Quick reference guide for the i-stat ® 1 handheld and i-stat chemistry cartridges for use with a clia certificate of waiver chem8+ crea g e3+ ec4+ 6+.
2 quick reference guide for the i-stat ® 1 handheld and i-stat chemistry cartridges for use with a clia certificate of waiver this quick reference guide is for use with the following clia-waived i-stat cartridges: chem8+, 6+, ec4+, e3+, crea, and g. These cartridges include various subsets of the fo...
3 control testing frequency • test one cartridge from each lot in each shipment upon receipt. • test a single cartridge from the refrigerator monthly. Select cartridge in the following order: chem8+, 6+, crea, ec4+, glucose, e3+ prepare control • control ampule: must be at room temperature for at le...
4 10. Withdraw the control fluid slowly into a 1 or 3 cc plain syringe until the syringe is half filled. Dispense 3 drops of control fluid onto a gauze pad. Draw fluid fill cartridge fold closure control testing procedure - continued review results • compare results on handheld to the value assignme...
5 prepare the sample • mix the sample tube by gently inverting it 2-3 times before filling a cartridge. Do not use a partially filled collection tube. • chem8+: test the sample within 10 minutes after collecting it. • 6+, ec4+, e3+, crea, and g: test the sample within 30 minutes after collecting it....
6 9. Fold the snap closure over the sample well until it clicks into place. Press on the tab. Do not press directly over the sample well. Pt: 12345 16:10 25janyy i-stat chem8+ na mmol/l 141 k mmol/l 4 . 2 cl mmol/l 1 0 4 ica mmol/l 1 . 2 2 t c o 2 m m o l / l 2 6 page pt: 12345 16:10 25janyy i-stat ...
7 abbott point of care inc. 400 college road east princeton, nj 08540 www.Abbottpointofcare.Com i-stat is a registered trademark of the abbott group of companies in various jurisdictions. 06f20-06 art: 716942-00k rev. Date: 13-feb-17 cartridge receipt • check the temperature strip included with each...