Van Os Medical B.V. Travelux Venture User Manual - User Manual
~ 31 ~
User Manual
© 2016 Excel, part of Van Os Medical UK Ltd. | Excel House | Ashbrooke Park |
Sherburn in Elmet| T: 01977 681 400 | E: sales@vanosmedical.com |
W: www.vanosmedical.nl/en
8
PROBLEM ANALYSIS AND SOLUTIONS
Your Travelux Venture power chair is inspected and adjusted so that you can use it
immediately.
Self-help guide
If a system fault occurs you can find the problem and solution below by counting the
number of LEDs on the battery gauge that are flashing.
Diagnostics
Below is a list of fault codes. Try to use this list before you contact your Travelux
dealer. Find the number of flashes below which matches the number of flashing LEDs
and follow the instructions.
Number of flashes
1 flash
The battery needs charging or there is a bad connection to the battery.
Remedy
Check the connections to the battery. If they are connected the correctly try to load
the batteries.
2 flashes
The left hand motor has a bad connection.
Remedy
Check the connections.
3 flashes
The left hand motor has a short circuit to a battery connection.
Remedy
Contact your Travelux dealer.
4 flashes
The right hand motor has a bad connection.
Remedy
Check the connections.
5
flashes
The right hand motor has a short circuit to a battery connection.
Remedy
Contact your Travelux dealer.
6
flashes
The power chair is stopped driving by an external signal.
Remedy
A possibility is that the battery charger is still connected. Disconnect the charger.