Section 2. POWER REQUIREMENTS
Figure 2.
LC-22C Back Panel.
LC-22C CONTROLLER
This instrument can be operated from either 115V/50-60Hz or 230V/50-60Hz. The line
voltage selector for 115V or 230V operation is on the back chassis panel. The line voltage
must be selected before applying power to the controller.
A 3-prong power cable on the LC-22C must be connected to a standard, grounded power
outlet for safety. If the instrument is operated from a power outlet without a ground
connection, an appropriate adapter should be used and the ground connection of this
adapter should be securely fastened to an earth ground.
LC-23A and LC-23C HEAT-
ING COMPARTMENTS
The LC-23A and LC-23C compartments are designed specifically for use with the LC-22C
Controller. Two cables (one for heater power and the other for the temperature sensor)
connect to the LC-22C to each heating device. If you have purchased a heating
compartment for use with an older model LC-22A, you may be unable to make the
connection directly to the controller without an additional adapter. A change in design
converted this from a 3-prong male plug to the current female plug. The adapter (EF-1575)
is available from BAS at an additional charge.
GND
LC-22C
Made in USA
GND
REMOTE
ADDRESS
OUTPUT
PROBE 1
PROBE 2
HTR 2
HTR 1
F1
F2
1 2 3 4
2
1
INH
Section 2. POWER REQUIREMENTS
LC-22C CONTROLLER
2