Abbott FreeStyle Navigator II User Manual - page 16
27
28
AC Adapter 5W USB
Abbott Diabetes Care PRT20992
Plug Kit AC adapter OUS
Abbott Diabetes Care PRT20993
Cable
Abbott Diabetes Care PRT23557-001 (ADC)
Ensure that you are using a personal computer that complies with the requirements of EN60950-1 .
Glossary of Symbols
Do Not Re-Use
Manufacturer
Sterile
Consult Operating Instructions
Use By
Batch Code
Serial Number
Catalogue Number
Caution
Authorised Representative in the European Community
Temperature Limitation
CE Mark
Do not use if package is damaged
IP27
Transmitter can withstand immersion into one metre of water for
up to 30 minutes
Type BF applied part
Keep dry
Non-ionising radiation
System Specifications
Your FreeStyle Navigator II System specifications are listed in the following table:
Operating Temperature
4 ºC to 40 ºC (41 ºF to 104 ºF)
Storage Temperature
Store the Transmitter and Receiver between -10 ºC (14 ºF) and 45 ºC (113 ºF) . Store Sensor Inserters and
Sensor Support Mounts between 3 ºC (37 ºF) and 30 ºC (86 ºF) . Store test strips between 4 ºC (40 ºF) and
30 ºC (86 ºF) . Control solution should be stored between 2 ºC (36 ºF) and 30 ºC (86 ºF)
Operating Humidity (Receiver)
10% to 93% (non-condensing)
Operating and Storage Altitude
Sea level to 3,048 metres (10,000 feet)
Operating Pressure
1 .0 bar (14 .7 psia, sea level) to 0 .7 bar (10 .1 psia, 3,048 metres (10,000 feet))
Sensor Life
Up to 5 days
Sensor Operating Skin Surface Temperature
25 ºC to 40 ºC (77 ºF to 104 ºF)
Glucose Result Range
20 to 500 mg/dL (1 .1 to 27 .8 mmol/L)
Receiver Memory
60 days of normal use including continuous glucose readings (stored every 10 minutes) and daily blood
glucose readings
Transmitter Battery and Battery Life
CR2032 Lithium Coin Cell, non-replaceable; up to 1 year of daily use under typical conditions
Receiver Battery and Battery Life
One Lithium-ion rechargeable, non-replaceable; Maximum voltage 4 .1Volts . Typical capacity of 830mAhrs .
Battery is UL 1642 Compliant . Charge sufficient for up to at least 3 days of typical use .
Wearing Transmitter Under Water
Up to 1 metre (3 feet) under water for no more than 45 minutes
Haematocrit (checking your Blood Glucose)
15% to 65%
Transmitter Skin Contact Material
The Transmitter has a temperature probe made of stainless steel that contacts the skin during wear . Surgical
stainless steel contains 8 –14% nickel
Radio Frequency
Transmission range: Up to 3 metres (10 feet) through RF transparent materials; up to 30 metres (100 feet) in
the absence of RF reflective materials
Transmitter transmission frequency and power: less than 100uW (-10dBm) at 433 .6MHz from at least 3 metres
Receiver bandwidth: 102 .4kHz bandwidth at 433 .6MHz
ART22683-101_rev-B.indd 27-28
1/7/15 11:43 AM