w / Air-Purifying Elements
Checking Airflow with the Airflow Indicator
(PA4AFI - sold separately)
With the blower switched ON and the filters/
cartridges mounted, take the free end of the
breathing tube in one hand, hold it upright and
place the Airflow Indicator into the end of the
tube. (See Figure 6).
Apply a light downward pressure to the Airflow
Indicator to get a reasonable seal at the
breathing tube end. Ensure that the air outlet
holes in the Airflow Indicator tube are not
blocked. Two hands may be used if preferred,
one to hold the breathing tube and one to hold
the Airflow Indicator.
The position of the ball in the Airflow Indicator
should be observed. If any part of the ball is
below the PASS LINE on the Airflow Indicator, check for:
• Blower malfunction.
• Clogged or damaged Air-Purifying filter elements on the HE filter. See
“Mounting and Replacing Cartridges on the Blower Unit” on page 7.
• Low battery or battery malfunction.
If the ball is completely above the PASS LINE on the Airflow Indicator, then the
system is ready for use.
WARNING
If the blower malfunctions during use in a hazardous area:
Remain calm and
LEAVE
the hazardous area immediately.
DO NOT
use a blower that fails the flow test (air flow indicator sold separately).
Use
ONLY
Bullard filter/cartridges which comply with and have the NIOSH
approval label and which are appropriate for the contaminant.
Failure to observe these warnings could result in death or serious injury.
EVA Series PAPR Air-Purifying Elements
Principle of Operation
The following filter/cartridge protection classification applies when used with any
EVA PAPR system.
Protection
Filter/Cartridge Type NIOSH / ANSI Color
Code for Cartridge Label
HE
PAPRFC3
Magenta
OV/CL/HC/
SD/CD/HF/HE
PAPRFC4
Olive and Magenta
HE particulate filters are 99.97% effective against all particulate aerosols.
Filters are supplied in quantities of six per box.
The following abbreviations are approved by NIOSH to indicate the particulates,
gases, or vapors which are removed by the gas/vapor cartridges:
HE High Efficiency Particulate Air Filter for Powered Air-Purifying Respirators
WARNING
Use only the filter/cartridge(s) described in the above table.
Do not change cartridges while in a hazardous atmosphere.
Incorrect cartridge selection will invalidate all performance statements and
approvals for this equipment.
Follow established cartridge change schedules to ensure that cartridges are
replaced before breakthrough occurs.
Failure to follow these warnings could result in death or serious injury.
Mounting and Replacing Filters
on the Blower Unit
High efficiency particulate filters must be
replaced when retained particles clog the
filters and reduce air flow below acceptable
levels, as indicated by testing with the Air
Flow Indicator as described at left.
To Replace Filters
• Remove the air-purifying element from its packaging, and inspect for
damage. If in doubt do not use.
• Check that the air-purifying element has not exceeded its “use-by” date.
• Check that the filter connecting thread and gasket are in good condition.
• Check that the air-purifying element is appropriate to the hazard. If in doubt
consult your respirator program administrator or supervisor.
• Check that the threads in the blower unit port are in good condition and
clear of contaminant.
• Screw the air-purifying elements into the receptacles (see Figure 8) until the
cartridge is hand tight. DO NOT OVERTIGHTEN.
• Check to see that the locking tab is secure. (see Figure 9)
To Replace Combination Filter/Cartridge
• Follow the steps above, but beware that the filter locking tab is beneath the
filter rim. (see Figure 10)
Installing and Removing the Belt on the
Blower Unit
To install the belt
• With the blower filter side down, orient the lever locks as shown in Figure 11
• Lay belt over blower as shown in Figure 12
• Rotate level locks until they are oriented as shown in Figure 13
To Remove the Belt
• With the blower filter side down, orient the lever locks as shown in
Figure 12
• Remove belt from blower
Figure 7
Figure 8
Figure 9
Figure 10
Locking
tab
Locking
tab
Figure 11
Figure 12
Figure 13
Figure 6