ValueStore US Portable Ultrasound V12 User Manual

Manual is about: Veterinary Ultrasound Scanner

Summary of Portable Ultrasound V12

  • Page 1

    1 www.Valuestore.Us customercare@valuestore.Us user manual portable ultrasound v12 veterinary ultrasound scanner.

  • Page 2: Table Of Contents

    2 www.Valuestore.Us customercare@valuestore.Us table of contents statement .................................................................................................................................................. 4 m anufacturer’s r esponsibility ...............................................

  • Page 3

    3 www.Valuestore.Us customercare@valuestore.Us 3.8.6 how to set the color of screen background? ............................................................................... 34 3.9 f unctions m enu .......................................................................................................

  • Page 4: Statement

    4 www.Valuestore.Us customercare@valuestore.Us statement the dodson group co. Ltd. (hereinafter called the dodson) owns the intellectual property rights to this the dodson product and this manual. This manual may refer to information protected by copyright of the dodson. The dodson maintains the con...

  • Page 5: Packing List

    5 www.Valuestore.Us customercare@valuestore.Us packing list s/no description quantity check list remarks 1 main unit 1 √ 2 multi-frequency mechanical sector probe 1 √ 3 rubber protector for main unit 1 √ 4 ac-adapter with cable 1 √ 5 rechargeable battery 1 √ 6 neck belts 1 √ 7 user manual 1 √ please...

  • Page 6: Warranty

    6 www.Valuestore.Us customercare@valuestore.Us warranty this warranty is exclusive and is in lieu of all other warranties. Expressed or implied, including warranties of merchantability or fitness for any particular purpose. Definitions • main unit: integrated facility that implements the specified f...

  • Page 7: Return Policy

    7 www.Valuestore.Us customercare@valuestore.Us return policy return procedure in case that it becomes necessary to return this product or part of this product, the following steps should be followed: 1. Return authorization: contact customer care at customercare@valuestore.Us and please provide the ...

  • Page 8: Important Information

    8 www.Valuestore.Us customercare@valuestore.Us important information 1. The responsibility for maintenance and management of the product after delivery resides with the customer who has purchased the product. 2. The warranty does not cover the following items, even during the warranty period: (1) da...

  • Page 9: Preface

    9 www.Valuestore.Us customercare@valuestore.Us preface to ensure safe and correct operation of the system, read carefully and understand this operator’s manual before operating the system 1.1 product information product name: veterinary ultrasound scanner model: v12 symbols in operator’s manual the ...

  • Page 10

    10 www.Valuestore.Us customercare@valuestore.Us video output usb port packing and transportation labels explanation: handle with care temperature limit upwards limited layers of stowage protect against wetness protect against heat.

  • Page 11: Safety Precautions

    11 www.Valuestore.Us customercare@valuestore.Us safety precautions the operators are asked to observe the following safety precautions. Danger : do not use flammable gasses such as anesthetic gas, oxygen or hydrogen, or flammable liquid such as ethanol, near the system, because there is danger of ex...

  • Page 12

    12 www.Valuestore.Us customercare@valuestore.Us 12. After disinfecting or sterilizing accessories, thoroughly clear chemicals from the accessories. Residual chemicals may damage the accessories and cause injury. 13. Do not let the patient contact the system or other equipments. If the system or othe...

  • Page 13

    13 note : 1. Do not use the system in the vicinity of strong electromagnetic field (for example a transformer), which may affect the performance of the monitor. 2. Using devices transmitting rf signals in the vicinity of the system may affect the system’s performance. Do not use or take any devices ...

  • Page 14: 1 System Basics

    14 1 system basics this ultrasound scanner is suitable for diagnosis on pigs, dogs, cats, sheep and other animals. 1.1 structure 1 2 3 front-side of scanner 9 8 8 7 6 5 4 left-side of scanner.

  • Page 15: 1.2 Main Features

    15 10 11 back side of scanner no. Part description 1 keyboard program the scanner 2 tft color lcd 5.5 inch screen 3 rubber protector around the scanner against dirt and impact from the outside 4 power switch 5 power indicator it will be lighten when the machine is turn on. 6 dc12.6 connection of ac-...

  • Page 16

    16 1.3 technical specifications model ultrasound scanner v12 probe 3.5mhz mechanical sector probe (multi-frequency) detect depth (mm) ≥140 resolution lateral ≤4 (depth≤80) ≤5 (80<depth≤130) (mm) axial ≤2 (depth≤80) blind zone (mm) ≤8 geometric position precision horizontal ≤20 vertical ≤10 monitor s...

  • Page 17: 1.4 Principle Diagram

    17 1.4 principle diagram electric principal block diagram.

  • Page 18: 2  System Connection

    18 2 system connection this chapter introduces how to operate & install system and the battery. 2.1 environmental requirements 1. Environment temperature range:+5℃~+40℃ 2. Relative humidity range:30%~80% 3. Atmosphere pressure range:700hpa~1060hpa when using, avoid strenuous vibration, keep it away ...

  • Page 19

    19 probe connection sketch warning : avoid by all means unplugging or plugging the probe when the main unit is turned on, it can cause permanent damage to the main unit and probe. Once the probe is connected to the main unit, do not unplug nor plug it unless the main unit is turned off. Warning : do...

  • Page 20

    20 2.3.2 battery installation & disassembly battery installation: set the battery into the battery slot and move the battery release key on its back to top till the battery is inserted completely and then release the key. 1. Push the battery into the rabbet like the arrowhead with little force atten...

  • Page 21: 2.4 Power Supply

    21 2.4 power supply there are two modes of the power supply for the device, by the adapter and the built-in battery. 2.4.1 supply via adapter 1. Check the input power cord plug of the adapter to see if it matches the eps outlet. 2. Check the eps to see if it is in the specified range and the power c...

  • Page 22

    22 2.5.2 charging via adapter 1. Take out the battery from the main unit or take out the spare battery. 2. Connect the flat end marked with an arrow of the auto-charger to the charging terminal of the battery. 3. Plug the other end of the auto-charger into the cigar lighter socket. 4. When the "char...

  • Page 23

    23 warning : if you find any abnormality, turn off the system and see if the section lists any solution for the problem. If not, contact customer care or your seller. O check whether the startup is normal. O check whether the system has any abnormal noise or smell or is overheated. O check by hand w...

  • Page 24: 3 Operating & Setting

    24 n 3 operating & setting this chapter describes how to use the keyboard function and preset parameters. You can preset system operating status, configuration parameters of each exam mode, post process and others. The system runs automatically with the preset parameters after each startup. Before e...

  • Page 25: 3.3 Operating Procedure

    25 7 annotations ● hosp: ---- hospital name. ● name: ---- patient name ● age: --- patient’s age. ● sex: --- patient’s sex by f and m defined. (see article 3.4.5 annotation menu) 8 current date current date of the scanner (the date displayed on the frozen image is the date when the image is frozen) 9...

  • Page 26: 3.6 General Setting

    26 4 ► 3 ▲ 3 ▲ 5 ◄ 6 ▼ 6 ▼ 3.6 general setting the parameters at the top right corner of the screen will be changed while you do the setting. Prob: sec-18 freq: 3.5mhz gain: 31 imo ☆ : 36 ▎ 30 ic0/cc0/bc0 3.6.1 how to set probe’s working frequency? ● press 4 ► and 5 ◄ to light “freq” as showing abov...

  • Page 27

    27 freeze menu key to ge 8. Default tset freez menu to gen v1.20 0. Name 1. Age 2. Sex 3. Comment 4. Time 5. Hosp 6. Language 7. Erase 8. Default tset 9. Beep-off 3.6.4 how to back to default setting? 1. Name 2. Age 3. Sex v1.20 4. Comment 5. Time 6. Hosp 7. Language 8. Erase 9. Default tset 10. Bee...

  • Page 28

    28 of screen. --- press clear to erase the wrong letter(s) input previously. --- press menu to confirm the characters chosen and the needed ones will be showing at the top screen in name item. ● age input --- choose 1. Age by pressing key /1 key. There is sub-menu shown at the bottom of screen. Plea...

  • Page 29: 3.7 Keyboard

    29 ● current time --- select 4. Time by pressing the key ►/4, the following sub-menu appears on the screen: yy-mm-dd ■ hh-mm-ss --- 2 bytes allowed in each item. Input “13” to indicate the year of 2017. --- 24 hours setting for the current time. --- the cursor will be switched to the next line for r...

  • Page 30

    30 keyboard of v12 freeze key press it to freeze/unfreeze the echo image. If freeze is righted at the lower corner of the screen, the image is at frozen status. M 0 ~ bm/9 alphanumeric keys all of them are dual functional keys. Function keys are effective when the system is in real-time. The descrip...

  • Page 31

    31 ● m 0 under menu key (when the scanner is frozen) --- press freeze first, then press menu; the main menu will be shown on the screen. --- reference article 3.6.5 for more details. ● /1 image magnification (available in real-time) --- press it to have the image magnified at difference depth when t...

  • Page 32: 3.8  Parameters Setting

    32 6. Equine: gsd 7. Bovine: bl 8. Bovine: sl 9. Bvine: hl 10. Swine: hl 11. Sheep: usd 1. Cat: hd 2. Cat: bd 3. Dog:gsd 4. Dog:crl 5. Dog:hd 6. Dog:bd --- select the number you need to choose the right item for the measurement. Reference article 2.4 for more details about the procedure. --- press c...

  • Page 33

    33 3.8.1 how to set probe frequency? --- press 5◄ or ►4 to light freq first. --- there are three frequencies 2.5/3.5/5.0 for your choice. Press ▲3 and ▼6 to make the selection. 3.8.2 how to set gain? --- press 5◄ or ►4 to light freq first. --- press ▲3 and ▼6 to make the selection. --- the adjustmen...

  • Page 34: 3.9 Functions Menu

    34 3.8.5 how to set the character ’s color? --- --- press key m / 0 to generate the function menu appeared on the screen. 0: save v1.2 1: svload 2: imageproc 3: color 4: area-volm 5: histogarm --- press key▲3 to light cc0 first. Cc is the short for character color. --- there are 2 difference color d...

  • Page 35

    35 3.9.1 image save ---- press freeze key first to lighten freeze on the bottom right screen. --- then press m / 0 key to generate function menu on the screen. --- select “0.Save’ by pressing the key of / 1. There is sentence showing on the left top of screen “saving …. 5”. This sentence will be dis...

  • Page 36

    36 3.9.3 image procession --- this function is not available when the system is not in real-time. --- press m / 0 to have function menu displayed on the screen. --- select “2. Image proc” by pressing the key of + 2. --- repeat + 2 key to select the 2 nd presetting, and the third pressing + 2 key to ...

  • Page 37: 3.10  Measurements

    37 histogram 3.9.6 ---- press freeze key first to lighten freeze on the bottom right screen. --- then press m / 0 key to generate function menu on the screen. --- select “5.Histogram ’ by pressing the key of 5◄ . There is a square frame appeared on the screen . ---- move this square frame by press ▲...

  • Page 38

    38 measure measure m / 0 --- the ratio between these four distances will be shown on its items when the measurements are done. --- press clear key to clear the measuring marks and lines on the screen. The echo image will be kept on there. ●mouse operation: --- press left button and the caliper displ...

  • Page 39

    39 ▲3 ►4 measure * the measure result(s) will be shown automatically at the left side of screen. ● by ellipse: --- press + / 2 to select the way by ellipse: * a caliper will be shown on the screen. * move this ellipse to the focus with the arrow keys ▲3, ►4, 5 ◄, ▼6. * press measure , the arrow keys...

  • Page 40

    40 by ellipse method illustration of volume measuring(3 axis method) procedures:( kidney as example) 1. Catch the cross and longitudinal sections of the kidney and freeze. 2. Measure the perimeter and area of cross and longitudinal sections. 3. The system will automatically complete the measuring, w...

  • Page 41: 3.11 Histogram

    41 measure / 1 or + / 2 clear freeze press reference key to gener reference 3.11 histogram --- freeze the image first. --- press m/0 to have function menu displayed on the screen. --- select “5.Histogram” by pressing the key of ◄/5, a sampling window displayed: --- press the arrow keys to move this ...

  • Page 42

    42 1 st page: 0. equine:gsd 1. bovine:bl 2. bovine:sl 3. bovine:hl 4. swine:hl 5. sheep:usd 2 nd page: 0. cat:hd 1. cat:bd 2. dog:gsd 3. dog:crl 4. dog:hd 5. dog:bd --- input the number and select the related ob menu and acquire the distance as per distance measurement method. The corresponding ga r...

  • Page 43

    43 6. Confirm the distance value as per distance measurement methods and the corresponding data display behind “g·a”. With this measurement to set up a chart to find the growth curves to estimate embryo size and ga. Here ga refers to the duration from the copulation instead of impregnation. ● bovine...

  • Page 44

    44 arly. The measurement method is giv 1. Keep the cow standing. 2. Put the probe against the abdomen side center, shift it a little bit left or right and hold it closely against the skin. Clean the abdomen skin if there is mud to ensure a clear display of the abdominal pelvic structure. 3. The maxi...

  • Page 45

    45 1. Keep the cow standing. 2. Put the probe, a little bit left or right of the center, on the ventral abdominal wall closely along the side of teats and skull to rear leg. If the is mud on this part, clean with water first incase the abdomen pelvic structure could not be displayed accurately. Swin...

  • Page 46

    46 abdomen check: 1. Abdomen examination can be done when the sheep is standing or lying or seating. Put the probe against the appointed abdomen center where there is no fur. 2. Clean the abdomen skin if there is mud to ensure a clear display of the abdominal pelvic structure. 3. Measure the length ...

  • Page 47: 3.13 Power Off

    47 note at ob measurement, when the distance is less than the following value, the ga of this animal will not display. Refer to the following table for detailed data: equine d1 bovine–bl d1 bovine–sl d1 bovine–hl d1 sheep d1 swine d1 cat–hd d1 cat–bd d1 dog–gsd d1 dog–crl d1 dog–hd d1 dog–bd d1 3.13...

  • Page 48: 4 Checking & Maintenance

    48 4 checking & maintenance 4.1 service life bases on the manufacturer’s design, production related files, this model’s use life is six years. The product’s material will gradually aging, if the product continually use over the designed use life, it may bring the problem of the performance reduced a...

  • Page 49: 4.4 Probe Maintenance

    49 4.4 probe maintenance ► probe is an expensive part and frangible. Never hit it or drop it on floor. ► please use medical ultrasound gel during diagnoses. Do daily inspection on the probe enclosure to see if it is cracked and avoid liquid leakage to spoil the built-in components ► check the probe ...

  • Page 50: 4.7 Battery Information

    50 must not use extender, ethylene oxide or any other organic solvent, which tend to deface the probe's protective foil. Keep device or probe from any type of liquid's infiltration. Must not clean device or probe by airing or heating. 4.6 correct usage of probe in order to prolong probe's service li...

  • Page 51

    51 16. Must not bead the battery terminals. 17. Must not decompound the battery in any way. 18. Must not place the battery in microwave oven or pressure vessels. 19. Must not combine the battery with primary battery (such as dry battery) or battery with different capacity, models and types. 20. Do n...

  • Page 52: Table

    52 appendix b acoustic output reporting table index mi tis tis tis tib tib tic mode scanning non- scanning non- scanning scanning non- scanning aaprt≤1 cm 2 aaprt> 1cm 2 acoustic working frequency (mhz) 3.69 3.69 3.60 - 3.69 3.60 output power (mw) 0.03 - 0.03 - bounded output power (mw) 24.7 24.7 at...

  • Page 53

    53 index mi tis tis tis tib tib tic mode scanning non- scanning non- scanning scanning non- scanning aaprt≤1 cm 2 aaprt> 1cm 2 acoustic working frequency (mhz) 4.11 4.11 - 4.02 4.11 4.02 output power (mw) - 0.02 0.02 - bounded output power (mw) 23.7 23.7 attenuated output power (mw) 0.009 0.01 spati...

  • Page 54: Appendix C Obstetrics

    Appendix c obstetrics gestational table 1: equine all measurements +/- 3 days measurement (mm) (gestational sac diameter) week day 6 1 4 8 1 4 10 1 5 12 1 6 14 1 6 16 2 0 18 2 0 20 2 1 22 2 2 24 2 3 26 2 5 28 4 1 30 4 2 32 4 3 34 4 4 40 5 0 42 5 2 44 5 3 46 5 4 48 5 5 50 5 6 52 6 1 54 6 2 56 6 3.

  • Page 55

    Gestational table 2: bovine all measurements +/- 3 days measurement (mm) (body length) week day 8 4 0 10 5 0 12 5 1 14 5 2 16 5 3 18 5 5 20 5 5 22 5 6 24 5 6 26 6 1 28 6 1 30 6 1 32 6 2 34 6 3 36 6 3.

  • Page 56

    Gestational table 3: sheep all measurements +/- 3 days measurement (mm) (umbilicus to spine distance) week day 15 7 1 18 7 3 21 7 6 24 8 1 27 8 4 30 9 0 33 9 2 36 9 4 39 10 0 42 10 2 45 10 5 48 11 3 51 11 5 54 12 1 57 12 2 60 12 4 63 12 6 66 13 2 69 13 4 72 14 2 75 14 4 78 15 0 81 15 2 84 15 5 87 16...