Valhalla 2300 Operation Manual - page 54
Sample Program #3 (Current Query):
10 OUTPUT
715;"W1" requests
φB data
20 OUTPUT
715;"T1" requests current data
30 ENTER
715;A$
controller reads buffer
40 PRINT
A$
display
the
data
The result is .7952E2,_ which is equal to
.7952 x 10
2
or 79.52 amps.
Sample Program #4 (Status-Word Query):
10 OUTPUT
715;"E"
requests status word
20 ENTER
715;A$
controller reads buffer
30 PRINT
A$
display
the
data
The result is Q0V1I8W1D0T1 which was
explained in Section 7-7-8.
7-8.
IEEE-488 Device-Dependent Command Set Summary for 2300/01
Listed below is a summary of the device-dependent command set for the 2300 Series of Power
Analyzers. Refer to the preceding paragraphs for greater details on use of the various commands. The
settings labeled "default" are the conditions that exist whenever the 2300 is initially powered up. The
default conditions are also restored upon returning the 2300 to Local mode. The commands are listed in
alphabetical order.
Command
Result
Dn
Output
data
terminator
command
D0
Carriage return and line feed (default)
D1
Carriage return and line feed with EOI
D2
Carriage
return
only
D3
Carriage
return
with
EOI
D4
EOI
only
D5
Line
feed
only
E
Programming
status
query
2300 responds with a configuration status word the next time it is read.
F
Finish
compiling
command
Stops compilation of average watts or watt-hour data.
In
Current
range
commands
I0
0.2 amp range (default)
I1
0.5
amp
range
I2
1
amp
range
I3
2
amp
range
I4
5
amp
range
I5
10
amp
range
I6
20
amp
range
I7
50
amp
range
I8
100
amp
range