WACO 2609C Operating Instructions Manual

Manual is about: DIGITAL CLAMP METER

Summary of 2609C

  • Page 1

    Operating instructions model 2609c digital clamp meter cat.Iv 600v cat.Iii 1000v 2000a clamp meter true rms off Ω hz v v hz hz hz % 1500a 660a 660a apo ac dc 0 20 10 30 40 50 60 n max range hold zero min 1500a max temp max 30v k-type cat iv 600v cat iii 1000v 750v 1000v 1832°f 1000°c temp Ω 2000a 15...

  • Page 2

    Contents introduction ---------------------------------------------------------- 1 unpacking and inspection ---------------------------------- - - 1 safety precautions ----------------------------------------------- 1 safety information ---------------------------------------------- - - 3 instrument...

  • Page 3

    Introduction this manual contains information and warnings which must be followed to ensure safe operation and retain the meter in safe condition. 1 the following safety precautions must be observed to ensure maximum personal safety during the operation, service and repair of this meter: safety prec...

  • Page 4

    2 1. Read these operating instructions thoroughly and completely be- fore operating your meter. Pay particular attention to warnings which will inform you of potentially dangerous procedures. The instructions in these warnings must be followed. 2. Always inspect your meter, test leads and accessorie...

  • Page 5

    3 safety information cleaning wipe the case with a damp cloth and mild detergent. Do not use abrosives or solvents. Dirt or moisture in the terminals can affect readings. Safety: conforms to iec 61010-1 (en 61010-1), iec 61010-2-032 (en 61010-2-032), cat iii 1000v, cat iv 600v, class ii, pollution d...

  • Page 6

    4 instrument layout 13 11 6 3 1 9 2 12 5 10 8 4 7 cat.Iv 600v cat.Iii 1000v 2000a clamp meter true rms off Ω hz v v hz hz hz % 1500a 660a 660a apo ac dc 0 20 10 30 40 50 60 n max range hold zero min 1500a max temp max 30v k-type cat iv 600v cat iii 1000v 750v 1000v 1832°f 1000°c temp Ω 2000a 1500a i...

  • Page 7

    1. V Ωhz% , voltage, ohms, frequency, duty cycle, capacitance, diode input terminal this is the positive input terminal for voltage, ohms, frequency, duty cycle, capacitance, diode measurements.Connection is made to it using the red test lead. 2. Com common terminal this is the negative (ground) inp...

  • Page 8

    6 1. Press the "inrush" button to toggle in the inrush mode, ane the "----" and "inrush" will be displayed. 2. Press the trigger to open transformer jaws and clamp onto single conductor only, and turn on the motor. 3. Read the inrush current directly from the display. 4. Depress the inrush button fo...

  • Page 9

    14.Other functions auto power off 1. Auto power off: approx. 30 minutes. 2. After auto power off, press any button to restart the meter, and the reading of measurement will be maintained in the display. Cancellation of auto power off feature: press and hold the (max/min) button while rotating functi...

  • Page 10

    Current measurements warning these snap-arounds are designed to take current measurements on circuits with a maximum voltage difference of 500vac between any conductor and ground potential. Using the snap-around for current measurements on circuits above this voltage may cause electric shock, instru...

  • Page 11

    Resistance measurements 1. Set the function/range switch to the " Ω " position. Use the " / / " button to select the resistance test. 2. Turn off power to the circuit under test. External voltage across the components causes invalid reading. 3. Connect the red test lead to the "v Ω " jack and the bl...

  • Page 12

    Capacitance measurements 1. Set the function/range switch to the " " position. 2. Connect the red test ldad to the “ v Ω “ jack and the black test lead to the “ com “ jack. 3. Touch the probes to the capacitor. Observe polarity when measur- ing polarized capacitors. 4. Read the capacitance directly ...

  • Page 13

    Specifications · display: 6600 counts, 66 segments analog bar-graph. · polarity: automatic, (-) negative polarity indication. · overrange indication: (ol) or (-ol) is displayed. · low battery indication: the " " is displayed when the battery voltage drops below accurate operating level. · measuremen...

  • Page 14

    Dc volts range 660mv 6.6v 66v 660v 1000v resolution 0.1mv 1mv 10mv 100mv 1v accuracy ± (0.5% rdg + 2d) ± (0.5% rdg + 2d) ± (0.5% rdg + 2d) ± (0.5% rdg + 2d) > 100m Ω 10m Ω 9.1m Ω 9.1m Ω 9.1m Ω overload protection: 1000v dc or 750v ac rms ac volts (true rms) range 660mv 6.6v 66v 660v 750v resolution ...

  • Page 15

    Dc current range resolution accuracy 0~660a±(2.0% rdg+5d) 660~1000a±(3.0% rdg+5d) 1000~2000a±(5.0% rdg+5d) 660a 2000a 0.1a 1a overload protection: 2000a dc for 60 seconds maximum. 13 ac current (true rms) range resolution accuracy (50 ~ 400hz) 0~660a±(2.0% rdg+10d)50~60hz 0~660a±(3.0% rdg+10d)61~400...

  • Page 16

    Range 6.6 nf 66 nf 660 nf 6.6 µf 66 µf 660 µf 6.6mf resolution 1pf 10pf 100pf 1nf 10nf 100nf 1uf accuracy ±(3.0% rdg + 30d) ±(3.0% rdg + 10d) ±(3.0% rdg + 30d) ±(3.0% rdg + 10d) ±(3.0% rdg + 10d) ±(3.0% rdg + 10d) ±(5.0% rdg + 10d) overload protection: 600v dc or ac rms capacitance 14 continuity tes...

  • Page 17

    15 %duty cycle range 5% to 95% resolution 0.1% accuracy ( 5v logic) ±(2.0% rdg+10d) >10us frequency range: 5% to 95% ( 40hz to 20khz ) the readings of % duty cycle measurements will show on the sub- display. Overload protection: 600v dc or ac rms pulse width frequency resolution 0.01hz 0.1hz 1hz 10h...

  • Page 18

    Maintenance maintenance consists of periodic cleaning and battery replacement. The exterior of the instrument can be cleaned with a dry clean cloth to remove any oil, grease or grime. Never use liquid solvents or detergents. Repairs or servicing not covered in this manual should only be performed by...

  • Page 19

    P/n: 7000-1879 2609c.P65 ?2009/12/15, 上午 09:15 19.