Bacharach Monoxor II Instruction

Manual is about: HCAC & Refrigeration

Summary of Monoxor II

  • Page 1

    Monoxor ii instruction 0019-9113 portable co analyzer part no. 0019-7034 & 0019-7039 rev. 8 - may 2010 bacharach, inc. 621 hunt valley circle, new kensington, pa 15068 ph: 1-800-736-4666 • fax: 724-334-5001 • web: www.Mybacharach.Com e-mail: help@mybacharach.Com printed in u.S.A. ® registered tradem...

  • Page 2

    Warranty bacharach, inc. Warrants to buyer that at the time of delivery this product will be free from defects in mater ial and manufac ture and will conform substan tially to bacharach inc.’s ap plicable specifi cations. Bacharach’s liability and buyer’s remedy under this warranty are limited to th...

  • Page 3

    1 warning! Because this instrument is used to detect and monitor materials and conditions which are listed by osha or others as potentially hazardous to personnel and property, the information in this manual must be fully understood and utilized to ensure that the instrument is operating properly an...

  • Page 4

    2 figure 2. Battery installation 2 technical characteristics co display range ............... 0-1999 ppm co accuracy .............................. ±10 ppm or ±5% of reading whichever is greater* response time ..................... 90% of fi nal value within 40 sec. Battery requirements ........... ...

  • Page 5

    Figure 3. Probe installation 3 if batteries are accidently installed in the wrong polarity, a positive temperature coeffi cient (ptc) thermistor will protect the instrument’s electronic circuitry. The instrument will operate once batteries are properly installed and the ptc thermistor is allowed to ...

  • Page 6

    4 4 operation to operate the monoxor ii, you simply . . . • set its power switch to on, • wait for the instrument to warm up (approx. 1 minute), • zero the display (if necessary), • take a gas sample. Detailed operating procedures are presented below: 4.1 power on/off turn on the instrument by slidi...

  • Page 7

    5 4.3 backlight on/off the lcd can be read in low-light areas by setting the front panel light switch to on. The backlight stays on until turned off, or until the power switch is set to off. 4.4 using the strap the instrument's elastic strap allows the unit to be either hand-held, or hung on nearby ...

  • Page 8

    6 overrange when the gas sample exceeds 1999 ppm, a “1” is displayed on the lcd. To clear an overrange condition, leave instrument turned on and sample fresh air until the lcd returns to displaying co. Low battery indications when the battery voltage becomes low, the “lo bat” indicator appears. Alth...

  • Page 9

    6 parts / service 6.1 parts/accessories list item part no. Battery cover 0019-3029 probe/hose/line dryer assy. 0019-3084 flexible probe tube (optional) 0019-3104 line dryer filter packing 0011-0122 6.2 bacharach sales / service centers 7 united states bacharach, inc. 621 hunt valley circle new kensi...

  • Page 10

    7 hazards of carbon monoxide carbon monoxide poisoning results in headache, nausea, chronic tired- ness, confusion, dizziness, and sometimes coma or death. It effects people by cutting off the supply of oxygen to their muscles and brain. The harmful effects of carbon monoxide exposure depend on both...