82 ) 12 0 ( 4 .7 2) 16 (0 .6 3) 50 (1.97) 32 ( 1. 26 ).
Page 31
Chapter 6 ____________________________________________________________ technical data vaisala _______________________________________________________________________ 29 40 (1.57) 37 (1.46) 2 (0.08) ø15 (0.59) 21 (0.83) 50 (1.97) 14x45° (0.55x45°) m3 m3 60 (2.36) 1 0 (0 .3 9) 14.20 (0.56) 19 (0.75) 5...
Page 32