9 Maintenance
52
S4
D184B122U02
9.8.3 Warning
Messages During Operation
1) Warning is displayed for 30s
2) Converter must include the communication "HART-Protocol" option
3) Warning is only displayed when the function is called using a HART-Command
Warning Code &
Clear Text Message
Priority
Description
Possible Causes
Corrective Measures
Warning: 1
Q Simulation
2
Before starting the flowrate, a flowrate value can
be simulated. The outputs correspond to the
simulated value set.
Simulation Mode turned on
After the Simulation program has
been completed, the parameter
„Simulation Mode“ should be
turned off.
Warning: 2
Totalizer reset
1
All totalizers (forward/reverse, difference and over-
flow counters) were reset
External totalizer reset was
initiated
Open switch at contact input
(terminals 81, 82)
Warning: 3
Test Mode
3
Only for testing a converter with Simulator
55XC4000.
Test Mode turned on
After completing the Simulation
test with the Simulator the
parameter „Test Mode“ is to be
turned off.
Warning: 4
Self Check
4
If with HART-Communication a contact output or
contact input test was initiated,
Warning "4" is displayed
Function test running
After completing the self check
test, exit the routine.
Warning: 7
1)
Warning primary data
loaded
9
Converter has recognized different flowmeter pri-
mary data and loaded it in the internal FRAM.
Loaded are the system and flowmeter primary
data
Defective FRAM, Repair
instrument, replace
instrument
Record old totalizer values,
totalizer should be reset.
Warning: 8a
1)
Update internal FRAM
10
Converter has recognized an error in the internal
FRAM and has repaired the contents and data
from the external FRAM
Defective FRAM, Repair
instrument, replace
instrument
Check the entry settings and
correct, if required.
Warning: 8b
1)
Update external FRAM
11
Converter has recognized an error in the external
FRAM and has repaired the contents and data
from the internal FRAM
Defective FRAM or data was
changed.
Check the entry settings and
correct, if required.
Warning: 9a
Overflow > Forward
5
The forward totalizer has reached the maximum
count limit and has initiated a reset and incre-
mented the overflow counter.
Max. count value exceeded,
1 overflow is counted.
Reset totalizer if necessary.
Warning: 9b
Overflow < Reverse
6
The reverse totalizer has reached the maximum
count limit and has initiated a reset and incre-
mented the overflow counter.
Max. count value exceeded,
1 overflow is counted.
Reset totalizer if necessary.
Overflow: 9c
Difference totalizer
7
The difference totalizer has reached the maximum
count limit and has initiated a reset and increment-
ed the overflow counter.
Max. count value exceeded,
1 overflow is counted.
Reset totalizer if necessary.
Warning: 10
Reverse Q
8
If the forward flow direction is set and reverse flow
occurs, this Warning is displayed
Flow direction is reverse,
possibly a defective check
valve.
Prevent reverse flow of change
flow direction setting to forward/
reverse.
Warning: 11
2)
Instrument Address
not zero
12
The HART Instrument Address was changed from
0. Current output is frozen at 4 mA
Address 1-15 selected.
Therefore the current
output frozen at 4 mA .
Select Address 0, if a 4-40 mA
output is required.
Warning. 12a
2)3)
Simulation Iout
13
Self Check current output. Check the value of the
current at the connection terminals with a digital
voltmeter or the process instrumentation.
Function test, enter current
output directly in mA for
simulation.
Exit function test current output
Warning. 12b
2)3)
Simulation pulse
output
14
Self Check pulse output. Check the value of out-
put frequency with a counter.
Pulse output simulation
turned on
Exit function test pulse output
Warning. 13
2)3)
Automatic adjustment
running
15
Warning displayed only in Service Code section
when an adjustment has been initiated (only
accessible to ABB Service).
Only for
ABB Service Personnel
Only for ABB Service Personnel
Warning. 14
Measurement value
halted
16
Warning is only displayed when the function
"Noise reduction" has been activated (only acces-
sible to ABB Service)
Noisy measurement signal
Please contact factory Service