Zeiss HBO 100 Operating Manual - 3 Start-Up
Microscope Illuminating System
Start-Up
HBO 100
Lamp Insertion and Change
Carl Zeiss
B 46-0063 e 03/06
11
3 Start-Up
3.1 Lamp Insertion and Change
Fig. 4
Inserting the HBO 103 lamp
− Before inserting the lamp, make sure the
illuminating system has been disconnected
from line power.
− The lamps may only be changed when they
are cooled down; allow the microscope
illuminator to cool down for at least
15 minutes to avoid the risks of burns and
explosion!
− The lamp may only be removed from the
packaging and inserted in the microscope
illuminator if a protection mask and safety
gloves are worn (see chapter 1.3).
− All clamping adjustments must be performed
carefully; strong heat during operation may
result in loose contacts.
1 Cooling
body
2 Fixation screws for top part of lamp
3 Lamp
mount
4 Clamping lever at lamp base
5 Control knob for collector
6 Lamp
7 Clamping lever at heat sink