Zeiss S8 User Manual - page 27
M-30-1382-en
S8 / S81 Ceiling Mounts
Issue 12.0
Printed on 19. 11. 2007
New installation of S8 / S 81 ceiling mount
27
Due to the large number of parameters involved and the variety of poten-
tial building/ceiling mount constellations, Carl Zeiss Surgical is unable to
give an absolute guarantee for the vibration-free suspension of the ceiling
mount, even if the building meets the applicable ISO standards.
However, it is highly improbable that constant vibrations are transferred
to the ceiling mount if all requirements regarding vibrations in the OR
ceiling are met.
e Max. vibration velocity (RMS) at the installation points for the ceiling
mount.
e V
max
max
agram) for the specified frequency range.
Sources: Carl Zeiss Surgical (in-house study), ISO 10811, recommenda-
tions for ORs.
4
5
6,3
8
10
12,5
16
20
25
31,5
40
50
63
90
4
5
6,3
8
10
12,5
16
20
25
31,5
40
50
63
90
1
0,1
0,01
100
10
4
1
RMS
o
f
vibrat
ion
velocit
y
(m
m
/
sec)
One-third octaveband center frequencles (Hz)
One-third octaveband center frequencles (Hz)
RM
S
o
f
vib
ratio
n
velo
city
(m
ili-in
ch
/
sec)
Summary of S8
Page 1
S8 / s81 ceiling mounts planning manual m-30-1382-en issue 12.0 printed on 19. 11. 2007.
Page 2
2 m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 key to symbols warning! The warning triangle indicates potential sources of danger which may constitute a risk of injury for the user or a health hazard. Caution: the square indicates situations which may lead to malfunction, ...
Page 3: Contents
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 contents new installation of s8 / s 81 ceiling mount 5 technical data 7 on-site preparations 24 constructional requirements 29 planning the electrical installation 33 planning the mechanical installation 42 confirmation of the s...
Page 4
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 300 mm (11.8") ceiling panel extension 115 on-site preparations 116 constructional requirements 118 confirmation of the structural calculation and execution of installation 119 ordering data 121 index 125.
Page 5
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 5 new installation of s8 / s 81 ceiling mount technical data 7 s8 ceiling mount, dimensions 7 s81 ceiling mount, dimensions 8 pre-installation set 10 installation position in the or 12...
Page 6
6 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 s81 ceiling mount for ors with false ceilings, ordering data 54 s81 ceiling mount for ors without false ceilings, ordering data 57.
Page 7: Technical Data
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 7 technical data s8 ceiling mount, dimensions key 1 false ceiling 1 : 20 620 mm / 24.4t 250 mm 1 350° ± 90° 270° 95 0 mm / 37. 4t 950 mm / 3 7.4t 700 mm / 2 7.6t max. 1000 mm / 39.4t m...
Page 8
8 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 s81 ceiling mount, dimensions key 1 false ceiling 2 floor a distance a is specifically defined for each project. B distance from floor we recommend different distances from the floor...
Page 9
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 9 max. 350° ± 90° 270° 600 mm / 23.6t max. 930 mm / 36.6t 36 0 mm a b 1 2 1 : 20 228 mm / 9t 23 8 m m / 9 .4 t 620 mm / 24.4t 250 mm 9.8t 14. 2t 360 m m 14.2t s81.
Page 10
10 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 pre-installation set key 1 stud bolt 2 spacers 3 ceiling anchor plate 4 floor 5 ceiling panel 6 mount flange 7 false ceiling 8 interface plate 9 structural ceiling a is the distance...
Page 11
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 11 75 150 1 2 3 5 6 7 8 9 4 a b 620 mm / 24.4t 45 m m / 1. 8t 100 m m / 3.9t 75 mm / 3t (1393 lbf) (7007 lbf.Ft).
Page 12
12 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation position in the or the optimum installation position of the ceiling anchor plate is dependent on the required position of the operating table in the or. The following o...
Page 13: S81
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 13 1500 ± 200 mm 1 2 3 4 5 5 s8 s81 1 : 50 1250 ± 100 mm 1500 ± 200 mm 59.1 ± 7.9t 49.2 ± 3.9t 59.1 ± 7.9t 1250 ± 100 mm 49.2 ± 3.9t.
Page 14
14 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 distance from floor 1 distance from floor we recommend different distances from the floor, depending on the discipline involved. E ent 1,650 mm (65") e ophthalmology 1,800 mm (70.9"...
Page 15: S81
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 15 ca. 1 800 mm 1 2800 - 3300 m m 2650 - 3500 m m s8 s81 1 : 50 110. 2 - 129.9t 104. 3 - 137.8t ca. 7 0 .9t.
Page 16
16 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 operating position the ceiling mounts are shown in a typical operating position. The angle between the microscope arm and the carrier arm is 90°. At this angle, min- imum effort is ...
Page 17: S81
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 17 3 1 2 4 1600 mm / 63t 1350 mm / 53.1t 1100 mm / 43.3t (s8) 1400 mm / 55.1t (s8) 1150 mm / 45.3t (s81) 850 mm / 33.5t (s81) 250 mm / 9.8t 50 mm / 2t s8 s81 1 : 50 s8.
Page 18
18 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 top view 1 s8 normal installation 2 s81 normal installation if, due to other installations, the space available on the ceiling is not suf- ficient for normal installation, you can u...
Page 19: S81
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 19 1600 mm / 63t 1350 mm / 53.1t 1100 mm / 43.3t (s8) 1400 mm / 55.1t (s8) 3 1 2 4 850 mm / 33.5t (s81) 1150 mm / 45.3t (s81) s8 s81 1 : 50 s8.
Page 20
20 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 operating range of the ceiling mount s8 ceiling mount the operating range specified here applies to normal installation. For installation in cramped conditions, the dimensions shown...
Page 21
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 21 250 mm / 9.8t 1250 ± 500 mm / 49.2 ± 19.7t 1 2 3 1 : 20 s8.
Page 22
22 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 s81 ceiling mount the operating range specified here applies to normal installation. For installation in cramped conditions, the dimensions shown on page 16 and page 18 can be used ...
Page 23
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 23 250 mm / 9.8t 1000 ± 400 mm 1 3 2 1 : 20 39.4 ± 15.7t s81.
Page 24: On-Site Preparations
24 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 on-site preparations customer's responsibilities note: zeiss service staff can install the ceiling mount only if all points of the fol- lowing checklist applicable to the relevant i...
Page 25
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 25 warning! G if an existing ceiling anchor plate or intermediate piece must be ex- changed, never re-use the old anchors. New anchor holes must be drilled. When calculating the effect...
Page 26
26 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 building vibrations the on-site requirements also include the low-vibration design of the ceiling in the or. This must be taken into account right during the plan- ning phase for th...
Page 27
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 27 due to the large number of parameters involved and the variety of poten- tial building/ceiling mount constellations, carl zeiss surgical is unable to give an absolute guarantee for ...
Page 28
28 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 customer's preparatory responsibilities, overview e mounting of the pre-installation set. E installation and connection of the systems's power cables on the pre- installation set. E...
Page 29: Constructional Requirements
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 29 constructional requirements the actual load on the ceiling depends on a large number of different fac- tors. The requirements to be met by the ceiling or substructure result from th...
Page 30
30 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 mounting a pre-installation set on ceilings with air-supply ceiling caution: if the carrier arm systems are installed in air-supply ceilings, the method of installation must be spec...
Page 31
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 31 1 7 6 2 4 5 3 8 1 s 7 5 2 4 6 3 9 8 m z a.
Page 32
32 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007.
Page 33
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 33 planning the electrical installation the power lines on the site of installation must be run up to the interface plate. In addition, the potential equalization lines must be run to ...
Page 34
34 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installing the conduits the wiring can be run either above the false ceiling or in conduits installed in the structural ceiling. There are two options for the installation of con- d...
Page 35
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 35 connection of speedfokus af note: g determine the exact length between the speedfokus af connection and the ccu+ monitor. If the cable length exceeds 10 m (394"), a signal amplifier...
Page 36
36 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 mounting the wall socket note: e we recommend mounting wall socket (3) on the wall located behind the surgeon. Height approx. 1.4 m (55.1"). E please also note the chapter "installi...
Page 37
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 37 1 2 6 6 1400 m m / 55. 1t 3 4 5.
Page 38
38 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 on-site electrical installation the power lines on the site of installation must be run up to the interface plate. In addition, the potential equalization lines must be run to the t...
Page 39
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 39 h i e f a d g a d g b l1 l2 pe c l1 l2 pe l1 l2 pe l1 l2 pe 6 pe 5 pe pe 7 pe 8 pe 1 pe 2 4 pe 3 pe.
Page 40
40 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 checking the power connection caution: the ceiling mount can be set at two points to the rated voltage used in the country of destination: e transformer terminal on interface plate....
Page 41
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 41.
Page 42
42 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 planning the mechanical installation mounting the pre-installation set note: g check the installation conditions, in particular with respect to the room height. G mount the pre-inst...
Page 43
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 43 2 1 3 4 mm inch 605 23.8 517 20.4 25 1.0 Ø400 Ø15.7 Ø560 Ø22.0 Ø26 Ø1.0
Page 44
44 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 mounting the spacers g mount eight spacers (2) in the ceiling anchor plate (1). G firmly tighten the spacers (2) using a torque wrench with a 55 mm (2.2") fork wrench bit. Apply a t...
Page 45
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 45 mounting the interface plate note: the connector for potential equalization (1) and the terminals for power line (2) are provided on interface plate (3). G check whether interface p...
Page 46
46 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 300 mm (11.8") ceiling panel extension ceiling panel extension (1) is required if the distance between the struc- tural ceiling and the false ceiling (i.E. Dimension "a" in the orde...
Page 47: Of Installation
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 47 confirmation of the structural calculation and execution of installation sales order no.: ..............................................................................................
Page 48: Ordering Data
48 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 ordering data s8 ceiling mount for ors with false ceilings, ordering data note: please note that the relevant dimensions have to be entered in the fol- lowing drawing for each purch...
Page 49
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 49 order sheet for s8 ceiling mount in ors with false ceilings a mm b mm min 2800 mm max 3300 m m mi n 2 0 0 mm ma x 8 0 0 m m dimensions in inches: a = " min 7.9", max 31.5" b = " min...
Page 50
50 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007.
Page 51
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 51 s8 ceiling mount for ors without false ceilings, ordering data you must order the ceiling mount using the following order sheet only if the or does not provide a false ceiling. If t...
Page 52
52 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 note: please note that the relevant dimensions have to be entered in the fol- lowing drawing for each purchase order. G enter the data and enclose a copy of the drawing with your pu...
Page 53
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 53 order sheet for s8 ceiling mount in ors without false ceilings 1 2 4 3 f mm h mm i 5 length to which the cover must be cut: f = h - i - 52 mm f = min 197 mm, max 800 mm length of pr...
Page 54
54 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 s81 ceiling mount for ors with false ceilings, ordering data note: please note that the relevant dimensions have to be entered in the fol- lowing drawing for each purchase order. G ...
Page 55
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 55 order sheet for s81 ceiling mount in ors with false ceilings mm mm mm mi n 2 0 0 mm ma x 8 00 m m column length (ordered length): d = b - c - 810 mm d = mm min. 40 mm, max. 800 mm* ...
Page 56
56 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007.
Page 57
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 57 s81 ceiling mount for ors without false ceilings, ordering data you must order the ceiling mount using the following order sheet only if the or does not provide a false ceiling. If ...
Page 58
58 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 note: please note that the relevant dimensions have to be entered in the fol- lowing drawing for each purchase order. G enter the data and enclose a copy of the drawing with your pu...
Page 59
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 new installation of s8 / s 81 ceiling mount 59 order sheet for s81 ceiling mount in ors without false ceilings 1 2 4 3 mm h 5 length (f) to which the cover must be cut: f = a - 7 mm. F = min 193 mm, max 800 mm mm pre-installatio...
Page 60
60 new installation of s8 / s 81 ceiling mount m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007.
Page 61: Accessory Interface
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 accessory interface 61 accessory interface accessory interface 62 coupling of the zeiss system with accessories from other manufacturers 62 installation position 63 mount flange and ceiling panel 64 mount flange (flange plate) 6...
Page 62: Accessory Interface
62 accessory interface m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 accessory interface next to the flange area for the carrier arm of the ceiling mount, the mount flange also features a flange area for an accessory carrier. This is the ac- cessory interface designed as a ...
Page 63: Installation Position
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 accessory interface 63 installation position caution: when defining the positions and ranges of movement of the accessory equipment, please pay attention to the geometry of the respective ceiling mount and to the risk of collisi...
Page 64
64 accessory interface m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 mount flange and ceiling panel the s8 or s81 ceiling mount is secured with mount flange (1) on the pre- installation set already installed in the ceiling. In addition to the flange area for the carrier arm...
Page 65
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 accessory interface 65 2 2 1 3 4 5.
Page 66
66 accessory interface m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 mount flange (flange plate) 1 accessory interface (flange area for an accessory carrier) with 6 tapped holes m12 in a hole circle of 270 mm (10.6") dia. Accessories (illuminators, monitor bracket etc.) can...
Page 67
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 accessory interface 67 configuration example: mount flange with flange tube 1 mount flange 2 ceiling panel only one piece of the ceiling panel is shown. 3 flange tube (example) a flange tube of this type can be used for mounting...
Page 68: Pre-Installation Set
68 accessory interface m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 pre-installation set 1 stud bolt 2 spacers 3 ceiling anchor plate 4 floor 5 ceiling panel 6 mount flange 7 false ceiling 8 interface plate 9 structural ceiling 10 dimension a is the distance between the st...
Page 69: Customer'S Responsibilities
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 accessory interface 69 customer's responsibilities note: zeiss service staff can install the ceiling mount only if all points of the fol- lowing checklist applicable to the relevant installation conditions have been fulfilled. T...
Page 70
70 accessory interface m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 forces and torques warning! The structural engineer must ensure in each individual case that the struc- tural ceiling has a sufficient load capacity for the forces and torques listed below. He must also ta...
Page 71
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 accessory interface 71 electrical connections for accessory modules the power supply for the electrical accessory components can be con- nected to connectors "a" on the interface plate (connectors for or illumi- nators, terminal...
Page 72
72 accessory interface m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 key a connectors for or illuminators b terminal 1 for or illuminator 1 maximum terminal cross section = 16 mm2 (awg 6) c terminal 2 for or illuminator 2 maximum terminal cross section = 16 mm2 (awg 6) for ...
Page 73: Of Installation
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 accessory interface 73 confirmation of the structural calculation and execution of installation sales order no.: ........................................................................................... Customer address / deli...
Page 74
74 accessory interface m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007.
Page 75
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 discipline-specific installation positions 75 discipline-specific installation positions installation positions for ophthalmology and ent 77 installation position for p&r, hand 90 installation position for neuro, spine, uro and ...
Page 76
76 discipline-specific installation positions m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 the various surgical disciplines and on-site conditions require different in- stallation positions of the s8 and s81 ceiling mounts. This chapter describes a number of recommended in...
Page 77
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 discipline-specific installation positions 77 installation positions for ophthalmology and ent the s8 or s81 ceiling mount is used for ophthalmology and ent. Recommendations for ophthalmology preferred installation position for ...
Page 78
78 discipline-specific installation positions m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 preferred installation position for ophthalmology and ent s8 ceiling mount positioned above the patient's foot area (centered rela- tive to the operating table). Flange area for acce...
Page 79
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 discipline-specific installation positions 79 preferred installation position for ophthalmology and ent s8 ceiling mount positioned above the patient's foot area (centered rela- tive to the operating table). Flange area for acce...
Page 80
80 discipline-specific installation positions m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 preferred installation position for ophthalmology and ent s81 ceiling mount positioned above the patient's foot area (centered rel- ative to the operating table). Flange area for acc...
Page 81
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 discipline-specific installation positions 81 preferred installation position for ophthalmology and ent s81 ceiling mount positioned above the patient's foot area (centered rel- ative to the operating table). Flange area for acc...
Page 82
82 discipline-specific installation positions m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 alternative installation position for ophthalmology s8 ceiling mount positioned laterally to the patient (centered relative to the patient's head). Flange area for accessory carrier ...
Page 83
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 discipline-specific installation positions 83 alternative installation position for ophthalmology s8 ceiling mount positioned laterally to the patient (centered relative to the patient's head).Flange area for accessory carrier (...
Page 84
84 discipline-specific installation positions m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 alternative installation position for ophthalmology s81 ceiling mount positioned laterally to the patient (centered relative to the patient's head). Flange area for accessory carrier...
Page 85
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 discipline-specific installation positions 85 alternative installation position for ophthalmology s81 ceiling mount positioned laterally to the patient (centered relative to the patient's head). Flange area for accessory carrier...
Page 86
86 discipline-specific installation positions m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 alternative installation position for ent s8 ceiling mount positioned behind the patient's head. Flange area for ac- cessory carrier (2) in outward position. S8 2 1 r 6 8.9 t r 1 75 ...
Page 87
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 discipline-specific installation positions 87 alternative installation position for ent s8 ceiling mount positioned behind the patient's head. Flange area for ac- cessory carrier (2) in inward position. S8 2 1 r 7 50 m m r 2 9.5...
Page 88
88 discipline-specific installation positions m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 alternative installation position for ent s81 ceiling mount positioned behind the patient's head. Flange area for accessory carrier (2) in outward position. S81 1 2 r 14 00 m m r 5 5...
Page 89
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 discipline-specific installation positions 89 alternative installation position for ent s81 ceiling mount positioned behind the patient's head. Flange area for accessory carrier (2) in inward position. S81 2 1 r 1 40 0 m m r 5 5...
Page 90
90 discipline-specific installation positions m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation position for p&r, hand the s8 ceiling mount is used for p&r and hand surgery. The installation positions recommended in this chapter are also suitable for the following ...
Page 91
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 discipline-specific installation positions 91 recommended installation position for p&r, hand s8 ceiling mount positioned laterally to the operating table, approximately midway along the table. Flange area for accessory carrier ...
Page 92
92 discipline-specific installation positions m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 recommended installation position for p&r, hand s8 ceiling mount positioned laterally to the operating table, approximately midway along the table. Flange area for accessory carrier ...
Page 93
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 discipline-specific installation positions 93 installation position for neuro, spine, uro and gyn the s8 ceiling mount is used for the disciplines of neurology, spine, uro- logy and gynecology. The installation positions recomme...
Page 94
94 discipline-specific installation positions m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 recommended installation position for neuro, spine, uro and gyn s8 ceiling mount positioned laterally to the operating table, approximately midway along the table. Flange area for ac...
Page 95
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 discipline-specific installation positions 95 recommended installation position for neuro, spine, uro and gyn s8 ceiling mount positioned laterally to the operating table, approximately midway along the table. Flange area for ac...
Page 96
96 discipline-specific installation positions m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007.
Page 97
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 97 installation on an existing substructure installation options 99 working ranges 99 mounting options 100 adapter plate (1071-924) 100 installation on an intermediate piece of s23 or emd...
Page 98
98 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007.
Page 99: Installation Options
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 99 installation options if certain requirements are met, the s8 and s81 ceiling mounts can be in- stalled on existing substructures provided for the s23, emd2, s5, s51, s6 or s61 ceiling ...
Page 100
100 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 mounting options using the adapter plate, the s8 or s81 ceiling mount can be mounted on: e existing intermediate piece (305658-0000.000) of an s23 or emd2 ceiling mount, with the adap...
Page 101
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 101 2 2 1 mm inch Ø14 Ø0.55 Ø15 Ø0.6 Ø270 Ø10.6 Ø160 Ø6.3 Ø440 Ø17.3 120 4.7 140 5.5 mm inch 250 9.8 500 19.7 90 3.5 123 4.8 150 5.9 275 10.8 550 21.7 t = 20 mm (0.8t).
Page 102
102 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an intermediate piece of s23 or emd2 mounting the adapter plate in an offset position adapter plate mounted with a 90 mm (3.5") lateral offset the following two illust...
Page 103
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 103 working range of the s81 ceiling mount with a 90 mm (3.5") lateral offset of the adapter plate 1 pivot axis the radii of the working range r = 600 mm (23.6") and r = 1,400 mm (55.1") ...
Page 104
104 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 adapter plate mounted with a lateral offset and electrical components mounted on the mounting bracket note: the mounting bracket can only be used to mount the electrical compo- nents ...
Page 105
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 105 1 2 3 4 5 min. 3 10 mm / 12. 2" 100 mm 3. 9" 45 m m 1. 8" mi n . 5 0 mm min. 2.0" 100 (50) mm 3.9 (2.0)" 240 (200) mm 9.4 (7.9)".
Page 106
106 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 mounting the adapter plate in a centered position electrical components mounted on the mounting bracket note: the mounting bracket can only be used to mount the electrical compo- nent...
Page 107
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 107 1 2 3 4 5 100 mm 3. 9" 45 mm 1. 8 " mi n . 5 0 m m mi n . 2 .0 " 150 mm 5.9" min . 310 mm / 12. 2".
Page 108
108 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 electrical components mounted on the adapter plate note: the electrical components can only be mounted on adapter plate (3) if the clearance between intermediate piece (1) and adapter...
Page 109
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 109 1 2 3 4 100 mm 3. 9" 45 m m 1.8 " mi n . 1 0 8 m m mi n. 4. 3" 150 mm 5.9".
Page 110
110 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing zdk intermediate piece the zdk intermediate piece (305942-9103.000) was used to install the s5, s51, s6 and s61 ceiling mounts. Mounting electrical compone...
Page 111
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 111 1 2 3 4 mi n . 3 1 0 m m / 1 2 .2 " 100 mm 3. 9" 45 mm 1.8" mi n . 5 0 mm min. 2.0".
Page 112
112 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 electrical components mounted on the adapter plate note: the electrical components can only be mounted on adapter plate (3) if the clearance between intermediate piece (1) and adapter...
Page 113
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 113 1 2 3 4 45 mm / 1.8" mi n. 108 mm mi n . 4 .3 " 10 0 mm 3.9 ".
Page 114
114 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on ceiling track mount you can install the s81 ceiling mount on an existing ceiling track mount. The s8 ceiling mount cannot be installed on a ceiling track mount. The fo...
Page 115
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 115 300 mm (11.8") ceiling panel extension ceiling panel extension (1) is required to completely cover any interme- diate pieces of the s23, emd2, s5, s51, s6 or s61 ceiling mount protru-...
Page 116: On-Site Preparations
116 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 on-site preparations customer's responsibilities note: zeiss service staff can install the ceiling mount only if all points of the fol- lowing checklist applicable to the relevant ins...
Page 117
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 117 warning! G if an existing flange plate or intermediate piece must be exchanged, never re-use the old anchors. New anchor holes must be drilled. When calculating the effective strength...
Page 118: Constructional Requirements
118 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 constructional requirements the actual load on the ceiling depends on a large number of different fac- tors. The requirements to be met by the ceiling or substructure result from the ...
Page 119: Of Installation
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 119 confirmation of the structural calculation and execution of installation sales order no.: ........................................................................................... C...
Page 120
120 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007.
Page 121: Ordering Data
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 121 ordering data mounting an s8/s81 ceiling mount on an existing intermediate piece caution: when ordering an s81 ceiling mount for installation on an existing inter- mediate piece, you ...
Page 122
122 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 mounting an s81 ceiling mount on an existing ceiling track mount note: please note that the relevant dimensions have to be entered in the fol- lowing drawing for each purchase order. ...
Page 123
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 installation on an existing substructure 123 order sheet for s81 ceiling mount on ceiling track mount mm d c mm column length: d = b - c - 910 mm d = mm min 40 mm, max 800 mm b dimensions (inches): b = " c = " column length: d =...
Page 124
124 installation on an existing substructure m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007.
Page 125: Index
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 index 125 index numerics 300 mm ceiling panel extension (1177-632) 115 a accessories from other manufacturers, coupling with zeiss system 62 accessories on opmi, cable installation 34 accessory components, power supply 71 access...
Page 126
126 index m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 electrical connections for accessory modules 71 electrical installation, on site 38 emd2 and s2 ceiling mounts 99 ent, alternative installation position 86, 87, 88, 89 execution of installation, confirmation of 47, 73,...
Page 127
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 index 127 mounting bracket (1124-211) 121, 122 mounting kit (1120-925) 121, 122 mounting kit, special 99 mounting options 100 mounting the spacers 44 n neurology, recommended installation positions 93 o on-site conditions 76 on-...
Page 128
128 index m-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 scale, same as layout plan 76 signature, confirm by 47, 73, 119 spacers 10 structural calculation 47, 73, 119 structural calculation, confirmation of 47, 73, 119 structural ceiling 10 structural engineer 47, 73, 119 st...
Page 129
M-30-1382-en s8 / s81 ceiling mounts issue 12.0 printed on 19. 11. 2007 129.
Page 130
Carl zeiss surgical gmbh a carl zeiss meditec company 73446 oberkochen germany fax: ++49 (0) 7364-20 48 23 email: surgical@meditec.Zeiss.Com internet: www.Meditec.Zeiss.Com m-30-1382-en printed in germany cg-st.