Instruction 0019-9166 combined service manual for: dioxor ® ii monoxor ® ii monoxor ® ii h (high range) nonoxor ® ii rev. 7 – may 2010 bacharach, inc. 621 hunt valley circle, new kensington, pa 15068 phone: 1-800-736-4666 • fax: 724-334-5001 • web: www.Mybacharach.Com printed in the u,s.A..
Page 2
® bacharach, dioxor, monoxor, and nonoxor are registered trademarks of bacharach, inc. Copyright © bacharach 1997-2010, all rights reserved. A.
Page 3
Table of contents 1.0 service manual contents.................................................................4 2.0 maintenance ............................................................................................5 2.1 opening the case.............................................................
Page 4
Single gas instrument service manual page 4 bacharach, inc. Instruction 0019-9166 1.0 service manual contents this manual contains the following service information for the . . . Dioxor ii (part no. 0019-7038 & 0019-7043) monoxor ii (part no. 0019-7034 & 0019-7039) monoxor ii h (part no. 0019-8015) ...
Page 5
Single gas instrument service manual instruction 0019-9166 bacharach, inc. Page 5 2.0 maintenance 2.1 opening the case figure 2-1. Opening and closing the case equipment needed: • phillips-head screwdriver, no. 1 procedure: 1. Lay the instrument face down on a workbench; then detach the metal clip f...
Page 6
Single gas instrument service manual page 6 bacharach, inc. Instruction 0019-9166 2.2 closing the case figure 2-2. P.C.B. And chamber assem- bly removal and installation equipment needed: • phillips-head screwdriver, no. 1 procedure: caution when replacing the p.C.B. And chamber assembly, the p.C.B....
Page 7
Single gas instrument service manual instruction 0019-9166 bacharach, inc. Page 7 2.3 cleaning the inside of the chamber assembly is the only part of the instrument that needs to be cleaned. To clean the chamber assembly, proceed as follows: equipment needed: • screwdriver, no. 1 phillips head • cle...
Page 8
Single gas instrument service manual page 8 bacharach, inc. Instruction 0019-9166 figure 2-3. Chamber as- sembly, exploded view figure 2-3a. Fan-blade assem- bly set screw a. If disassembled, push the fan assembly onto the motor shaft until it seats (fig 2-3). Be sure the set screw (fig 2-3a) is not...
Page 9
Single gas instrument service manual instruction 0019-9166 bacharach, inc. Page 9 2.4 sensor replacement when the instrument can no longer be spanned, as described in section “3.0 sen- sor calibration,” the sensor has reached the end of its useful life, and should be replaced. The life of a sensor i...
Page 10
Single gas instrument service manual page 10 bacharach, inc. Instruction 0019-9166 2.5 printed circuit board (p.C.B.) replacement figure 2-4. Sensor & motor wiring diagram equipment needed (refer to section “5.0 replacement parts” for part nos.): • phillips-head screwdriver, no. 1 • printed circuit ...
Page 11
Single gas instrument service manual instruction 0019-9166 bacharach, inc. Page 11 2.6 chamber assembly servicing to disassemble the chamber assembly, proceed as follows: equipment needed (refer to section “5.0 replacement parts” for part nos.): • screwdriver, phillips head, no. 1 • screwdriver, 1/8...
Page 12
Single gas instrument service manual page 12 bacharach, inc. Instruction 0019-9166 3.0 sensor calibration 3.1 monoxor ii co calibration figure 3-1. Calibration kit setup the following calibrates the monoxor ii to a known concentration of carbon monoxide. Equipment needed (refer to section “5.0 repla...
Page 13
Single gas instrument service manual instruction 0019-9166 bacharach, inc. Page 13 3. Connect the co discharge hose of the calibration equipment to the monoxor's probe tip; then open the co cylinder regulator valve and adjust the regulator for a fl owmeter indication of 1 to 2 scfh. 4. Allow gas to ...
Page 14
Single gas instrument service manual page 14 bacharach, inc. Instruction 0019-9166 4.0 troubleshooting chart symptom cause remedy batteries dead batteries installed backwards p.C.B. Defective backlight or its switch is defec- tive sensor connections loose or broken sensor membrane blocked with water...
Page 15
Single gas instrument service manual instruction 0019-9166 bacharach, inc. Page 15 symptom cause remedy erratic display unit won't zero and/or span low batteries faulty motor sensor connections loose or broken sensor depleted or defective p.C.B. Defective too much gas present to zero probe kinked or...
Page 16
Single gas instrument service manual page 16 bacharach, inc. Instruction 0019-9166 symptom cause remedy sensor connections loose motor has excessive arcing in its carbon brush/commutator interface causing rfi chamber assembly screws loose fan-blade assembly loose inlet and/or outlet port(s) blocked ...
Page 17
Single gas instrument service manual instruction 0019-9166 bacharach, inc. Page 17 5.0 replacement parts replacement parts for the instrument are listed below. Please refer to the ap- propriate fi gure(s) for part locations. Item description part no. 1 probe assembly, standard 0019-3076 -- probe ass...
Page 18
Single gas instrument service manual page 18 bacharach, inc. Instruction 0019-9166 5.2 instrument assembly (figs 5-2 & 5-3) figure 5-2. Replacement parts, chamber assem- bly exploded view item description part no. 1 adapter, motor shaft 0024-0412 2 battery clip, push on 0004-3113 3 battery cover wit...
Page 19
Single gas instrument service manual instruction 0019-9166 bacharach, inc. Page 19 figure 5-3. Replacement parts, back and front exploded views.
Page 20
Description part no. Calibration kit 0024-7059 carbon monoxide gas cylinder, 500 ppm 0024-0492 potentiometer adjustment tool 0006-9456 custom carrying case (1 instrument) 0019-0377 custom carrying case (2 instrument) 0019-3040 instruction manual, (dioxor ii) 0019-9121 instruction manual, (monoxor ii...