User manual CP-4900 Micro-GC Power Supply
Page: 3
Pre-installation requirements
In order to assure a quick, safe and uncomplicated installation, we kindly
request you to make provisions as stated below before our Varian, Inc. service
engineer install(s) your instrument(s). See also CP-4900 Micro-GC pre-
installation requirements document CP501289.
Environmental requirements
-
Pollution degree: 2
-
Humidity: 5% to 95% RH
-
Temperature: 5 ° to 40 °C operating, -20 ° to 65 °C non-operating
-
The CP-4900 Micro-GC Power Supply should be protected from
corrosive chemicals or gases, dust/particulate accumulation, and direct
venting of air conditioners, heaters, furnaces or fans.
Space requirements
-
Allow sufficient bench space to permit installation of workstations,
integrators and other Micro-GC equipment. The table below lists the
physical dimensions and weight of the CP-4900 Micro-GC Power
Supply and the peripheral instruments which may be installed near it.
-
Allow 10-20 cm of space at the sides and rear of the CP-4900 Micro-
GC Power Supply to permit free air circulation.
Height Width Depth Weight
Instrument
Inch. Cm Inch. Cm Inch. Cm Lb. Kg
CP-4900 Micro-GC 2-CH
11
28
6.5
16
12
30
14
6
CP-4900 Micro-GC 4-CH
11
28
6.5
16
21.5
55
22
10
Power Supply
2.5 6.4
4
9.5
7
17.8
2
1
Field case 2-CH
15
38
12
30
16
41
35
16
Field case 4-CH
(with trolley)
18.5 47 15 38 28.5 73 68 30
Chromatography Workstation
(computer with monitor, approximate
values)
17
43
17
43
21
53
35
16