AccuTools DS-220 User Manual - Tank Capacity Mode
DS-220 User’s Guide
12
6. When the applied (or removed) weight exceeds the programmed amount, the alarm will
sound and the DONE indicator will flash on the display to indicate the program is complete.
7. Press any key to stop the alarm.
8. To cancel the RUN mode at any time, press the
key.
NOTE: The RUN mode can only be initiated from the SET mode. If you cancel the RUN mode by
pressing
, and wish to RUN again, press
(the previous set point will be restored) followed
by
again. The display will be zeroed when the program restarts. You may program the SET
mode without starting the RUN mode by pressing
after setting the program set point. The
new value will be saved until next time the SET mode is activated.
can be used during the
RUN mode similarly as described in the HOLD MODE section. The CM-500 Charging Module
provides additional functionality when using the RUN mode. Refer to the CM-500 Charging
Module User’s Guide for details.
Tank Capacity Mode
Sometime it is convenient to know the relative amount of material remaining in a container (tank).
The DS-220’s Tank Capacity Mode allows a continuous display of the percentage of material
remaining in a tank. In order to compute the percentage accurately, the DS-220 must be trained
to know the weights of an empty and full tank. To train the DS-220 Tank Capacity Mode, perform
the following steps:
Tank Capacity Mode Training
1. While holding
down, press
. The SET TANK indicator will turn on, and OFF will flash
on the display. This indicates that the Tank Capacity Mode is currently off.
2. Turn the Tank Capacity Mode on by pressing
. The display will flash ON, and the
Alternate Numeric Display and Percent Indicator will turn on.
3. Press
to proceed to the first step of training. 0% will flash in the Alternate Numeric
Display indicating that the weight of an empty container must be registered.
4. Apply an empty container (or a load representing the weight of an empty container) on the
platform. Wait 5 seconds for the load to stabilize, then press
. 100% will flash in the
Alternate Numeric Display indicating that the weight of a full container must be registered.