3M Bair paws 875 Service Manual - page 7
3
English 202036A
Check the 3M
TM
Bair Paws
TM
system website to ensure you have the most recent version of this document.
www.bairpaws.com reorder #202036A
Manufacturer
This system is subject to European WEEE Directive 2002/96/EC
This product contains electrical and electronic components and must
not be disposed of using standard refuse collection Please consult local
directives for disposal of electrical and electronic equipment
Consult instructions for use
Follow instructions for use
Recycle to avoid environmental contamination This product contains
recyclable parts For information on recycling, please contact your nearest
3M Service Center for advice
Keep dry
Temperature limit
Explanation of Signal Word Consequences
WARNING:
Indicates a hazardous situation which, if not avoided, could result in death or serious injury
CAUTION:
Indicates a hazardous situation which, if not avoided, could result in minor or moderate
injury
NOTICE:
Indicates a situation which, if not avoided, could result in property damage only
Contraindication
To reduce the risks associated with thermal energy:
•
Do not apply heat to lower extremities during aortic cross-clamping as thermal injury
may occur if heat is applied to ischemic limbs
WARNING:
1
Do not use a forced-air warming device over transdermal medications; increased drug
delivery, patient death, or injury may occur
2
To reduce the risks associated with thermal energy hazards:
• Do not continue use of the unit if the yellow Over-temp indicator light illuminates
or an audible alarm sounds; contact a biomedical technician or call 3M Patient
Warming technical service
• Use only the Bair Paws warming gowns with this warming unit
•
Do not warm patients with the warming unit’s hose alone
Always connect the hose to a Bair Paws warming gown before providing
patient warming