Start-Up
N HBO 103 and N XBO 75
Carl Zeiss
XBO 75 Lamp Insertion/Change
Microscope Illuminating Systems
12
B 40-065 e 08/99
3.2
N XBO 75 Microscope Lamp
Lamp Insertion/Change
Fig. 6
Insert XBO 75 lamp
−
Before inserting the lamp, make sure that
no connection to the line is available.
−
The lamps may only be changed when they
are cooled down; allow the microscope
illuminator to cool down for 10...15 minutes
to avoid the risks of burns and explosion!
−
The XBO 75 lamp is under high pressure
even if it is not switched on.
It may only be removed from the packaging
and inserted in the microscope illuminator if
a protection mask and safety gloves are
worn. (See section 1.3)
−
The safety regulations of the lamp
manufacturer must be observed when the
lamp is inserted or changed.
(See section 1.3 - Notes on use)
−
All clamping adjustments must be performed
carefully;
pronounced heat during operation may
result in loose contacts.
1
Fixation screw for lamp housing
2
Control wheel for collector
3
Cooling body
4
Fixation screw for lamp mount in cooling body
(screw is hidden in the illustration by the cooling body)
5
Spring lever
6 Lamp
mount
7
Lamp