Connection of peripheral devices
Read the operating instructions of the devices to be connected very carefully and follow
the installation instructions of the respective products. The cable lengths of the devices
connected must not exceed 3m, otherwise correct operation cannot be guaranteed. En-
sure the connectors are seated properly. If you have bought your Pro-Line II PC with an
additional graphic card the display connections on the mother board are disconnected.
Connect the monitor to be used to one of the horizontally lying graphics card connectors.
These are to found in the lower area at the back of the computer. With connection to the
motherboard sockets for the built in graphic device your monitor will not function.
! Warning: With devices that provide the Pro-Line II with additional power via the con-
necting plug, the supply unit must also be checked according to EN60601-1.
Upgrade and installation of components
<< Remove the mains supply plug before opening the housing! >>
Any opening of the housing is solely permitted by one of our authorised company’s resp.
technicians. After opening the unit, and according to applicable legal requirements, a new
safety test must be carried out in accordance with EN60601-1.
Before opening the housing it is imperative to remove the mains supply plug and to wait for
approx. 1 minute until all electrical current inside the unit has dissipated. Otherwise there
will be a danger to life!!! A subsequent change to the configuration of the unit requires
authorisation by e-medic Support and only components that have been tested and valida-
ted by us may be used. An unauthorised change to the unit configuration will result in an
immediate loss of conformity and can lead to a danger to the health of the user and / or the
patients and other people as the electrical safety can no longer be guaranteed!
Action with the ingress of foreign objects and liquids
With the inadvertent ingress of foreign objects, liquids or with the development of conden-
sation on the unit it must immediately be separated from the mains supply by removing
the plug. Under no circumstances switch the unit on again and contact the in-house tech-
nician, resp. your e-medic dealer or contact our Support team.