    Instruction manual cr10xtcr thermocouple reference revision: 1/03 c o p y r i g h t ( c ) 1 9 9 5 - 2 0 0 3 c a m p b e l l s c i e n t i f i c , i n c ..

  • Page 2: Warranty and Assistance

    Warranty and assistance the cr10xtcr thermocouple reference is warranted by campbell scientific, inc. To be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless specified otherwise. Batteries have no warranty. Campbell scient...

  • Page 3: Cr10Xtcr Table of Contents

    I cr10xtcr table of contents pdf viewers note: these page numbers refer to the printed version of this document. Use the adobe acrobat® bookmarks tab for links to specific sections. 1. General ........................................................................1 1.1 specifications..................

  • Page 4

    This is a blank page..

  • Page 5: 1.  General

    1 cr10xtcr thermocouple reference 1. General the model cr10xtcr probe contains a thermistor for measuring temperature and is used as a reference temperature for thermocouples measured with the cr10x. It is installed between the two terminal stripes in the wiring panel. After all the thermocouples ar...

  • Page 6: 3.  Installation and Wiring

    Cr10xtcr thermocouple reference 2 figure 1. Error produced by polynomial fit to published values 3. Installation and wiring the wiring diagram for the cr10xtcr is shown in figure 1. The probe requires a single-ended analog input channel and an excitation channel. After the cr10xtcr and thermocouple ...

  • Page 7: 4.  Programming

    Cr10xtcr thermocouple reference 3 cs i/0 cr10x wiring panel m ad e in u sa w ir in g pa ne l n o. G 1 2v po w er in se de f se de f earth cround g g h l ag h l ag e3 ag g g l ag h 4 7 8 5 9 10 6 11 12 ag g g g sw 12v ctrl sw 12v g g h l ag h l ag e3 ag g l ag h 1 1 2 2 3 4 3 5 6 g g h l ag h l ag e3...

  • Page 8: 5.  Instruction 11 Details

    Cr10xtcr thermocouple reference 4 example 1. Sample cr10x instructions 01: temp 107 probe (p11) 01: 1 rep 02: 1 in chan 03: 3 excite all reps w/exchan 3 04: 15* loc [:ref_temp] 05: 1 mult 06: 0 offset * proper entries will vary with program and input location. 5. Instruction 11 details understanding...

  • Page 9

    Cr10xtcr thermocouple reference 5 table 3. Temperature, resistance, and datalogger output temperature °c resistance ohms output °c -40.00 4067212 -39.18 -38.00 3543286 -37.55 -36.00 3092416 -35.83 -34.00 2703671 -34.02 -32.00 2367900 -32.13 -30.00 2077394 -30.18 -28.00 1825568 -28.19 -26.00 1606911 ...

  • Page 10

    Cr10xtcr thermocouple reference 6 table 4. Polynomial coefficients coefficient value c0 -53.4601 c1 90.807 c2 -83.257 c3 52.283 c4 -16.723 c5 2.211 6. Electrically noisy environments ac power lines can be the source of electrical noise. If the datalogger is in an electronically noisy environment, th...

  • Page 11

    This is a blank page..

  • Page 12

    Campbell scientific companies campbell scientific, inc. (csi) 815 west 1800 north logan, utah 84321 united states www.Campbellsci.Com info@campbellsci.Com campbell scientific africa pty. Ltd. (csaf) po box 2450 somerset west 7129 south africa www.Csafrica.Co.Za sales@csafrica.Co.Za campbell scientif...