Verity Medical NeuroTrac Operator's Manual - Quick Start Instructions
NeuroTrac™ TENS Operation Manual
7
1.
Insert a 9 volt PP3 Alkaline battery. Alternatively insert a rechargeable
Nickel Hydride battery [Which has a much longer life than the Ni-Cad
rechargeable batteries] into the battery compartment.
2.
Insert lead wire/s to channel A and B if both channels are to be used.
3.
Switch on the unit by pressing the ON/OFF button
4.
Press the PRG [Programme] button to select one of the programmes as
detailed in table 1 and table 2 on page 8.
5.
To start press channel A + and B + button if you are using both
channels.
6.
To stop the programme, press the ON/OFF button which will turn the
unit off.
Setting up your own Constant mode parameters.
1.
Select PC1 or PC2 by pressing the PRG button on the front panel
Remove the battery lid where you will see two buttons SET and ESC.
Press the SET button and the Hz symbol will flash IN/OFF, then press
the + or – button on the front panel to adjust the Pulse Rate from 2-
200 Hz
2.
Press the SET button again and the µS symbol will flash on/off, then
press the + or – button to adjust the Pulse Duration from 50 - 300µS
3.
Press the SET button again and the Clock symbol will flash ON/OFF,
then press the + or – button to adjust the time. Channel A + or – button
to hours and Channel B + or – button to change minutes.
After setting up the customised programme parameters, press the ESC button
to store the information. Simply repeating the above procedure can
reprogramme customised programmes.