• Substances that release caustic vapors such as acids
should be analyzed in the smallest sample size possible to
still achieve satisfactory results. If vapors condense on the
VP-Series Moisture Analyzers housing parts, stop the
testing, wipe the surface clean with a suitable neutralizer
and place the unit in a fume hood capable of removing the
vapors. The flow system in the instrument is composed
of aluminum and brass fittings connected with clear
tubing. Check for compatibility of substances to be tested
before inserting them into the instrument.
• The user has the responsibility for carrying out
appropriate decontamination if hazardous material is spilt
on or inside the VP-Series Moisture Analyzers.
CLEANING
• Clean the VP-Series Moisture Analyzers according to the
cleaning instructions only. Use of strong detergents and
“oven cleaners” will damage the case and heater housing.
• Before using any cleaning or decontamination method
except those recommended by Arizona Instrument, check
with AZI Customer Service at 1-602-470-1414 or 1-800-
528-7411, to ensure that the proposed method will not
damage the equipment.
• The outside housing of the VP-Series Moisture Analyzers
should be cleaned with a mild household detergent such
as “409” and a soft, lint-free cloth. Paper towels should
never be used, as they will scratch the instrument’s finish
and the clear lens over the display. If methods that are
more stringent are required, call AZI Customer Service at
1-602-470-1414 or 1-800-528-7411 for recommendations.
• Cleaning of the flow system components is normally a
factory operation. If components become contaminated,
follow the directions contained in this manual beginning
on page 42.
• Every precaution has been taken to prevent contaminants
from entering the unit. However, if liquids or powders do
enter the case, call AZI Customer Service at 1-602-470-
1414 or 1-800-528-7411 for recommendations.
The user shall be liable and responsible for any damage that
arises in connection with the use this moisture analyzer.
ENVIRONMENTAL CONDITIONS
Storage and Shipping
• Temperature should be between 0 °C to +40 °C (+32 °F to
+104
°F).
• Relative Humidity should be between 10% and 80%.
OPERATIONAL CONDITIONS – INDOOR USE.
• Optimum results will be achieved when the unit is set on
a smooth, level surface in a non-condensing, non-
explosive environment of 0-40º C @ 50% relative
humidity and 0-31º C @ 80% relative humidity.
• Do not expose the moisture analyzer unnecessarily to
extreme temperatures, moisture, shocks, blows or
vibration.
UNPACKING THE MOISTURE ANALYZER
• After unpacking the moisture analyzer, check it
immediately for any visible damage as a result of rough
handling during shipment.
• Save the box and all parts of the packaging to use when
returning your moisture analyzer for calibration. Only the
original packaging provides the AZI recommended
protection for shipment. Before packing your moisture
analyzer for shipment, unplug all connected cables to
prevent damage during transit.
WARNING SYMBOLS
- Protective Ground
Identifies the connection on the Chassis and Power Inlet
Module for connection of the safety ground (green/yellow)
wire.
- Warning
Information or procedure that must be observed.
- Electrical Shock Hazard.
Observe all steps of the procedure to prevent electrical shock.
- Warning, Hot surface or area of possible severe
burns.
Use listed precautions when operating the instrument and
handling samples to prevent injury.