Varian Turbo-V 969-9831 Instruction Manual - Turbo-V Vent Device

Manual is about: Vent Device

Summary of Turbo-V 969-9831

  • Page 1

    87-900-806-01 (f) may 2005 instruction manual turbo-v vent device model 969-9831 vacuum technologies.

  • Page 2: Turbo-V Vent Device

    Turbo-v vent device.

  • Page 5

    Instruction for use ................................................................................................... 1 technical information .............................................................................................. 2 general.......................................................

  • Page 6

    Instructions for use 87-900-806-01 (f) 1 general information operators and service personnel must be aware of all hazards associated with this equipment. They must know how to recognize hazardous and potentially hazardous conditions, and know how to avoid them. The consequences of unskilled, imprope...

  • Page 7

    Technical information 87-900-806-01 (f) 2 general the 969-9831 turbo-v vent device, consisting of a vent control unit and a vent valve, is a complete unit suitable for automatic venting of the turbo-v pump when it is switched off or during a power failure switch (refer to the following figure). The ...

  • Page 8

    Technical information 87-900-806-01 (f) 3 vent device specifications control unit input: voltage frequency power fuse 120 v ± 10% 50 to 60 hz 8 va t315 (ma) output: voltage power (max) 22.5 vdc ± 3% 1.2 w delay time factory set to 15 seconds (adjustable up to 36 minutes) venting time factory set to ...

  • Page 9

    Technical information 87-900-806-01 (f) 4 control unit installation place the control unit free standing near the turbo-v controller and connect the 1.5 metres long supply cable to the power interconnecting connector (120 vac vent valve out, see the relevant turbo-v controller instruction manual). C...

  • Page 10

    Technical information 87-900-806-01 (f) 5 time setting ! High voltage present in the control unit can cause severe injury or death. Service must be carried out only by qualified and authorized personnel. Disconnect the vent controller from the power interconnecting connector on the turbo-v controlle...

  • Page 11

    Technical information 87-900-806-01 (f) 6 vent control unit description the control unit has three main sections: power circuit, protection circuit, and time circuit (refer to following figure). Power circuit the mains voltage is stepped down to 12 vac, which is rectified by the full wave bridge rec...

  • Page 12

    Technical information 87-900-806-01 (f) 7 turbo-v vent device replacement parts description part number vent valve (complete) 03-661718-02 gas filter 28-900008-01 battery 6 v 1ah 74-103100-01 relay 6v 72-114110-01 transformer t1 (120/12v, 8va) 03-660295 l.E.D. (yellow) lt4223 67-560005 l.E.D. (red) ...

  • Page 13

    Request for return 1. A return authorization number (ra#) will not be issued until this request for return is completely filled out, signed and returned to varian customer service. 2. Return shipments shall be made in compliance with local and international shipping regulations (iata, dot, un). 3. T...

  • Page 14

    Request for return failure report turbo pumps and turbocontrollers position parameters does not start noise vertical power: rotational speed: does not spin freely vibrations horizontal current: inlet pressure: does not reach full speed leak upside-down temp 1: foreline pressure: mechanical contact o...

  • Page 15

    Sales and service offices argentina varian argentina ltd. Sucursal argentina av. Ricardo balbin 2316 1428 buenos aires argentina tel: (54) 1 783 5306 fax: (54) 1 786 5172 australia varian australia pty ltd. 679-701 springvale road mulgrave, victoria zz 3170 australia tel: (61) 395607133 fax: (61) 39...