A&D LCC12T010 Instruction Manual - page 2
- 8 -
3. Specifications
Rated
capacities.....................................100 kN, 200 kN, 300 kN
Rated
output...........................................2mV/V±0.1%
Maximum safe overload .........................150%R.C.
Permissible
horizontal force ...................75 kN
Permissible
lifting force...........................80 kN
Combined
error ......................................±0.03%R.O.
Zero
balance ..........................................±1%R.O.
Compensated
temperature range...........-20
°C to 60°C
Minimum
excitation voltage ....................5VDC
Recommended
excitation
voltage ..........5 to 12VDC
Maximum
excitation voltage ...................15VDC
Input
terminal resistance ........................800
Ω±80 Ω
Output
terminal resistance......................2200
Ω±10 Ω
Insulation
resistance...............................5000M or greater
Ω at 50VDC
Temperature
effect - Zero.......................0.019%R.O./10
°C Typ.
Temperature
effect - Span......................0.010%R.O./10
°C Typ.
Cable
diameter/ length ...........................
φ8/12m
Dustproof/Waterproof .............................IP68 (100h at 1.5m immersion)
Cable connection color
Red................................................Excitation + (Input)
White .............................................Excitation - (Input)
Green ............................................Signal + (Output)
Blue ...............................................Signal - (Output)
Yellow............................................Shield
4. Parts Description
Stopper bolt
(One each on the
left and right side)
Upper mount fixture
(Maximum movable distance: ±5 mm in all directions)
Anti-vibration/lift-off
protection stopper unit
Lower mount fixture
Load cell
Anti-vibration/lift-off protection
stopper unit
1WMPD4000785B