3M Bair paws 875 Service Manual

Manual is about: warming unit

Summary of Bair paws 875

  • Page 1

    Warning: electrical shock hazard. There are electrically live parts within the warming unit when it is connected to a power source, even when the handheld temperature controller is in the off position. Please forward to biomedical engineering department 3 bair paws ™ model 875 warming unit service m...

  • Page 2

    Check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a.

  • Page 3: Bair Paws

    3 bair paws model 875 warming unit service manual 43 º c 1 model 875 warming unit service manual 2 appareil de réchauffement, modèle 875 manuel d’entretien 3 erwärmungssystem modell 875 wartungshandbuch check the 3m tm bair paws tm system website to ensure you have the most recent version of this do...

  • Page 4: Table of Contents

    Table of contents introduction to the 3m™ bair paws™ patient adjustable warming system 1 bair paws model 875 warming unit ...

  • Page 5

    1 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a introduction to the 3m™ bair paws™ patient adjustable warming system the 3m tm bair paws tm patient adjustable warming system consists of the m...

  • Page 6: Definition Of Symbols

    Bair paws model 875 warming unit service manual 2 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a important information about the bair paws model 875 warming unit indications the model 875 warming unit is indi...

  • Page 7

    3 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a manufacturer this system is subject to european weee directive 2002/96/ec this product contains electrical and electronic components and must ...

  • Page 8

    Bair paws model 875 warming unit service manual 4 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a • do not allow the patient to lie on the warming unit hose or allow the hose to contact the patient’s skin duri...

  • Page 9

    5 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a 7 to reduce the risks associated with fire: • always follow standard safety protocols when using high intensity heat sources near a bair paws ...

  • Page 10

    Bair paws model 875 warming unit service manual 6 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a proper use and maintenance arizant healthcare inc , a 3m company assumes no responsibility for the reliability,...

  • Page 11: Preparing The Bair Paws

    7 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a preparing the bair paws model 875 warming unit for use before using the warming unit, make sure that it is safely placed on a flat, hard surfac...

  • Page 12

    Bair paws model 875 warming unit service manual 8 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a placing the warming unit on the wall-mount bracket 1 snap the warming unit into position by sliding the clip o...

  • Page 13

    9 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a mounting the warming unit on an iv pole caution: to prevent tipping, mount the model 875 warming unit on an iv pole at a height that ensures st...

  • Page 14

    Bair paws model 875 warming unit service manual 10 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a mounting the warming unit on a bedrail 1 slide the clip on the back of the warming unit into the groove of th...

  • Page 15

    11 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a mounting the warming unit with a rail-mount to the wall caution: the rail-to-wall bracket is intended for use only with a bair paws 800 series...

  • Page 16

    Bair paws model 875 warming unit service manual 12 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a mounting the warming unit with a suction wall mount adapter 1 slide the bair paws suction wall mount adapter ...

  • Page 17

    13 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a using the handheld temperature controller after the hose has been connected to the bair paws gown, the patient can adjust the air temperature ...

  • Page 18: Service Procedures

    Bair paws model 875 warming unit service manual 14 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a service procedures calibrating the operating temperatures service frequency • always recalibrate after perform...

  • Page 19

    15 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a method 1 turn the warming unit on and select the maximum airflow and temperature by turning the temperature controller completely clockwise (...

  • Page 20

    Bair paws model 875 warming unit service manual 16 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a 6 insert a flat-headed, 2-mm screwdriver into the calibration access hole to a depth of approximately 1/2 inc...

  • Page 21

    17 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a method note: the ot threshold temperature cannot be adjusted if problems occur, contact 3m patient warming customer service 1 if continuing...

  • Page 22

    Bair paws model 875 warming unit service manual 18 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a replacing the filter service frequency every six months tools/equipment needed • medium phillips screwdriver ...

  • Page 23

    19 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a replacing the temperature controller and/or hose tools/equipment needed • small phillips screwdriver • medium phillips screwdriver method 1 ...

  • Page 24

    Bair paws model 875 warming unit service manual 20 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a 7 to replace the hose or controller remove the two screws that connect the wire collar to the mid-hose couple...

  • Page 25

    21 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a replacing the fuses tools/equipment needed • small phillips screwdriver • medium phillips screwdriver method 1 disconnect the warming unit fr...

  • Page 26

    Bair paws model 875 warming unit service manual 22 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a replacing the power cord tools/equipment needed • medium phillips screwdriver method 1 disconnect the warmin...

  • Page 27: General Maintenance

    23 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a general maintenance calibrating the operating temperatures calibrate the operating temperatures of the model 875 warming unit (1) after perfor...

  • Page 28

    Bair paws model 875 warming unit service manual 24 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a technical support and customer service u.S. Customer service tel: 800-228-3957 outside of the usa contact your...

  • Page 29: Technical Specifications

    25 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a technical specifications physical characteristics dimensions of warming unit 13 in high x 4 in deep x 7 7 in wide 33 cm high x 10 2 cm deep...

  • Page 30

    Bair paws model 875 warming unit service manual 26 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a safety system characteristics thermostat independent electronic and heater (electromechanical) overcurrent pro...

  • Page 31: Maintenance Log

    27 english 202036a check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a maintenance log date maintenance action performed.

  • Page 32: Maintenance Log

    Bair paws model 875 warming unit service manual 28 check the 3m tm bair paws tm system website to ensure you have the most recent version of this document. Www.Bairpaws.Com reorder #202036a maintenance log date maintenance action performed.

  • Page 33

    Made in the usa of globally sourced material by 3m health care. 3m is a trademark of 3m company, used under license in canada. Bair hugger, bair paws and the bair paws logo are trademarks of arizant healthcare inc., used under license in canada. ©2013 arizant healthcare inc. All rights reserved. 3m ...