Wavecom TNT+ Instruction Manual - Specifications For:
Page 31
Specifications for:
TnT Portable Appliance Testers
Appliance Tester Specifications (for TnT, TnT-el & TnT+ models):
Lead continuity and Polarity test @ 240Vac
Circuit continuity test 0 – 5M
Ω (appliance sense)
15A DUT Socket for Appliance testing only (Not Load Testing)
Insulation
Voltage
500Vdc + < 5%
250Vdc + < 5%
Range 0 – 10.0M
Ω + < 5% of reading
Pass Limit > 1M
Ω
Earth Bond
Current 200Ma dc
Range 0 – 10.0
Ω + < 5% of reading
Pass Limit < 1.0
Current Load 10A (resistive load 10amp HRC fuse)
(X-Models Only 20A, Thermal overload protected manual reset).
Leakage Tester (for TnT-el & TnT+ models only):
Earth Leakage
240Vac supply mA
Range 0 – 30.0mA + < 5% of reading
Pass limit – dependent on test requirements
Load current max 10A resistive (80A inrush rated)
HBC fuse protected 500Vac 10A (M205)
(X-Models Only 20A, Thermal overload protected manual reset).
Power Test (for TnT+ model only):
Voltage 220 – 270 Vac + < 5% of reading
Current 0 -10 A + < 5% of reading (X-Models Only, 0 – 20A)
Load current max 10A resistive.
HBC fuse protected 500Vac.
(X-Models Only 20A, Thermal overload protected manual reset).
RCD Test (for TnT+ model only):
Trip Current 2 – 500mA + < 5% of value
Trip Time 0 – 300 ms + < 1ms
Battery powered when testing RCD
Display (for TnT, TnT-el & TnT+ models):
Backlit Graphics. 64 x 128 LCD PVC protected (Blue White)
40 x 70mm
Mechanical (for TnT, TnT-el & TnT+ models):
Case ABS plastic
Size: 200L x 175W x 70D
Weight: 1.5Kg (X-Models Only, 2.5Kg)
Safety (for TnT, TnT-el & TnT+ models):
The instrument in general meets the requirements of relevant
IEC, AS & CE.
Safety Class II (Double insulated)
CAT III 600V
DUT Test Fuse Protected at 10A. Only use HBC fuses warranty void if not used.
Tester Fuse Protected. 1A Non user serviceable.