A&D SW-D Setup Instructions

Manual is about: Weighing Indicator For Super Washdown Scales

Summary of SW-D

  • Page 1

    S s w w - - d d weighing indicator f f o o r r s s u u p p e e r r w w a a s s h h d d o o w w n n s s c c a a l l e e s s 1wmpd4002414.

  • Page 2

    Warning definition the warning described in this manual has the following meaning: danger an imminently hazardous situation which, if not avoided, will result in death or serious injury. © 2011 a&d company, limited. All rights reserved. No part of this publication may be reproduced, transmitted, tra...

  • Page 3: Contents

    1 contents 1. Compliance ..................................................................................................... 2 2. Introduction .................................................................................................. 3 3. Features..............................................

  • Page 4: 1. Compliance

    2 1. Compliance compliance with fcc rules please note that this equipment generates, uses and can radiate radio frequency energy. This equipment has been tested and has been found to comply with the limits of a class a computing device pursuant to subpart j of part 15 of fcc rules. These rules are d...

  • Page 5: 2. Introduction

    3 2. Introduction this manual describes how this product works and how to get the most out of it in terms of performance. The sw-d weighing indicator may be used to build a water-proof digital scale to specification ip69k. It has similar specifications and functions as an indicator of the sw series....

  • Page 6: 4. Unpacking

    4 4. Unpacking unpack the indicator carefully and keep the packing material if you are likely to transport the indicator again in the future. When unpacking, check whether all of the following items are included: model label (1083013259-2) sw-d capacity section write the scale’s ratings, etc.. Sw se...

  • Page 7

    5 5. Installation and precautions 5.1. Precautions for installing the indicator danger ground the indicator so that the user will not be subjected an electric shock. Do not handle the main power cord with wet hands. The ac plug is not water-resistant. Install it in an area where it does not get wet....

  • Page 8

    6 4. Solder the wires to the pin contacts of the connector. A 5 m or less load cell cable is recommended as the sw-d uses a 4 wire system (no remote sensing). 5. Tighten the cable clamp locking nut to secure the load cell cable. 6. Close the load cell panel by securing the four screws and cover the ...

  • Page 9

    7 example of calculation: in the case of designing a scale with a capacity of 60 kg and minimum division of 0.005 kg, using one load cell with a rated capacity of 100 kg and rated output of 1 mv/v: rated capacity of load cell: a = 100 kg rated output: b = 1 mv/v minimum division: d = 0.005 kg excita...

  • Page 10: 6. Description of Each Part

    8 6. Description of each part symbols symbol description stable ○ turns on when the weight value is stable. Net ○ turns on when the net weight is displayed. (the tare operation is in progress.) zero ○ turns on when zero is displayed. Turns on when the comparator results are displayed. Weighing units...

  • Page 11

    9 key description on/off key turns the power on or off. When turned on, the scale will be automatically set to zero (power-on zero). To turn the power off, press and hold the key. Note: if pressing the key does not turn the power on immediately, keep pressing the key. Sample key in the counting mode...

  • Page 12: 7. Calibration

    10 when calibrating, refer to this set up instructions. (the contents of sw instruction manual are not the same as this set up instructions.) 7.1. Items of the calibration mode the calibration mode has the following three functions. •setting of the weighing capacity, minimum division, and decimal po...

  • Page 13

    11 the chart below shows the relation between each item and the key operation in the calibration mode. Press the units key to select an item and press the tare key to execute it. To exit from the calibration mode, press the zero 1 key or the cal switch. The indicator turns the power off. (or press t...

  • Page 14

    12 7.2. Setting of the weighing capacity / minimum division / decimal point position set the weighing capacity, minimum division, and decimal point position. Be sure to perform these setting first when connecting to a load cell (base unit). The indicator does not require setting each time, if nothin...

  • Page 15

    13 7.3. Gravity acceleration correction when the scale is first used or has been moved to another location, it should be calibrated using a calibration weight. But if a calibration weight is not available, the gravity acceleration correction will compensate the scale. Change the gravity acceleration...

  • Page 16

    14 7.4. Calibration using a weight this function adjusts the scale (indicator) for accurate weighing. Calibration must be done when the sw-d is initially connected to a load cell (base unit) or the combination is changed. Calibration may also be required according the changes in the environment. Whe...

  • Page 17

    15 7. Place the calibration weight with the same value as displayed on the weighing pan, and wait for the stable indicator to turn on. 8. Press the tare key. The scale calibrates span and “ end ” is displayed. After displaying “ cal ”, the setting is complete. 9. To finish the setting procedure, pre...

  • Page 18

    16 8. Setting functions or units for use the indicator has cf function settings to specify the indicator performance. After installing the indicator, the indicator does not require resetting or changing of this cf function as far as it is used normally. If you want to maintain only the needed units,...

  • Page 19

    17 8.2. Setting unit for use 1. refer to “7.1. Items of the calibration”, to enter the calibration mode. The indicator displays “ cal5et ”. 2. Press the mode key three times to display “ cf ”. 3. Press the tare key to display the setting item. 4. To proceed without storing this setting item, press t...

  • Page 20

    18 8.3. Cf function list item parameter description used internally cf 1 3 not available for setting change used internally cf 2 0 not available for setting change 0 not available for setting of function “trc” (it is fixed to “trc 1”) zero tracking cf 3 1 available for setting of function “trc” when...

  • Page 21: 9. Options

    19 9. Options for details of hc-02i, sw-03, and sw-04, refer to the “options” in the sw series instruction manual. 9.1. Sw-11 display stand / wall-mount sw-11 provides the function as a display stand or a wall-mount for the sw-d. Attaching to sw-11 1. Attach the sw-11 to the display using the four s...

  • Page 22: 10. Maintenance

    20 10. Maintenance 10.1. Notes on maintenance do not disassemble the indicator. Contact your local a&d dealer if the indicator needs service or repair. Use the original packaging for transportation. Do not use organic solvents to clean the indicator. Use a warm lint free cloth dampened with a mild d...

  • Page 23: 11. Specifications

    21 11. Specifications 11.1. Specifications input sensitivity 0.2 µv / d min. (d: minimum division) input signal range -16 mv ~ 16 mv load cell excitation voltage 5 v dc ±5% (no remote sensing) load cell drive capacity up to 4 x 350 ohm load cells zero ± (0.2 µv ±0.001% of zero adjustment voltage) / ...

  • Page 24

    22 11.2. External dimensions ● when attaching to sw-11 unit: mm 255 20 .5 15 8 118.9 50 52 .5 160 52 .5 47.5 47.5 4-m4 depth 8mm 315 283 15 2.5 19 7 .5 250 90 70 6- 5 φ 135.9 110 160 50 65 4-m4 x 8 truss screw.

  • Page 25: Gravity Acceleration Map

    23 gravity acceleration map values of gravity at various locations amsterdam 9.813 m/s 2 manila 9.784 m/s 2 athens 9.807 m/s 2 melbourne 9.800 m/s 2 auckland nz 9.799 m/s 2 mexico city 9.779 m/s 2 bangkok 9.783 m/s 2 milan 9.806 m/s 2 birmingham 9.813 m/s 2 new york 9.802 m/s 2 brussels 9.811 m/s 2 ...

  • Page 26

    24 world map.