A&D FS-02i Instruction Manual

Manual is about: FS-i Series check weighting scales option

Summary of FS-02i

  • Page 1

    Wm+pd4001368 f f s s - - i i s s e e r r i i e e s s c c h h e e c c k k w w e e i i g g h h i i n n g g s s c c a a l l e e s s o o p p t t i i o o n n op-02 / op-03/ op-04.

  • Page 2

    - 1 - contents 1. Op-02 battery .............................................................................................. 2 1-1. Using the op-02 sla battery.................................................................... 2 2. Op-03 rs-232c / relay output .......................................

  • Page 3: 1. Op-02 Battery

    - 2 - 1. Op-02 battery 1-1. Using the op-02 sla battery the scale can be operated with an sla (sealed lead acid) battery that will be commercially available. The scale (with no other options) can be operated with a fully charged battery for; lcd backlight off and comparator light off: approximately ...

  • Page 4

    - 3 - 1. Disconnect the main power cord from the outlet. 2. Remove the ten hexagon head screws and open the rear cover. Take care not to drop the rear cover or the wires connecting between front display and rear cover will be damaged. 3. Loosen the screws fixing the battery support and remove it. 4....

  • Page 5

    - 4 - 2. Op-03 rs-232c / relay output this interface allows the fs-i series to be connected with a multifunction printer or a personal computer, and the relay outputs for comparator result are obtained. The op-03 unit includes an interface board, a connector cable, a cable gland and two screws (m3x6...

  • Page 6

    - 5 - 1. Disconnect the main power cord from the outlet. 2. Remove the ten hexagon screws and open the rear cover. 3. Connect the cable from external device through the cable gland to the terminal blocks on the interface board. Also connect a yellow/green wire inside the display pod directly to “fg”...

  • Page 7

    - 6 - circuit diagram data format s t , + 0 0 0 0 0 . 0 0 _ k g c r l f header data unit terminator there are 3 headers for the weighing data. St: stable weighing data us: unstable weighing data ol: out of weighing range the data consists of 9 characters including polarity and decimal point. There a...

  • Page 8

    - 7 - out of rage (+) o l , + 9 9 9 9 9 9 9 9 _ k g c r l f data output mode (f06) stream mode (f06-0) data is sent continuously. The data-update rate is approximately 20 times per second. But for 2400 bps, it is too slow to send 20 times. If necessary, set the baud rate 4800 bps or more. Command mo...

  • Page 9

    - 8 - command definition notes ct clear tare weight. Including preset tare. ?Pt send a preset tare weight in-use. ?Tr send a tare weight in-use. ?Ok send a target weight in-use. ?Hi send a hi limit value or upper limit weight in-use. ?Lo send a lo limit value or lower limit weight in-use. Send a set...

  • Page 10

    - 9 - set a preset tare weight. (no reply for f20-1.) command p t , + 0 0 1 2 0 0 c l f decimal point follows the weight display. Reply p t , + 0 0 1 2 0 0 c l f fs-15ki : the preset tare weight 1.200 kg is subtracted and the display shows net weight. Clear tare weight. (no reply for f20-1.) command...

  • Page 11

    - 10 - reply o k , + 0 0 1 0 0 0 c l f fs-15ki : target weight 1.000 kg will be set. Set a hi limit value or upper limit weight. (no reply for f20-1.) f07-0 or f07-1 is set, the command should have “+” and 6 digit number without decimal point. Command h i , + 0 0 0 2 0 0 c l f decimal point follows ...

  • Page 12

    - 11 - command m l , 0 1 , + 0 0 1 0 0 0 , + 0 0 0 2 0 0 , + 0 0 0 1 0 0 c l f reply m l , 0 1 , + 0 0 1 0 0 0 , + 0 0 0 2 0 0 , + 0 0 0 1 0 0 c l f fs-15ki : store target weight 1.000 kg, upper limit weight +0.200 kg and lower limit weight +0.100 kg into the memory number “01”. F07-2 is set, the co...

  • Page 13

    - 12 - the maximum number of text data is 300 characters. ‰ to use the ufc function, the text data has to be sent to the scale from the computer using “pf” command in the command mode (f06-1) in advance. Then, connect the scale with the printer and set the f06-2, f06-3 or f06-4. When the print key i...

  • Page 14

    - 13 - if you will send more than 2 lines of data, attach “&” to the end of the first line. Then, the scale considers the data to be continued. A “space” or “,” will be used to separate these data. You can skip them to reduce the number of characters, but you cannot skip “,” after “pf”. You must sta...

  • Page 15

    - 14 - 3. Op-04 rs-422 / 485 / relay output this interface allows a personal computer to connect and control up to 16 fs-i scales, and the relay outputs for comparator result are obtained. The op-04 unit includes an interface board, a connector cable, two cable glands and two screws (m3x6). The rela...

  • Page 16

    - 15 - installation is similar to the op-03. See “2-1. Installation”. Set the function parameters f04, f05, f06, f18, f19 and f20 according your application. The function setting f19-1 must be set for rs-422 and f19-2 must be rs-485. To connect more than one scale with a computer, set a different ad...

  • Page 17

    - 16 - example of connection rs-422 rs-485 the polarity (a, b) of the host computer signal depends on model. Check the technical manual of the computer. 3-3. Data output data and commands for the rs-422/485 are same as the rs-232c except following. When used with the setting f19-1 (rs-422) or f1-2 (...

  • Page 18

    - 17 - examples of command and reply ( “_” shows “space” (20h).) following examples are for f20-0 (reply to command is sent.) the address ## = 23 (f18-23). Request a weight data. Command @ 2 3 q c l f reply @ 2 3 s t , + 0 0 1 2 . 3 4 5 _ k g c l f stable data @ 2 3 u s , + 0 0 0 7 . 8 9 0 _ k g c l...