    Lc-22c march, 2000 mf-9010 instruction manual temperature controllers for liquid chromatography b i o a n a l y t i c a l s y s t e m s , i n c 2 7 0 1 ke n t av e n u e w e s t l a f a y e t t e i n d i a n a 4 7 9 0 6.

  • Page 2

    Manufacturer’s note the bas logo is a registered trademark of bioanalytical systems, inc. Copyright march 2000 no portion of this manual may be reproduced without express, written permission of bioanalytical systems, inc. All rights reserved bioanalytical systems, inc. 2701 kent avenue west lafayett...

  • Page 3: Lc-22C

    Section 1. Preface the bioanalytical systems lc-22c temperature controller and lc-23a and lc-23c column heating compartments are versatile tools for the experimenter who desires precise, reproducible temperature control for liquid chromatography. The lc-22c dual channel temperature controller is a g...

  • Page 4

    Section 2. Power requirements figure 2. Lc-22c back panel. Lc-22c controller this instrument can be operated from either 115v/50-60hz or 230v/50-60hz. The line voltage selector for 115v or 230v operation is on the back chassis panel. The line voltage must be selected before applying power to the con...

  • Page 5

    Section 3. Installation lc-22c controller this section concerns installation for stand-alone operation of the lc-22c with the optional universal probe. For operation of the other accessories, see subsequent sections. 1. Use the cable adapter (ef-1575) to convert the special safety receptacle on the ...

  • Page 6

    Lc-23a column heating compartment 1. Place the lc-23a heating compartment in the desired location. Six holes with captive 6-32 threads are provided for secure mounting on your liquid chromatograph, if desired. The holes are positioned to fit early model (pre 1986) bas liquid chromatography support s...

  • Page 7

    Lc-23c column heating compartment the lc-23c is specifically designed to fit within the bas cc-5 column compartment. The cc-5 is pre-drilled to accept a rheodyne injection valve on the front panel. It has several grooves on both sides for convenient access to external detectors. If you are installin...

  • Page 8

    Installing the lc-23c in a cc-5 compartment 1. Lower the cc-5 lid and remove the utility cover from the back panel of the cc-5 cabinet (four screws). 2. On some models, pressed-in, threaded nuts will be located on the outside of the back panel (figure 6). These must be removed. Do not remove if they...

  • Page 9

    Making plumbing connections for simple arrangements, use the innermost trough of the lc-23c. This location minimizes detector and valve hookup tubing. If you are coupling columns together, the longer trough will be more useful. A special coil of tubing is provided with each lc-23c (see figure 7). At...

  • Page 10

    Section 4. Operation for purposes of instruction, we will reference channel 1, although channel 2 is identical. 1. With the heater and probe cables disconnected, turn on the main power to the lc-22c. 2. Rotate the display selector knob to illuminate “set 1" in the red display window. 3. Using the “s...

  • Page 11

    Universal liquid probe 1. Ensure that the stainless steel probe is plugged into the jacks marked “probe 1" or ”probe 2" and observe the color code used on the jacks. 2. Turn the main power on to the lc-22c. Rotate the display selector knob to “sens 1" or ”sens 2", depending upon which channel is bei...

  • Page 12

    Section 5. Safety and maintenance checks the lc-22c, lc-23a, lc-23c and cell preheater are all essentially maintenance free instruments. Bas does advise that you attempt to keep these pieces in good condition by following these suggestions: 1. Avoid corrosive laboratory environments. 2. Don’t stack ...

  • Page 13

    Terminals) will be shown. Note that the center and right decimal points are illuminated indicating a closed circuit. If your unit fails either step 3 or 4, please contact the bas service coordinator to return both the lc-22c and the probe used to make these tests for service. Do not return any equip...

  • Page 14

    Section 6. Specifications temperature setpoint 0.00°c to 99.9°c in 0.1°c increments temperature measurement -50°c to 115°c readout front panel led max. Load 75 watts, each channel repeatability 0.1°c with the same temperature probe. Output temperature calibration output is calibrated to 100 mv per °...