Wrist digital blood pressure monitor model ub-351 instruction manual manuel d’instructions manual de instrucciones manuale di istruzioni 使用手冊 1wmpd4002389b original traduction traducción traduzione 翻譯.
Page 2: Contents
English 1 contents dear customers.............................................................................................................................................. 2 preliminary remarks .........................................................................................................
Page 3: Dear Customers
English 2 dear customers congratulations on purchasing a state-of-the-art a&d blood pressure monitor, one of the most advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen. We recommend that you read through this manu...
Page 4
English 3 when reusing the device, confirm that the device is clean. Do not modify the device. It may cause accidents or damage to the device. To measure blood pressure, the wrist must be squeezed by the cuff hard enough to temporarily stop blood flow through the artery. This may cause pain, numbnes...
Page 5: Parts Identification
English 4 parts identification battery cover cuff display memory button start button alkaline batteries (lr03 or aaa) heart mark memory who classification indicator and pressure bar indicator i.H.B. Symbol (irregular heartbeat symbol) battery indicator systolic pressure diastolic pressure pulse rate...
Page 6: Symbols
English 5 symbols symbols that are printed on the device case symbols function / meaning recommended action standby and turn the device on battery installation guide direct current sn serial number date of manufacture type bf: device and cuff are designed to provide special protection against electr...
Page 7: Symbols
English 6 symbols symbols that appear on the display symbols function / meaning recommended action appears while measurement is in progress. It blinks when the pulse is detected. Measurement is in progress. Remain as still as possible. Irregular heartbeat symbol (i.H.B.) appears when an irregular he...
Page 8: Operation Mode
English 7 operation mode 1. Normal measurement press the start button. Blood pressure is measured and the data is stored in memory. This device can store the last 90 measurements in memory. 2. Recalling the memory data measurements stored in memory are assigned an index number in the order of the ne...
Page 9: Using The Monitor
English 8 using the monitor installing / changing the batteries 1. Remove the battery cover. 2. Remove the used batteries and insert new batteries into the battery compartment as shown, taking care that the polarities (+ and -) ae correct. Use only lr03 or aaa batteries. 3. Attach the battery cover....
Page 10: Using The Monitor
English 9 using the monitor applying the cuff 1. Wrap the cuff around your wrist about 1 cm above your hand as shown in the figure at the right. 2. Fasten the cuff tightly using the velcro strip. Note: for proper measurements, fasten the cuff tightly and measure on a bare wrist. How to take accurate...
Page 11: Measurements
English 10 measurements before measurement, read “notes for accurate measurement” on page 11. Normal measurement 1. Sit comfortably with the cuff at the same level as your heart and relax. 2. Press the start button. All of the display segments are displayed. Then the display changes, as indicated in...
Page 12: Measurements
English 11 measurements for those whose systolic pressure is over 150 mmhg 1. Wrap the cuff around your wrist. Raise your hand so that the cuff is at the same level as your heart. 2. Press and hold the start button until a number about 30 to 40 mmhg higher than your expected systolic pressure appear...
Page 13: Recalling The Memory Data
English 12 average number of measurements stored in memory recalling the average reading the ub-351 provides an average blood pressure reading based on the total measurements stored in memory. Press the memory (“m”) button to see the average blood pressure reading. The average blood pressure reading...
Page 14: Pressure Bar Indicator
English 13 what is an irregular heartbeat? Model ub-351 blood pressure monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressur...
Page 15: Who Classification Indicator
English 14 who classification indicator each six segments of the bar indicator correspond to the who blood pressure classification described on the next page. About blood pressure what is blood pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressur...
Page 16
English 15 to prevent hypertension or keep it under control: do not smoke exercise regularly reduce salt and fat intake have regular physical checkups maintain proper weight why measure blood pressure at home? Blood pressure measured at a clinic or doctor's office may cause apprehension and can prod...
Page 17: Troubleshooting
English 16 troubleshooting problem possible reason recommended action batteries are drained. Replace all batteries with new ones. Nothing appears in the display, even when the device is turned on. Battery terminals are not in the correct position. Reinstall the batteries with negative and positive t...
Page 18: Maintenance
English 17 maintenance do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, contact the authorized dealer in your area or our customer service department. The a&d customer...
Page 19
1-243 asahi, kitamoto-shi, saitama 364-8585 japan telephone: [81] (48) 593-1111 fax: [81] (48) 593-1119 a&d instruments ltd. Unit 24/26 blacklands way, abingdon business park, abingdon, oxfordshire ox14 1dy united kingdom telephone: [44] (1235) 550420 fax: [44] (1235) 550485 a&d engineering, inc. 17...